1、Designation: D4127 12D4127 18Standard Terminology Used withIon-Selective Electrodes1This standard is issued under the fixed designation D4127; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in p
2、arentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope Scope*1.1 This terminology covers those terms recommended by the International Union of Pure and Applied Chemistry (IUPAC),2and is intended to provid
3、e guidance in the use of ion-selective electrodes for analytical measurement of species in water,wastewater, and brines.1.2 This international standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Developme
4、nt of International Standards, Guides and Recommendations issuedby the World Trade Organization Technical Barriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:3D1129 Terminology Relating to Water3. Terminology3.1 DefinitionsDefinitions: For other definitions used in this termi
5、nology, refer to Terminology D1129.3.1.1 For definitions of terms used in this standard, refer to Terminology D1129.3.2 Definitions of Terms Relevant to Ion-Selective Electrode Technology:acid error, nin very acid solutions, the activity of water is reduced (less than unity) causing a non-Nernstian
6、response in glasselectrodes.DISCUSSIONA positive error in the pH reading results.activity, nthe thermodynamically effective concentration of a free ion in solution.DISCUSSIONIn dilute solutions, ionic activity, and concentration are practically identical, but in solutions of high ionic strength, or
7、in the presence of complexingagents, activity may differ significantly from concentration. Ionic activity, not concentration, determines both the rate and the extent of chemicalreactions.activity coefficient, na factor, , that relates activity, A, to the concentration, C of a species in solution:A 5
8、CDISCUSSIONThe activity coefficient is dependent on the ionic strength of the solution. Ions of similar size and charge have similar activity coefficients.1 This terminology is under the jurisdiction ofASTM Committee D19 on Water and is the direct responsibility of Subcommittee D19.05 on Inorganic C
9、onstituents in Water.Current edition approved Aug. 1, 2012Feb. 1, 2018. Published August 2012May 2018. Originally approved in 1982. Last previous edition approved in 20062012 asD4127 06.D4127 12. DOI: 10.1520/D4127-12.10.1520/D4127-18.2 Recommendations for Nomenclature of Ion-Selective Electrodes, I
10、UPAC Commission on Analytical Nomenclature, Pergamon Press, Oxford, 1976.3 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the A
11、STM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consu
12、lt prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken,
13、PA 19428-2959. United States1alkaline error, nin alkaline solutions, where hydrogen ion activity becomes very small, some glass electrodes respond to othercations, such as sodium.DISCUSSIONA negative error in the pH reading results. By changing the composition of the glass, the affinity of the glass
14、 for sodium ion can be reduced. Suchelectrodes are known as lithium glass, high-pH, or full-range electrodes.analate, nthe sample being analyzed; used in the terms “analate addition” and “analate subtraction.”DISCUSSIONThis term differs from the term “analyte,” which describes the chemical species o
15、f interest in an analytical test.asymmetry potential, nthe potential across a glass pH electrode membrane when the inside and outside of the membrane arein contact with solutions of identical pH.DISCUSSIONThis term has also been used to define the observed potential differences between identical ele
16、ctrode pairs placed in identical solutions.calibration curve, na plot of the potential (emf or E) of a given ion-selective electrode cell assembly (ion-selective electrodecombined with an identified reference electrode) versus the logarithm of the ionic activity (or concentration) of a given species
17、.DISCUSSIONFor uniformity, it is recommended that the potential be plotted on the ordinate (vertical axis) with the more positive potentials at the top of the graphand that pa A (log activity of the species measured, A) or pcA (log concentration of species measured, A) be plotted on the abscissa (ho
18、rizontal axis)with increasing activity to the right. See Fig. 1 and Fig. 2. Region I of Fig. 1 represents the segment of the curve where the potential no longer changesin response to changes of the measured species activity and the electrode no longer demonstrates Nernstian response.IUPACactivity st
19、andard, na standardizing solution whose value is reported in terms of ionic activity.DISCUSSIONIf the electrode is calibrated using activity standards, the activity of the free, unbound ion in the sample is determined.concentration standard, na standardizing solution whose value is reported in terms
20、 of total concentration of the ion of interest.DISCUSSIONIf the electrode is calibrated using pure-concentration standards and measurements made on untreated samples, results must be corrected for the sampleionic strength and the presence of complexing agents. More commonly, a reagent is added to al
21、l standards and samples before measurement in orderto fix the ionic strength, thus avoiding the need for bination electrode, nan electrochemical apparatus that incorporates an ion-selective electrode and a reference electrode ina single assembly thereby avoiding the need for a separate reference ele
22、ctrode.FIG. 1 Example Response Curve (log activity on the horizontal axis)D4127 182concentration, nthe actual amount of a substance in a given volume of solution.DISCUSSIONWhen measuring ionic concentrations by electrode, a distinction is made between the concentration of the free, unbound ion, and
23、total concentrationthat includes ions bound to complexing agents.dissociation constant, na number indicating the extent to which a substance dissociates in solution.DISCUSSIONFor a simple two-species complex AB, the constant is given by the product of the molar concentrations of A and of B divided b
24、y the molarconcentrations of the undissociated species AB. For example, with hydrofluoric acid:H1#F2#!/HF#!5K 56.731024 at 25CThe smaller the value of K, the less the complex is dissociated. K varies with temperature, ionic strength, and the nature of the solvent.drift, nthis is the slow nonrandom c
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD41272018STANDARDTERMINOLOGYUSEDWITHIONSELECTIVEELECTRODES 离子 选择 电极 使用 标准 术语 PDF

链接地址:http://www.mydoc123.com/p-516336.html