ASTM D6854-2003 Standard Test Method for Silica&8212 n-Dibutyl Phthalate Absorption Number《硅石的标准试验方法 正邻苯二甲酸二吸收值》.pdf
《ASTM D6854-2003 Standard Test Method for Silica&8212 n-Dibutyl Phthalate Absorption Number《硅石的标准试验方法 正邻苯二甲酸二吸收值》.pdf》由会员分享,可在线阅读,更多相关《ASTM D6854-2003 Standard Test Method for Silica&8212 n-Dibutyl Phthalate Absorption Number《硅石的标准试验方法 正邻苯二甲酸二吸收值》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 6854 03Standard Test Method forSilican-Dibutyl Phthalate Absorption Number1This standard is issued under the fixed designation D 6854; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A numb
2、er in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of then-dibutyl phthalate (DBP) absorption number of silica.1.2 The values stated in SI units ar
3、e to be regarded as thestandard. The values given in parentheses are for informationonly.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practice
4、s and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:D 1799 Practice for Carbon BlackSampling PackagedShipments2D 1900 Practice for Carbon BlackSampling Bulk Ship-ments2D 2414 Test Method for Carbon BlackOil AbsorptionNumber23. Summary o
5、f Test Method3.1 In this test method, DBP is added by means of aconstant-rate buret to a sample of silica in the mixer chamberof an absorptometer. As the sample absorbs the DBP, themixture changes from a free-flowing state to one of a semi-plastic agglomeration, with an accompanying increase invisco
6、sity. This increased viscosity is transmitted to the torque-sensing system of the absorptometer. The test is stopped whena predetermined torque level has been reached. Preferably thetorque versus volume of DBP is recorded by a penwriter or bya data acquisition system allowing a reliable determinatio
7、n ofthe endpoint. The volume of DBP per unit mass of silica is theDBP absorption number.4. Significance and Use4.1 The DBP absorption number of a silica is related to theprocessing and vulcanizate properties of rubber compoundscontaining the silica.5. Apparatus35.1 Balance, analytical, with 0.001-g
8、sensitivity.5.2 Sieve, 500 m (U.S. standard No. 35), having a 200-mm(8 in.) diameter and 25 mm (1 in.) height.5.3 Bottom Receiver Pan.5.4 Oven, gravity-convection type, capable of temperatureregulation within 61C at 125C and temperature uniformitywithin 65C.5.5 Spatula, rubber, 100-mm.5.6 Absorptome
9、ter,4equipped with a constant-rate buret thatdelivers 4 6 0.024 cm3/min.5.7 Desiccator, with silica gel as desiccant.6. Reagents and Standards6.1 n-Dibutyl Phthalate,5having a density of 1.042 to 1.047mg/m3(g/cm3) at 25C.6.2 Silica, commercial grade with a nitrogen surface area of175 6 10 m2/g.7. Sa
10、mpling7.1 Samples shall be taken in accordance with PracticesD 1799 and D 1900.8. Calibration8.1 Absorptometer6 The absorptometer is composed ofcomponents that influence calibration: the dynamometer torquespring or the load cell, the torque-limit or the indicator setpoint, the oil damper (absorptome
11、ters Type E and Type H areequipped with electronic damping), and the mixer-measuringhead.7It is necessary that each of the components be in goodcondition or proper adjustment to achieve acceptable calibra-tion.1This test method is under the jurisdiction of ASTM Committee D11 on Rubberand is the dire
12、ct responsibility of Subcommittee D11.20 on Compounding Materialsand Procedures.Current edition approved Jan. 10, 2003. Published February 2003.2Annual Book of ASTM Standards, Vol 09.01.3All apparatus is to be operated and maintained in accordance with themanufacturers directions for optimum perform
13、ance.4Available from C. W. Brabender Instruments, Inc., 50 E. Wesley St., Hacken-sack, NJ 07606, website: , and HITEC Luxembourg, 5 ruede lEglise, L-1458 Luxembourg, website: www.hitec.lu.5Technical grade has turned out to be suitable for the test, provided that thedensity is in the specified range.
14、6Mechanical absorptometers (type A or type B) can be used for the test;however, they are no longer commercially available. Refer to the instructions of thesupplier for calibration procedure.7The rotor motor speed is 1.31 rad/s (125 r/min).1Copyright ASTM International, 100 Barr Harbor Drive, PO Box
15、C700, West Conshohocken, PA 19428-2959, United States.NOTE 1Stainless steel mixing chambers8have been found satisfac-tory for this test when they are manufactured to a roughness average (Ra)of 2.5 6 0.4 m (100 6 15 in.) based upon eight measurements. Nosingle measurement should be greater than 3.6 m
16、 (140 in.) or less than1.5 m (60 in.). Stainless steel bowls purchased with an absorptometerhave been pre-polished for 16 h to minimize bowl surface changesaffecting calibration during their initial use. It is recommended that newreplacement stainless steel bowls should also be pre-polished to minim
17、izethe bowl surface effects on calibration (see Annex A1).8.1.1 The torque indicator is the primary component used tocorrect calibration. The load cell tension is adjusted by varyingthe alarm shut-off set point. Proper adjustment on the torqueindicator should provide repeatable values for a silica s
18、amplededicated to internal reference.8.1.2 The maximum torque span is set at 10 000 mNm(10 000 units) torque value. The torque-limit alarm is initiallyset at 5000 mNm (5000 units), but for testing silicas it will benecessary to adjust this setting to a lower value in order toobtain reproducible resu
19、lts. Use an internal silica sample witha nitrogen surface area of 175 6 10 m2/g to set the torque limitalarm which should correspond to approximately 70 % of themaximum torque developed during the test. After calibration,this setting should not be changed.NOTE 2It is generally recommended to use the
20、 absorptometer inconjunction with a penwriter or preferably with a data acquisition system(see 9.9 for further details).8.1.3 All digital signals are preset at 3 s damping for thetorque sensing system.8.1.4 Properly maintain the surface finish of the mixingchamber. If a new mixer chamber is installe
21、d, frequentlymonitor the instrument for any drift in calibration.8.2 Constant-Rate Buret:8.2.1 The delivery rate of the buret is to be 4 cm3/min. SeeAnnex A1 for detailed instructions on the procedure forcalibration check of the constant-rate buret.9. Procedure9.1 Pass a suitable amount of the sampl
22、e through Sieve No.35 (500 m), using a brush in order to deglomerate largerparticles. Use2gofthesieved material to test the moisturecontent (see 9.2).9.2 Weigh 12.5 g of the sample to the nearest 0.01 g.NOTE 3For silicas with an extraordinary high pour density it may benecessary to increase the samp
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD68542003STANDARDTESTMETHODFORSILICA8212NDIBUTYLPHTHALATEABSORPTIONNUMBER 硅石 标准 试验 方法 正邻苯二 甲酸 吸收

链接地址:http://www.mydoc123.com/p-523776.html