《ASTM C1432-2003(2008) Standard Test Method for Determination of Impurities in Plutonium Acid Dissolution Ion Exchange Matrix Separation and Inductively Coupled Plasma-Atomic Emissi.pdf》由会员分享,可在线阅读,更多相关《ASTM C1432-2003(2008) Standard Test Method for Determination of Impurities in Plutonium Acid Dissolution Ion Exchange Matrix Separation and Inductively Coupled Plasma-Atomic Emissi.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C 1432 03 (Reapproved 2008)Standard Test Method forDetermination of Impurities in Plutonium: Acid Dissolution,Ion Exchange Matrix Separation, and Inductively CoupledPlasma-Atomic Emission Spectroscopic (ICP/AES) Analysis1This standard is issued under the fixed designation C 1432; the nu
2、mber immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scop
3、e1.1 This test method covers the determination of 25 ele-ments in plutonium (Pu) materials. The Pu is dissolved in acid,the Pu matrix is separated from the target impurities by an ionexchange separation, and the concentrations of the impuritiesare determined by inductively coupled plasma-atomic emis
4、sionspectroscopy (ICP-AES).1.2 This test method is specific for the determination ofimpurities in8MHNO3solutions. Impurities in other pluto-nium materials, including plutonium oxide samples, may bedetermined if they are appropriately dissolved (see PracticeC 1168) and converted to8MHNO3solutions.1.3
5、 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-p
6、riate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2C 757 Specification for Nuclear-Grade Plutonium DioxidePowder, SinterableC 758 Test Methods for Chemical, Mass Spectrometric,Spectrochemical, Nuclear,
7、and Radiochemical Analysis ofNuclear-Grade Plutonium MetalC 759 Test Methods for Chemical, Mass Spectrometric,Spectrochemical, Nuclear, and Radiochemical Analysis ofNuclear-Grade Plutonium Nitrate SolutionsC 1168 Practice for Preparation and Dissolution of Pluto-nium Materials for AnalysisD 1193 Spe
8、cification for Reagent Water3. Summary of Test Method3.1 A sample of plutonium metal is dissolved in a smallvolume of 6 M hydrochloric acid (HCl). Then, 10 M (HNO3)/0.03 M hydrofluoric acid (HF) is added to the dissolvedplutonium to oxidize the plutonium to the Pu (IV) state. Thesample solution is l
9、oaded onto a nitrate anion exchange resinand eluted with8MHNO3/0.006 M HF. The rinses contain thetarget metallic impurities and less than 15 g/mL Pu. Theplutonium is stripped from the anion exchange resin with 0.1M HCl. The rinses containing the metallic impurities areanalyzed by ICP-AES.4. Signific
10、ance and Use4.1 This test method can be used on plutonium matrices innitrate solutions.4.2 This test method has been validated for all elementslisted in Test Methods C 757 except sulfur (S) and tantalum(Ta).4.3 This test method has been validated for all of the cationelements measured in Table 1. Ph
11、osphorus (P) requires avacuum or an inert gas purged optical path instrument.5. Interferences5.1 Plutonium concentrations of less than 50 g/mL in thefinal aqueous phase do not significantly affect the analyticalresults for most elements. Interference studies should be madeto determine the degree of
12、Pu and other elemental interferenceson the target analytes; background and interelement correctionsmay be required.1This test method is under the jurisdiction of ASTM Committee C26 on NuclearFuel Cycle and is the direct responsibility of Subcommittee C26.05 on Methods ofTest.Current edition approved
13、 Dec. 1, 2008. Published January 2009. Originallyapproved in 1999. Last previous edition approved in 2003 as C 1432 03.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer
14、to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6. Apparatus6.1 An ICP-AES equipped with a Charge Injection Device(CID) detector or an ICP-AES with a spectral bandpass of 0.05
15、nm or less is required to provide the necessary spectralresolution.3The spectrometer may be either a simultaneousmultielement or a sequential spectrometer. The spectrometermay be either an inert gas-path or vacuum instrument; theappropriate spectral lines should be selected for each specificinstrume
16、nt. Either an analog or digital readout system may beused.6.2 The ICP-AES is interfaced to a glovebox. The torch boxis glovebox enclosed, since plutonium containing materialscome in direct contact with the torch. This setup is described inASTM STP 951.46.3 Vacuum manifold set at approximately 23 cm
17、Hg (9 in.Hg) is optional.5A gravity system is also acceptable.6.4 15 mL plastic disposable ion exchange columns.66.5 50 mL plastic vials.6.6 Plastic micro and macro pipettes.6.7 1000 mL plastic volumetric flasks.7. Reagents and Materials7.1 Purity of ReagentsReagent grade chemicals shall beused in a
18、ll tests. Unless otherwise indicated, it is intended thatall reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society(ACS), where such specifications are available.7Other gradescould be used, provided it is first ascertained that the reag
19、ent isof sufficiently high purity to permit its use without lesseningthe accuracy of the determination.7.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean laboratory accepteddemineralized or deionized water as described by Type 1 ofSpecification D 1193.7.3 U
20、ltra high purity acids shall be used for sample disso-lution and calibration standards preparation unless otherwisenoted.3The sole source of supply of the apparatus known to the committee at this timeis Thermo Jarrell Ash PolyScan Iris spectrometer (Thermo Electron Spectroscopy,Franklin, MA), or an
21、Applied Research Laboratories 3580 ICP-AES instrument(Dearborn, MI). If you are aware of alternative suppliers, please provide thisinformation to ASTM International Headquarters. Your comments will receivecareful consideration at a meeting of the responsible technical committee,1whichyou may attend.
22、4Edellson, M. C., and Daniel, J. Leland, “Plasma Spectroscopy of the Analysisof Hazardous Materials: Design and Application of Enclosed Plasma Sources,”Conference Proceedings, ASTM STP 951, ASTM, 1986.5The sole source of supply of the apparatus known to the committee at this timeis Eichrom Technolog
23、ies Vacuum Box System (Part # AC-24-BOX), EichromTechnologies Inc., Darien. IL. If you are aware of alternative suppliers, pleaseprovide this information to ASTM International Headquarters. Your comments willreceive careful consideration at a meeting of the responsible technical committee,1which you
24、 may attend.6The sole source of supply of the apparatus known to the committee at this timeis Ion exchange columns from either Applied Separation or Bio-Rad Inc. If you areaware of alternative suppliers, please provide this information to ASTM Interna-tional Headquarters. Your comments will receive
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC143220032008STANDARDTESTMETHODFORDETERMINATIONOFIMPURITIESINPLUTONIUMACIDDISSOLUTIONIONEXCHANGEMATRIXSEPARATIONANDINDUCTIVELYCOUPLEDPLASMAATOMICEMISSIPDF

链接地址:http://www.mydoc123.com/p-464635.html