1、Designation: D 6851 02Standard Test Method forDetermination of Contact pH with Activated Carbon1This standard is issued under the fixed designation D 6851; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision.
2、A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This method is to be used in the determination of the pHof water on initial contact with activated carbon. This testmethod is not me
3、ant as a replacement for Test Method D 3838and may give a different value.1.2 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and to determ
4、ine theapplicability of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:D 1193 Specification for Reagent Water2D 1293 Test Methods for pH of water2D 3838 Test Method for pH of Activated Carbon3E 300 Practice for Sampling Industrial Chemicals43. Summary of Test Method3.1
5、 A sample of carbon is stirred with water and the pH ofthe suspension is measured.4. Significance and Use4.1 The determination of contact pH can be used as a simpleand facts measurement that can allow activated carbon produc-ers and users to have a standard method for assessing the effectvarious car
6、bons will have on the initial pH of the water incontact with the carbon. It has been determined that there is abias between this method and Test Method D 3838; they arenot equivalent.5. Interferences5.1 pH electrodes used to measure this quantity can even-tually become clogged over time with carbon
7、fines. Suitableelectrodes can be found which have a detachable junctionallowing the user when necessary to discard a fouled pHmembrane.5.2 Distilled water can become acidic on standing. Makesure that the water used meets the minimum requirements forASTM Type II water. Determine the pH as indicated i
8、n TestMethod D 1293.6. Apparatus6.1 A pH meter (ambient temperature of 25 C is assumed;if otherwise, temperature compensation is required for accuratepH measurement.6.2 A combination pH electrode, or glass-calomel elec-trodes used together.6.3 A 100 mL (TD) graduated cylinder6.4 A 250 mL glass beake
9、r (for each sample)6.5 A polymer-coated magnetic stir bar (for each sample)6.6 Magnetic stir plate6.7 A balance capable of accurately measuring to 0.1 g7. Reagents and Materials7.1 Distilled or de-ionized water that meets the ASTM TypeII requirements7.2 Buffers to calibrate the pH meter; typically p
10、H 4.0, pH7.0, and/or pH 10.08. Hazards8.1 The water in contact with the carbon may have either alow or high pH. Take precautions accordingly and wearnecessary protective equipment to prevent injuries from spillsand splashes.9. Sampling, Test Specimens, and Test Units9.1 Follow Practice E 300 in the
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD68512002STANDARDTESTMETHODFORDETERMINATIONOFCONTACTPHWITHACTIVATEDCARBON 活性碳 测定 接触 PH 标准 试验 PDF

链接地址:http://www.mydoc123.com/p-523773.html