1、AUTOMATED CRUSH STRENGTH OF EXTRUDED CATALYSTS UOP Method 973-01 SCOPE This method is for determining the crush strength of extruded and shaped catalysts using a Chatillon motorized sample stand and force gauge with a chisel anvil. The measurement range is from 1 Newton (N) (0.22 lbf) to 250 N (56 l
2、bf). The range can be extended to 450 N (101 lbf) when using an alternative force gauge. Test apparatus other than the Chatillon may be used provided comparable data can be obtained. For statistical purposes, the number of extruded pieces crushed is set at a minimum of fifty; however, more may be sp
3、ecified. Data generated using this method cannot be directly compared to those generated using UOP Method 914, Automated Crush Strength of Catalysts or Molecular Sieves. This method will yield higher values for crush strength. Proper comparisons can only be made after tests are run to determine the
4、offset between the two methods. Such offsets are expected to vary from product to product. OUTLINE OF METHOD A representative sample is riffled to obtain approximately, but not less than, fifty extrudates. This riffled sample is treated at 500C (930F) in a dish for one hour and then transferred in t
5、he dish to a hot plate set at 200C (400F) to avoid moisture pick-up. Each extrudate is placed on a flat plate at a perpendicular orientation to the chisel attachment that is 1 mm across at the contact point. It is then subjected to a compressive load, and the force required to crush the extrudate is
6、 recorded. The procedure is repeated until a minimum of fifty extrudates in the riffled sample have been crushed, and the average crush strength per extrudate is calculated. The percent of extrudates having a crush strength above or below a given value (product specification or customer driven) can
7、be reported, as well as the maximum and minimum crush strength for individual extrudates within the set. In addition, the crush strength data can be graphically displayed as a histogram. Results can be reported in either Newtons (N) or pounds-force (lbf) as measured using a 1-mm chisel. APPARATUS Re
8、ferences to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Brush, round, camels hair, Fisher Scientific, Cat. No. 03-670 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF
9、 REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIBED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR
10、 ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2001 UOP LLC ALL RIGHTS RESERVED UOP Methods are available through ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken PA 19428-2959,
11、United States. The Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE. 2 of 6 973-01 Calibration weights, Class 3, Fisher Scientific, Cat. No. 02-225-26P Computer, personal, for running Chatill
12、on software, and capable of running Excel, Minitab or similar software to develop a histogram and perform post-analysis calculations, JLW Instrument Dish, porcelain, Coors 60230, Fisher Scientific, Cat. No. 08-693A Forceps, blunt-pointed, straight, 127-mm length, Fisher Scientific, Cat. No. 08-890 H
13、ot plate, capable of 200C Muffle furnace, capable of 500C Printer, any model compatible with the computer Riffler, Fisher Scientific, Cat. No. 04-942D Sieves, wire cloth, brass frame, 8-inch diameter, 2-inch depth, U.S. Standard sieves Nos. 8, 10, 12, 16, 20, and 30, Fisher Scientific, Cat. Nos. 04-
14、881F, G, H, K, M, and P, respectively; with cover and pan, Fisher Scientific, Cat. Nos. 04-886A and B, respectively. The sieves meet current ASTM E 11 and ISO 565/3310-1 specifications. Software (optional), Minitab or other equivalent for displaying histogram, and Excel or equivalent for post-analys
15、is calculations Software (required), Chatillon, AutoTest, to control TCD-200 and collect data from DFA50/100 Surface temperature probe, Spot Check Thermometer, PTC Instruments, Model 573F Test stand and gauge, Chatillon, includes Cat. No. TCD-200 Motorized Test Stand, Cat. No. DFA50 Turbo gauge incl
16、uding 1 x 6-mm chisel anvil tip, Cat. No. DFA100 Turbo gauge (optional), JLW Instrument Tongs, crucible, Fisher Scientific, Cat. No. 15-207 Tray, utility, stainless steel, 420- x 250- x 64-mm, Fisher Scientific, Cat. No. 13-361C REAGENTS AND MATERIALS References to catalog numbers and suppliers are
17、included as a convenience to the method user. Other suppliers may be used. Reference catalysts, extrudate, the same types as samples routinely run, UOP PROCEDURE The Chatillon instrument manual must be thoroughly understood before operation or calibration. The instrument must be validated per Chatil
18、lon instructions. Read and understand DFA50, or DFA100, force gauge limits, and TCD-200 operational settings, overloads, etc. 3 of 6 973-01 It is recommended that the user select one or more extrudate reference catalysts, similar to those being analyzed, for weekly quality control evaluation (in wee
19、ks when samples are being analyzed). Sufficient reference material should be obtained to last for at least six months, preferably one to five years. The reference should be well riffled and stored in sealed containers to reduce moisture absorption. The reference data obtained can be used to develop
20、a control chart that is valuable in detecting analytical variations. Validate instrument performance per the manufacturers instructions monthly when samples are being analyzed, or when indicated by rule violations, trend shifts, or out of control SPC data. Instrument validation is accomplished by in
21、verting the force gauge and placing known weights on the platform and reading the applied force on the gauge LED. The gauge should be sent to the manufacturer or approved representative yearly for calibration and certification. Weight calibration tolerances: 0.5 kg: 4.9 0.4 N, 1 kg: 9.8 0.3 N, 2 kg:
22、 19.6 0.3 N. Instrument Setup 1. Set the Chatillon reference point to 18.5 mm, following manufacturers instructions. The reference point and travel distance may vary from site to site. Statistical equivalence must be determined before applying parameter variations. 2. Install the chisel anvil. 3. Se
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9732001AUTOMATEDCRUSHSTRENGTHOFEXTRUDEDCATALYSTSPDF

链接地址:http://www.mydoc123.com/p-1074594.html