ASTM D5374-2013 Standard Test Methods for Forced-Convection Laboratory Ovens for Evaluation of Electrical Insulation《电气绝缘评估用实验室强制对流炉的标准试验方法》.pdf
《ASTM D5374-2013 Standard Test Methods for Forced-Convection Laboratory Ovens for Evaluation of Electrical Insulation《电气绝缘评估用实验室强制对流炉的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5374-2013 Standard Test Methods for Forced-Convection Laboratory Ovens for Evaluation of Electrical Insulation《电气绝缘评估用实验室强制对流炉的标准试验方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D5374 13 An American National StandardStandard Test Methods forForced-Convection Laboratory Ovens for Evaluation ofElectrical Insulation1This standard is issued under the fixed designation D5374; the number immediately following the designation indicates the year oforiginal adoption or,
2、 in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 These test methods cover procedures for evaluating thecharacteristics of forced-
3、convection ventilated electrically-heated ovens, operating over all or part of the temperaturerange from 20C above the ambient temperature to 500C andused for thermal endurance evaluation of electrical insulatingmaterials.1.2 These test methods are based on IEC Publication 216-4-1, and are technical
4、ly identical to it. This compilation of testmethods and an associated specification, D5423, have replacedSpecification D2436.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish app
5、ro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D2436 Specification for Forced-Convection Laboratory Ov-ens for Electrical Insulation (Withdrawn 1994)3D5423 Specification for Forced-Convection La
6、boratory Ov-ens for Evaluation of Electrical Insulation2.2 Other Document:IEC Publication 216-4-1 Guide for the Determination ofThermal Endurance Properties of Electrical InsulatingMaterials, Part 4Aging Ovens, Section 1Single-Chamber Ovens43. Terminology3.1 Refer to the terminology section of Speci
7、ficationD5423.4. Significance and Use4.1 Ovens used for thermal evaluation of insulating materi-als are to be capable of maintaining uniform conditions oftemperature and air circulation over the extended periods oftime that are required for conducting these tests. SpecificationD5423 specifies the pe
8、rmissible deviations from absoluteuniformity that have been generally accepted internationallyfor these ovens. These test methods include procedures formeasuring these deviations and other specified characteristicsof the ovens.5. Apparatus5.1 Multi-Point Recording Potentiometer, having provisionsfor
9、 at least nine iron-constantan or chromel-alumelthermocouples, with scale readings to 0.1C or less. Use of adata processor or a data logger is helpful in reducing thenumber of calculations required.5.2 Calibrated Iron-Constantan or Chromel-AlumelThermocouples, using 0.5-mm diameter or smaller wire a
10、ndhaving a junction size not over 2.5 mm long.5.2.1 If calibrated thermocouples are not available, usethermocouples made from a single spool of thermocouple wirethat gives values for temperature that do not differ from eachother by more than 0.2C when placed within 10 mm of eachother without touchin
11、g inside an oven chamber at 200C.5.3 It is acceptable to use a temperature measuring systemother than thermocouples and a potentiometer, provided thatthe sensitivity, accuracy, and response time are equivalent tothat of the equipment described above, and that the objectivesof 6.2.3 relative to minim
12、ization of heat conduction effects aremet.5.4 Thermal Lag Time Specimen, consisting of a solid brasscylinder, 10 mm in diameter and 55 mm long, with onejunction of a differential thermocouple soldered to the surfacemidway between the ends. The other junction of the thermo-couple must be capable of b
13、eing moved at least 80 mm awayfrom the brass cylinder. An appropriate temperature indicator1These test methods are under the jurisdiction of ASTM Committee D09 onElectrical and Electronic Insulating Materials and are the direct responsibility ofSubcommittee D09.17 on Thermal Characteristics.Current
14、edition approved Sept. 1, 2013. Published September 2013. Originallyapproved in 1993. Last previous edition approved in 2005 as D5374 93 (2005).DOI: 10.1520/D5374-13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual B
15、ook of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The last approved version of this historical standard is referenced onwww.astm.org.4Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036.Copy
16、right ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1(as in 5.1, or other) must be provided for indication oftemperature differences to the nearest 0.1C as measured by thedifferential thermocouple.5.5 Watt-Hour Meter, of the appropriate voltag
17、e and phase,capable of reading to the nearest 1.0 Wh or less.6. Procedures6.1 Rate of Ventilation:6.1.1 Summary of Test MethodThe rate of ventilation iscalculated using determinations of (1) the average powerrequired to maintain the oven at a given temperature with itsports open and (2) the average
18、power required to maintain theoven at the same temperature with its ports closed. The test isconducted at 100C and at the maximum temperature at whichthe oven is used.6.1.2 Seal all openings into the oven, including, but notnecessarily limited to, the vent ports, door, thermometer ports,and the spac
19、e around the blower shaft (if the blower motor ismounted externally).6.1.3 Install a watt-hour meter, as described in 5.5,intheoven electrical supply line.6.1.4 Install a temperature sensor, such as a thermometer, 2m to 3 m away from the oven, at least 1 m away from any solidobject, and approximatel
20、y level with the oven air intake. Usethe oven temperature indicator to measure the internal tem-perature of the oven.6.1.5 Raise the oven temperature to 100 6 2C. When thetemperature of the oven has stabilized, measure the consump-tion of power over a measured period of 30 to 40 min. Beginand end th
21、e measuring period at corresponding points of thecyclic temperature fluctuation; for example, the moment whenthe heaters are switched on by the thermostat in the case of an“on/off” control. Measure and record the room temperature,which must not vary by more than 2C during the test.6.1.6 Remove the s
22、eals to restore the oven to its normaloperating condition. If necessary, adjust the vents and dampersto positions estimated to provide the specified rate of ventila-tion.6.1.7 Repeat 6.1.5. The average ambient air temperaturemust be within 2C of the average ambient temperaturemeasured in 6.1.5.6.1.8
23、 Calculate the rate of ventilation in the oven using thefollowing equation:N 5 3.59 P22 P1!/VT! (1)where:N = number of air changes per hour,P1= average power consumption, with no ventilation, ob-tained by dividing the energy consumption determinedfrom the watt-hour meter readings by the duration oft
24、he test in hours, W,P2= average power consumption during ventilation, calcu-lated in the same manner, W,V = total volume of air circulated within the oven, m3(seeNote 1), = density of the ambient room air during the test, kg/m3(see Note 2), andT = difference in temperature between the oven and theam
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便