《ASTM D6122-2013 1250 Standard Practice for Validation of the Performance of Multivariate Online At-Line and Laboratory Infrared Spectrophotometer Based Analyzer Systems《多变量联机和实验室基于.pdf》由会员分享,可在线阅读,更多相关《ASTM D6122-2013 1250 Standard Practice for Validation of the Performance of Multivariate Online At-Line and Laboratory Infrared Spectrophotometer Based Analyzer Systems《多变量联机和实验室基于.pdf(30页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D6122 13Standard Practice forValidation of the Performance of Multivariate Online, At-Line, and Laboratory Infrared Spectrophotometer BasedAnalyzer Systems1This standard is issued under the fixed designation D6122; the number immediately following the designation indicates the year ofor
2、iginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONOperation of a laboratory or process stream analyzer system t
3、ypically involves four sequentialactivities. (1) Analyzer CalibrationWhen an analyzer is initially installed, or after majormaintenance has been performed, diagnostic testing is performed to demonstrate that the analyzermeets the manufacturers specifications and historical performance standards. The
4、se diagnostic testsmay require that the analyzer be adjusted so as to provide predetermined output levels for certainreference materials. (2) CorrelationOnce the diagnostic testing is completed, process streamsamples are analyzed using both the analyzer system and the corresponding primary test meth
5、od(PTM). A mathematical function is derived that relates the analyzer output to the primary test method(PTM). The application of this mathematical function to an analyzer output produces a predictedprimary test method result (PPTMR). (3) Probationary ValidationOnce the relationship betweenthe analyz
6、er output and PTMRs has been established, a probationary validation is performed using anindependent but limited set of materials that were not part of the correlation activity. This probationaryvalidation is intended to demonstrate that the PPTMRs agree with the PTMRs to within user-specifiedrequir
7、ements for the analyzer system application. (4) General and Continual ValidationAfter anadequate number of PPTMRs and PTMRs have been accrued on materials that were not part of thecorrelation activity, a comprehensive statistical assessment is performed to demonstrate that thePPTMRs agree with the P
8、TMRs to within user-specified requirements. Subsequent to a successfulgeneral validation, quality assurance control chart monitoring of the differences between PPTMR andPTMR is conducted during normal operation of the process analyzer system to demonstrate that theagreement between the PPTMRs and th
9、e PTMRs established during the General Validation ismaintained. This practice deals with the third and fourth of these activities.1. Scope*1.1 This practice covers requirements for the validation ofmeasurements made by laboratory or process (online or at-line)near- or mid-infrared analyzers, or both
10、, used in the calculationof physical, chemical, or quality parameters (that is, properties)of liquid petroleum products. The properties are calculatedfrom spectroscopic data using multivariate modeling methods.The requirements include verification of adequate instrumentperformance, verification of t
11、he applicability of the calibrationmodel to the spectrum of the sample under test, and verificationof equivalence between the result calculated from the infraredmeasurements and the result produced by the primary testmethod used for the development of the calibration model.When there is adequate var
12、iation in property level, the statis-tical methodology of Practice D6708 is used to provide generalvalidation of this equivalence over the complete operatingrange of the analyzer. For cases where there is inadequateproperty variation, methodology for level specific validation isused.1.2 Performance
13、Validation is conducted by calculating theprecision and bias of the differences between results from theanalyzer system (or subsystem) produced by application of themultivariate model, (such results are herein referred to asPredicted Primary Test Method Results (PPTMRs), versus thePrimary Test Metho
14、d Results (PTMRs) for the same sampleset. Results used in the calculation are for samples that are not1This practice is under the jurisdiction of ASTM Committee D02 on PetroleumProducts and Lubricants and is the direct responsibility of Subcommittee D02.25 onPerformance Assessment and Validation of
15、Process Stream Analyzer Systems.Current edition approved May 1, 2013. Published June 2013. Originallyapproved in 1997. Last previous edition approved in 2010 as D6122 10. DOI:10.1520/D6122-13.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harb
16、or Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1used in the development of the multivariate model. Thecalculated precision and bias are statistically compared touser-specified requirements for the analyzer system applica-tion.1.2.1 For analyzers used in product release or prod
17、uctquality certification applications, the precision and bias re-quirement for the degree of agreement are typically based onthe site or published precision of the Primary Test Method.NOTE 1In most applications of this type, the PTM is the specification-cited test method.1.2.2 This practice does not
18、 describe procedures for estab-lishing precision and bias requirements for analyzer systemapplications. Such requirements must be based on the critical-ity of the results to the intended business application and oncontractual and regulatory requirements. The user must estab-lish precision and bias r
19、equirements prior to initiating thevalidation procedures described herein.1.3 This practice does not cover procedures for establishingthe calibration model (correlation) used by the analyzer.Calibration procedures are covered in Practices E1655 andreferences therein.1.4 This practice is intended as
20、a review for experiencedpersons. For novices, this practice will serve as an overview oftechniques used to verify instrument performance, to verifymodel applicability to the spectrum of the sample under test,and to verify equivalence between the parameters calculatedfrom the infrared measurement and
21、 the results of the primarytest method measurement.1.5 This practice teaches and recommends appropriate sta-tistical tools, outlier detection methods, for determiningwhether the spectrum of the sample under test is a member ofthe population of spectra used for the analyzer calibration. Thestatistica
22、l tools are used to determine if the infrared measure-ment results in a valid property or parameter estimate.1.6 The outlier detection methods do not define criteria todetermine whether the sample or the instrument is the cause ofan outlier measurement. Thus, the operator who is measuringsamples on
23、a routine basis will find criteria to determine that aspectral measurement lies outside the calibration, but will nothave specific information on the cause of the outlier. Thispractice does suggest methods by which instrument perfor-mance tests can be used to indicate if the outlier methods arerespo
24、nding to changes in the instrument response.1.7 This practice is not intended as a quantitative perfor-mance standard for the comparison of analyzers of differentdesign.1.8 Although this practice deals primarily with validation ofinfrared analyzers, the procedures and statistical tests describedhere
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD612220131250STANDARDPRACTICEFORVALIDATIONOFTHEPERFORMANCEOFMULTIVARIATEONLINEATLINEANDLABORATORYINFRAREDSPECTROPHOTOMETERBASEDANALYZERSYSTEMS

链接地址:http://www.mydoc123.com/p-521615.html