SANS 6164-2009 Determination of lead (inorganic and tetra-alkyl) in workplace air by atomic absorption spectrophotometry《以原子吸收光谱法测定工作场所空气中的铅(无机铅及四烷基铅)》.pdf
《SANS 6164-2009 Determination of lead (inorganic and tetra-alkyl) in workplace air by atomic absorption spectrophotometry《以原子吸收光谱法测定工作场所空气中的铅(无机铅及四烷基铅)》.pdf》由会员分享,可在线阅读,更多相关《SANS 6164-2009 Determination of lead (inorganic and tetra-alkyl) in workplace air by atomic absorption spectrophotometry《以原子吸收光谱法测定工作场所空气中的铅(无机铅及四烷基铅)》.pdf(14页珍藏版)》请在麦多课文档分享上搜索。
1、 Collection of SANS standards in electronic format (PDF) 1. Copyright This standard is available to staff members of companies that have subscribed to the complete collection of SANS standards in accordance with a formal copyright agreement. This document may reside on a CENTRAL FILE SERVER or INTRA
2、NET SYSTEM only. Unless specific permission has been granted, this document MAY NOT be sent or given to staff members from other companies or organizations. Doing so would constitute a VIOLATION of SABS copyright rules. 2. Indemnity The South African Bureau of Standards accepts no liability for any
3、damage whatsoever than may result from the use of this material or the information contain therein, irrespective of the cause and quantum thereof. ISBN 978-0-626-21956-7 SANS 6164:2009Edition 1.1Any reference to SABS SM 1164 is deemedto be a reference to this standard(Government Notice No. 1373 of 8
4、 November 2002)SOUTH AFRICAN NATIONAL STANDARD Determination of lead (inorganic and tetra-alkyl) in workplace air by atomic absorption spectrophotometry Published by SABS Standards Division 1 Dr Lategan Road Groenkloof Private Bag X191 Pretoria 0001Tel: +27 12 428 7911 Fax: +27 12 344 1568 www.sabs.
5、co.za SABS SANS 6164:2009 Edition 1.1 Table of changes Change No. Date Scope Amdt 1 2009 Amended to change the designation from SABS to SANS, with no technical changes. Foreword This South African standard was approved by National Committee SABS TC 146, Air Quality, in accordance with procedures of
6、the SABS Standards Division, in compliance with annex 3 of the WTO/TBT agreement. This document was published in April 2009. This document supersedes SABS SM 1164:1990 (first edition). SANS 6164:2009 Edition 1.1 1 Determination of lead (inorganic and tetra-alkyl) in workplace air by atomic absorptio
7、n spectrophotometry 1 Scope This standard specifies a method for the determination of the lead in air samples collected by filters or impingers. The method is applicable to personal sampling as well as static sampling. 2 Principle 2.1 Filter (cellulose ester membrane) samples are wet-ashed using nit
8、ric acid and hydrogen peroxide. The ash is dissolved in dilute nitric acid and diluted to a known volume. 2.2 Impinger samples are quantitatively transferred to a volumetric flask and diluted to a known volume. 2.3 The diluted samples are aspirated into an air-acetylene flame of an atomic absorption
9、 spectrophotometer and the absorbances measured at 217,0 nm. NOTE The lead content may also be determined by means of other suitable analytical techniques, which include colorimetry, polarography, anodic stripping voltammetry, atomic absorption spectrophotometry with electrothermal atomization and i
10、nductively-coupled plasma (ICP) atomic emission spectrophotometry. 3 Personal sampling of lead, inorganic dusts and fumes 3.1 Equipment. A sampling train that includes the components given in 3.1.1-3.1.7 (inclusive). NOTE Assembly instructions are given in 3.2.3. 3.1.1 A battery-operated portable pu
11、mp, that a) is capable of delivering, for a period of 8 h, 2 L of air per minute, with a filter in the sampling train; NOTE The maximum deviation from the prescribed flow rate shall not exceed 5 %. b) is intrinsically safe; c) includes a calibrated flow meter or a mechanism for maintaining a constan
12、t flow rate; and d) has an on-off switch and flow rate control or adjustment mechanism that is tamperproof. SANS 6164:2009 Edition 1.1 2 3.1.2 Cellulose ester membrane filters, of diameter 37 mm and having a pore size of 0,8 m. 3.1.3 Filter support pads, of diameter 37 mm. 3.1.4 A closed-face filter
13、 holder, of the same size as the filters (see 3.1.2 and 3.1.3) fitted with an adaptor that provides a leakproof fit for connecting flexible tubing. 3.1.5 Flexible tubing, of sufficient length, and a) of appropriate wall thickness to prevent it collapsing when air is drawn through it; and b) of appro
14、priate inner diameter to provide a leakproof connection to the filter holder adaptor and to the inlet port adaptor of the portable pump. 3.1.6 A mercury thermometer, for ambient air temperature measurement, accurate to 1 C. 3.1.7 A barometer, for measuring atmospheric pressure, accurate to 5 mm merc
15、ury. 3.2 Procedure 3.2.1 Preparation of pumps a) Fully charge the pump battery pack in accordance with the manufacturers instructions, before use in any workshift. If the pump is used for more than one shift, substitute a fully charged battery pack for the used one. b) Calibrate flow meters or const
16、ant flow devices on the pumps at 2 L/min (using a primary standard such as a bubble meter and the procedures recommended in the manufacturers instructions, and with the filter and filter holder assembly in the line), 1) at intervals not exceeding 40 h of operation; 2) immediately after they have rec
17、eived unusually harsh or potentially damaging treatment; and 3) immediately after the pump or the sampling train has been disassembled for maintenance. 3.2.2 Preparation of filter holders a) Assemble the clean filter holders in a lead-free environment. b) Use only clean flat-tipped tweezers to handl
18、e the filter support pad and the filter. Place a filter support pad and then a cellulose ester filter in the base of a filter holder. c) Ensure that, if necessary, the joints of the filter holder are sealed with non-porous, shrinkable cellulose band or tape. d) Label the filter holder with an approp
19、riate sample identification. e) Use suitable plugs to seal the filter holder until it is placed in position in the field and sampling is ready to commence. SANS 6164:2009 Edition 1.1 3 3.2.3 Sampling a) Assemble the sampling train by fixing one end of the flexible tubing to the inlet port of the pum
20、p and the other end with the adaptor to the outlet port of the filter holder. b) Check for air leaks by switching on the pump. The air flow indicator on pumps equipped with a flow meter should drop to the zero flow position and oscillate slightly. If this does not occur, find and repair the leak. c)
21、 Preferably attach the pump to the workers belt, or place it in his pocket. d) Place the flexible tubing over the workers shoulder and so clip the filter holder to his clothing that the inlet port faces downwards, as near as possible to the breathing zone. e) Ensure that the equipment will not inter
22、fere with the workers movements in the area. f) Remove the plug from the filter holders inlet port and switch on the pump. Note and record the following: 1) Initial air flow rate; 2) ambient temperature; 3) atmospheric pressure; and 4) time of initiation. g) Maintain the flow rate within 5 % of the
23、nominal value throughout the sampling period. Since it is possible for a filter to become clogged by heavy particulate loading or by the presence of oil mists or other liquids in the air, check the flow indicator of the pump at least once per hour for any drop in flow rate. NOTE 1) This requirement
24、may be relaxed when a flow-stabilized pump is used. 2) Terminate sampling at any evidence of a problem. Do not re-adjust the flow rate. h) Sample for a total duration of 6-8 h, if feasible, to obtain a time-weighted average concentration. Where necessary, sequential air samples may be taken as oppos
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- SANS61642009DETERMINATIONOFLEADINORGANICANDTETRAALKYLINWORKPLACEAIRBYATOMICABSORPTIONSPECTROPHOTOMETRY

链接地址:http://www.mydoc123.com/p-1031632.html