《ASTM D6579-2011 6250 Standard Practice for Molecular Weight Averages and Molecular Weight Distribution of Hydrocarbon Rosin and Terpene Resins by Size-Exclusion Chromatography《用尺寸筛.pdf》由会员分享,可在线阅读,更多相关《ASTM D6579-2011 6250 Standard Practice for Molecular Weight Averages and Molecular Weight Distribution of Hydrocarbon Rosin and Terpene Resins by Size-Exclusion Chromatography《用尺寸筛.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D6579 11Standard Practice forMolecular Weight Averages and Molecular WeightDistribution of Hydrocarbon, Rosin and Terpene Resins bySize-Exclusion Chromatography1This standard is issued under the fixed designation D6579; the number immediately following the designation indicates the year
2、 oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This practice covers the determination of apparentmolec
3、ular weight (MW) averages and molecular weight distri-butions (MWD) for THF-soluble hydrocarbon, rosin and ter-pene resins by size-exclusion chromatography (SEC). Thistechnique is not absolute; it requires calibration with standardsof known molecular weight. This practice is applicable toresins cont
4、aining molecular-weight components that have elu-tion volumes falling within the elution volume range definedby polystyrene standards.NOTE 1SEC is also known as gel permeation chromatography (GPC).1.2 SEC systems employ low-volume liquid chromatogra-phy components and columns packed with relatively
5、small(generally 3 to 20 m) microporous particles. High-performance liquid chromatography instrumentation and auto-mated data handling systems for data acquisition and process-ing are also required.1.3 The values stated in SI units are to be regarded as thestandard. The values given in parentheses ar
6、e for informationonly.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior
7、to use.2. Referenced Documents2.1 ASTM Standards:2D804 Terminology Relating to Naval Stores, Including TallOil and Related ProductsD3016 Practice for Use of Liquid Exclusion Chromatogra-phy Terms and RelationshipsD6440 Terminology Relating to Hydrocarbon Resins3. Terminology3.1 For definitions of si
8、ze-exclusion chromatography terms,see Practice D3016.3.2 For definition of terpene resin, see Terminology D804.3.3 For definitions of resin terms, see Terminology D6440.4. Summary of Practice4.1 In this practice, a dilute solution of a hydrocarbon, rosinor terpene resin sample is injected into a liq
9、uid mobile phasecontaining the same solvent used to prepare the resin solution.The mobile phase transports the resin into and through achromatography column (or set of columns connected inseries) packed with a rigid or semirigid, porous substrate thatseparates the molecules according to their size i
10、n solution. Adetector monitors the eluate as a function of elution volume (ortime). Upon emerging from the column(s), the fractions ofsize-separated molecules are detected and their elution vol-umes (or times) and (usually) concentrations recorded.Through calibration, the elution volumes (or times)
11、are con-verted to apparent molecular weights, and various molecularweight parameters for the sample resin are calculated from themolecular weight/concentration data.5. Significance and Use5.1 The MW averages and the MWD are important charac-teristics of a resin. They may be used for a variety ofcorr
12、elations for fundamental studies, processing, or productapplications. The MW and MWD values may also be used forproduction quality control of resins.5.2 LimitationsComparison of SEC molecular weight val-ues should be made only if the data were obtained underidentical chromatographic conditions.6. Ap
13、paratus6.1 Solvent ReservoirThe solvent reservoir must holdsufficient solvent to ensure consistency of composition for anumber of analyses. The reservoir should isolate the solvent1This practice is under the jurisdiction of ASTM Committee D01 on Paint andRelated Coatings, Materials, and Applications
14、 and is the direct responsibility ofSubcommittee D01.34 on Pine Chemicals and Hydrocarbon Resins.Current edition approved June 1, 2011. Published June 2011. Originallyapproved in 2000. Last previous edition approved in 2006 as D6579 06. DOI:10.1520/D6579-11.2For referenced ASTM standards, visit the
15、ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1*A Summary of Changes section appears at the end of this standard.Copyright ASTM International, 100 Ba
16、rr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.from the atmosphere, permit control of the environment incontact with the solvent, and be inert to the solvent employed.Some means of agitation (for example, magnetic stirring) isrecommended to ensure uniform composition.6
17、.2 Solvent Pumping SystemThe principal requirement ofthe pumping system is production of a relatively constant andpulseless flow of solvent through the columns. In general, therate of flow should be adjustable between 0.1 and 5.0 mL/min,and back pressures should not exceed limits specified by thecol
18、umn manufacturer. If the elution volume is not beingmeasured directly or corrected for systematic changes, theprecision in the flow rate must be at least 60.3 % under theconditions and over the time interval required for running atypical analysis.6.3 Sample InjectorThe purpose of an injection system
19、 isto generate a sharply defined zone of solution containing theresin when introducing the resin into the flow stream. Avalve-and-loop assembly or any of a number of commerciallyavailable high-performance liquid chromatography automaticinjection systems can be used for this purpose. It is requiredth
20、at contribution to band spreading be minimal and that theinjector be able to operate at the back pressure generated by thecolumns.6.4 ColumnsStainless steel columns with uniform andhighly polished inside walls should be used. Columns withlengths ranging from 15 to 50 cm, plus special end fittings,fr
21、its, and connectors designed to minimize dead volume andback-mixing are recommended. Micro-particulate, semirigidorganic gels, and rigid, solid, porous packing materials areused for SEC. Generally, the packing materials have narrowparticle size distributions, with particle sizes in the range from3 t
22、o 20 m. Packing materials are available in a variety ofshapes and pore sizes. Columns may be packed with particlesof relatively uniform pore size or with a “mixed bed” ofparticles to produce a broad range of pore sizes for polymerseparation. If a set of columns is used, it is recommended thatthe col
23、umns be connected, starting from the injector outlet,from columns having the smallest to those having the largestpacking pore size.NOTE 2Select the number of columns and pore sizes based on themolecular weight range of the resins being analyzed, and on the degree ofresolution required.6.5 DetectorsT
24、he purpose of the detector is to continu-ously monitor the concentration of solute eluting from thechromatographic column(s). Consequently, the detector mustbe sufficiently sensitive and respond linearly to the soluteconcentration. Additionally, the detector must not appreciablydistort the concentra
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD657920116250STANDARDPRACTICEFORMOLECULARWEIGHTAVERAGESANDMOLECULARWEIGHTDISTRIBUTIONOFHYDROCARBONROSINANDTERPENERESINSBYSIZEEXCLUSIONCHROMATOGRAPHY

链接地址:http://www.mydoc123.com/p-522923.html