1、Designation: E 1615 08Standard Test Method forIron in Trace Quantities Using the FerroZine Method1This standard is issued under the fixed designation E 1615; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision
2、. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of iron in therange from 0.01 to 0.2 g/g using FerroZine2reagent solution.The range may
3、 be extended through the use of a 5- or 10-cmcell or by suitable dilution of the sample solution.1.2 This test method is intended to be general for the finalsteps in the determination of iron and does not includeprocedures for sample preparation.1.3 The values stated in SI units are to be regarded a
4、sstandard. No other units of measurement are included in thisstandard.1.4 Review the current material safety data sheets (MSDS)for detailed information concerning toxicity, first-aid proce-dures, and safety precautions.1.5 This standard does not purport to address all of thesafety concerns, if any,
5、associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specificwarning statements, see 7.4.2. Referenced Documents2.1 ASTM Standards:3D 1193 Speci
6、fication for Reagent WaterE60 Practice for Analysis of Metals, Ores, and RelatedMaterials by Molecular Absorption SpectrometryE 180 Practice for Determining the Precision of ASTMMethods for Analysis and Testing of Industrial and Spe-cialty ChemicalsE 200 Practice for Preparation, Standardization, an
7、d Stor-age of Standard and Reagent Solutions for ChemicalAnalysisE 275 Practice for Describing and Measuring Performanceof Ultraviolet, Visible, and Near-Infrared Spectrophotom-eters3. Summary of Test Method3.1 This test method is based upon a photometric determi-nation of the FerroZine2complex with
8、 the iron (II) ion.4,5Thesample is dissolved in a suitable solvent and the iron is reactedwith FerroZine reagent solution which will convert the dis-solved iron compounds to form a magenta color iron (II)complex. The iron content of the sample solution is determinedby measurement of the magenta colo
9、r at 560 nm using asuitable photometer.4. Significance and Use4.1 This test method is suitable for determining traceconcentrations of iron in a wide variety of products, providedthat appropriate sample preparation has rendered the iron andsample matrix soluble in water or other suitable solvent. Eac
10、hsample matrix must be investigated for suitability using thistest method.4.2 This test method assumes that the amount of colordeveloped is proportional to the amount of iron in the testsolution. The calibration curve is linear over the specifiedrange.5. Interferences5.1 Any ion that absorbs light a
11、t 560 nm will interfere withthe determination. Anionic interferences include oxalate inconcentrations over 500 g/g, cyanide, and nitrate.45.2 Of copper, cobalt, calcium, magnesium, lead, silver,molybdenum, aluminum, nickel, zinc, arsenic, manganese,hexavalent chromium, trivalent chromium, divalent c
12、obalt andmonovalent copper are the only metals other than iron thatform colored species with FerroZine under test conditions. Atleast 1000 mg/L of the alkali metals and the alkaline earths hadno effect on the determination. Many heavy metals will reactwith FerroZine in competition with iron, but wit
13、h the excessreagent used in the test there is no effect on the results.45.3 The pH range of the final solution should be from 4 to9 to give the best test results.4,51This test method is under the jurisdiction of ASTM Committee E15 onIndustrial and Specialty Chemicals and is the direct responsibility
14、 of SubcommitteeE15.01 on General Standards.Current edition approved April 1, 2008. Published May 2008. Originallyapproved in 1994. Last previous edition approved in 2005 as E 161505.2FerroZine is a trademark of Hach Chemical Company.3For referenced ASTM standards, visit the ASTM website, www.astm.o
15、rg, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.4Stookey, L. L., “FerroZineA New Spectrophotometric Reagent for Iron,”Analytical Chemistry, Vol 42, No. 7, June 1970, pp. 779 7
16、81.5Gibbs, C. R., “Characterization and Application of FerroZine Iron Reagent asa Ferrous Iron Indicator,” Analytical Chemistry , Vol 48, No. 8, July 1976, pp.11971201.1*A Summary of Changes section appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700,
17、 West Conshohocken, PA 19428-2959, United States.5.4 All glassware used in this test method must be iron-freeand scrupulously clean by precleaning with dilute hydrochloricacid and FerroZine reagent solution followed by a water rinse.6. Apparatus6.1 Photometer, capable of measuring light absorption a
18、t560 nm and holding a 5-cm or 10-cm cell. Check theperformance of the photometer at regular intervals according tothe guidelines given in Practice E 275 and the manufacturersmanual.6.2 Absorption Cells, 5-cm or 10-cm light path.NOTE 1A discussion of photometers and photometric practice isgiven in Pr
19、actice E60.7. Reagents7.1 Unless otherwise indicated, it is intended that all re-agents shall conform to the specifications of the Committee onAnalytical Reagents of the American Chemical Society, wheresuch specifications are available.6Other grades may be used,provided it is first ascertained that
20、the reagent is of sufficientlyhigh purity to permit its use without lessening the accuracy ofthe determination.7.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean Type II or Type IIIreagent water conforming to Specification D 1193.7.3 Iron, Standard Solution
21、, 1 mL = 1 g Fe (see Note 2 andNote 3)Dissolve 0.1000 g of iron wire in 10 mL ofhydrochloric acid (HCl, 1 + 1) and 1 mL of saturated brominewater (400 mL water + 20 mL bromine). Boil until the excessbromine is removed. Add 200 mL of HCl, cool, and dilute to 1L in a volumetric flask. Dilute 10 mL of
22、this solution to 1 L.NOTE 2The preparation of this reagent is also described in PracticeE 200.NOTE 3As an alternative, the standard iron solution may be preparedby diluting 1.00 mL of commercially available iron standard stocksolution (1000 mg iron/L) to 1 L with water.7.4 FerroZine Reagent Solution
23、Contains FerroZine colorreagent 3-(2-pyridyl)-5,6-bis(4-phenylsulfonic acid)-1,2,4-triazine, monosodium salt, monohydrate, buffer, and a reduc-ing agent. (WarningThis solution contains thiols as reduc-ing agents. Wear butyl rubber or neoprene gloves whenhandling the solution and avoid inhalation of
24、the vapors.)7.4.1 Alternatively, the individual solutions can be preparedas described below.47.4.1.1 Reducing AgentHydroxylamine hydrochloride, 10percent by weight solution in hydrochloric acid: Dissolve 10 gof reagent grade hydroxylamine hydrochloride (NH2OH.HCL)in 30 g of deionized water in a plas
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME16152008STANDARDTESTMETHODFORIRONINTRACEQUANTITIESUSINGTHEFERROZINEMETHOD 使用 铁合金 测定 微量 标准 试验 方法

链接地址:http://www.mydoc123.com/p-529055.html