FORD FLTM BU 017-1-2000 NONDESTRUCTIVE INSPECTION (NDI) OF ADHESIVE BONDS.pdf
《FORD FLTM BU 017-1-2000 NONDESTRUCTIVE INSPECTION (NDI) OF ADHESIVE BONDS.pdf》由会员分享,可在线阅读,更多相关《FORD FLTM BU 017-1-2000 NONDESTRUCTIVE INSPECTION (NDI) OF ADHESIVE BONDS.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、 FORD LABORATORY TEST METHOD BU 017-01 Date Action Revisions 2000 11 13 Revised Editorial no technical change A. Cockman 1992 08 10 Printed copies are uncontrolled Page 1 of 8 Copyright 2000, Ford Global Technologies , Inc. NONDESTRUCTIVE INSPECTION (NDI) OF ADHESIVE BONDS Application This procedure
2、 is used to locate unbonded regions in adhesively bonded plastic assemblies including fiber - reinforced composite assemblies. Apparatus and Materials Required Bond Tester Auto mation Industries model S - 2B Sondicator or equivalent, including: . Battery charger/eliminator and/or . Rechargeable batt
3、ery, . 25 kHz two - transducer/alarm probe, . Instruction manual. Reference Specimens Three adhesively bonded lap - joints are required. They must be made of the same type of adhesive and material as the assemblies to be inspected. Each must have minimum bond dimensions of 25 x 150 mm and the panel
4、size must be larger than 150 x 150 mm. Two of these reference specimens must have bonds of known acceptable bond integrity (selected according to the procedure described in Appendix A). 1. One should be kept with the instrument for hourly use as a reference or “calibration standard“. 2. The other sh
5、ould be retained as a “primary standard“ and kept in a protected place for daily comparison to the first, in order to detect unequal bond degradation of either reference specimen. The third reference specimen must contain an unbonded area about 25 x 25 mm. Refer to Appendix A for information on th e
6、 preparation, selection and certification of adhesive bond NDI reference specimens. FORD LABORATORY TEST METHOD BU 017-01 Page 2 of 8 Copyright 2000, Ford Global Technologies, Inc. Conditioning and Test Conditions All test values indicated herein are based on material conditioned in a controlled atm
7、osphere of 23 +/ - 2 C and 50 +/ - 5 % relati ve humidity for not less than 24 h prior to testing and tested under the same conditions unless otherwise specified. Procedure 1. Connect the 25 kHz dual transducer/alarm probe to the instrument. 2. Insert a charged battery pack or battery eliminator/c h
8、arger in the compartment located on the side of the instrument. Lock it in place. 3. Turn power on and select “PHASE + AMPTD“ as the alarm mode. Note: The BTRY indicator indicates a fully charged battery when the needle deflects fully to the right (w hite area). As the battery discharges, the needle
9、 will begin to drop, indicating that inspection time is becoming limited. To provide maximum instrument stability and battery life, the instruments power supply is programmed to automatically turn the ins trument off at a predetermined state of battery discharge. At this time, the BTRY indicator nee
10、dle drops to the center portion of the scale. The battery pack will rejuvenate itself to a degree, when line voltage is not available for recharging, if the in strument is switched off for approximately 15 or 20 minutes, possibly allowing the operator to finish a test. Battery Recharging To recharge
11、 the battery, proceed as follows: (a) Interconnect the battery eliminator/charger and the battery pack. Use the c harging cable supplied. The cable is coded for proper interconnection. (b) Connect the charger to a 115 volt AC, 60 Hz source using the power cable supplied with the instrument. (c) Mome
12、ntarily press the small red button at the rear of the charger to st art the charging cycle. When the battery has reached essentially full charge, the red light - emitting diode (LED) on the rear of the charger will light, indicating that the charging power has been reduced to trickle charge. The cha
13、rger will remain in this mode until AC power is disconnected. Approximately 10 hours are required to bring the battery to full charge. A full charge can provide up to eight hours of operation. To check for adequate battery charge, press the red button at the rear of the charge r, and then release. I
14、f the red LED does not go off when the button is released, the battery has accumulated a full charge. FORD LABORATORY TEST METHOD BU 017-01 Page 3 of 8 Copyright 2000, Ford Global Technologies, Inc. CAUTION DO NOT USE ANY OTHER MEANS OF CHARGING THE BATTERY THAN THE BATTERY ELIMINATOR/CHARGER SUPPLI
15、ED. THIS CHARGER IS DESIGNE D TO RECHARGE THE BATTERY ACCORDING TO THE BATTERY MANUFACTURERS RECOMMENDED SPECIFICATIONS. THE USE OF ANY OTHER TYPE OF CHARGER MAY DAMAGE THE BATTERY PACK. 4. Allow the instrument to equilibrate to ambient temperature which must be above 5 C. 5. All materials to be ins
16、pected or used as reference specimens must be clean and free of loose dirt, paint flakes, heavy grease or oil buildup, and other detrimental residues. 6. Adjust the warmed - up instrument according to the following operating instructions : (a) Place the probe down firmly over a known good area on th
17、e hourly reference or calibration standard. A probe hold -down force of approximately 9.8 N is recommended. The transducer contacts should be at least 6 mm in from the bond edge. Set the AMPL ITUDE potentiometer to 250 and the PHASE potentiometer to zero. Then rotate the phase potentiometer clockwis
18、e (CW) and observe as the phase meter peaks in deflection to the right and then falls back to the left. Continue rotating the PHASE potentiometer CW until it reaches the maximum stop, while counting the number, N, of full peaks observed on the phase meter. Then rotate the PHASE potentiometer counter
19、clockwise (CCW) until it again reaches its minimum stop at zero. Now, rotate the PHASE potentiometer CW again until N - 2 peaks on the phase meter have been observed and passed. Continue rotating the PHASE potentiometer CW, slowly approaching the N - 1 peak, until a phase meter reading of 0.6 is att
20、ained on the ascending side of the next - to - the - last peak. Readjust the AMPLITUDE potentiometer to obtain an amplitude meter reading of 0.7. Because these controls interact to a degree, continue readjusting them alternately until the meters read the prescribed values. Note: A fixed location on
21、the bond of both the hourly and daily reference specimens, and a fixed orientation of the transducer probe, should be identified (marked) and used for each referencing, or calibrating operation, in order to insure reproducible results. (b) The alarm level is set while the transducer probe is held do
22、wn firmly over the identified area on the hourly reference specimen. Turn the alarm switch ON. Rotate the ALARM LEVEL potentiometer clockwise until the alarm is activated. Then rotate the ALARM LEVEL knob CCW until the alarm turns off. Set the ALARM LEVEL knob two numbered scale divisions CCW from t
23、his point; i.e., if the alarm turns off when knob points to 7, setting for testing then is 5. (c) Check the alarm to ascertain that it is activated by a known unbond about 25 x 25 mm in size. FORD LABORATORY TEST METHOD BU 017-01 Page 4 of 8 Copyright 2000, Ford Global Technologies, Inc. 7. Monitor
24、the identified location on the daily reference specimen and record the amplitude and phase meter readings in a permanent log. A day - to - day comparison of these meter readings will provide monitoring of the continued acceptability of the reference specimens. 8. Inspect each bond by moving the tran
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- FORDFLTMBU01712000NONDESTRUCTIVEINSPECTIONNDIOFADHESIVEBONDSPDF

链接地址:http://www.mydoc123.com/p-747524.html