ISA 37 10-1982 Specifications and Tests for Piezoelectric Pressure and Sound Pressure Transducers《压电式压力传感器和声压传感器的规范和试验 原标准号ISA-S37 10-1982 (R1995)》.pdf
《ISA 37 10-1982 Specifications and Tests for Piezoelectric Pressure and Sound Pressure Transducers《压电式压力传感器和声压传感器的规范和试验 原标准号ISA-S37 10-1982 (R1995)》.pdf》由会员分享,可在线阅读,更多相关《ISA 37 10-1982 Specifications and Tests for Piezoelectric Pressure and Sound Pressure Transducers《压电式压力传感器和声压传感器的规范和试验 原标准号ISA-S37 10-1982 (R1995)》.pdf(36页珍藏版)》请在麦多课文档分享上搜索。
1、Specifications and Tests forPiezoelectric Pressure and Sound Pressure TransducersReaffirmed 29 September 1995ISA37.101982 (R1995)Formerly ISAS37.101982 (R1995)STANDARDISA The Instrumentation,Systems, andAutomation Society TMCopyright 1982 by ISA. All rights reserved. Printed in the United States of
2、America. No part ofthis publication may be reproduced, stored in a retrieval system, or transmitted in any form or byany means (electronic, mechanical, photocopying, recording, or otherwise), without the priorwritten permission of the publisher.ISA67 Alexander DriveP.O. Box 12277Research Triangle Pa
3、rk, North Carolina 27709ISA-37.10-1982 (R1995), Specifications and Tests for Piezoelectric Pressure and Sound-Pressure TransducersISBN 0-87664-382-9ISA-S37.10-1982 (R1995) 3PrefaceThis preface, as well as all footnotes and annexes, is included for informational purposes and is not part of ISA-37.10-
4、1982 (R1995).This standard has been prepared as a part of the service of ISA, the international society for measurement and control, toward a goal of uniformity in the field of instrumentation. To be of real value, this document should not be static but should be subject to periodic review. Toward t
5、his end, the Society welcomes all comments and criticisms and asks that they be addressed to the Secretary, Standards and Practices Board; ISA; 67 Alexander Drive; P.O. Box 12277; Research Triangle Park, NC 27709; Telephone: (919) 549-8411; Fax: (919) 549-8288; E-mail: standardsisa.org.The ISA Stand
6、ards and Practices Department is aware of the growing need for attention to the metric system of units in general, and the International System of Units (SI) in particular, in the preparation of instrumentation standards, recommended practices, and technical reports. The Department is further aware
7、of the benefits to USA users of ISA Standards of incorporating suitable references to the SI (and the metric system) in their business and professional dealings with other countries. Towards this end, this Department will endeavor to introduce SI and acceptable metric units in all new and revised st
8、andards to the greatest extent possible. The Metric Practice Guide, which has been published by the Institute of Electrical and Electronics Engineers as ANSI/IEEE Std. 268-1992, and future revisions, will be the reference guide for definitions, symbols, abbreviations,and conversion factors.It is the
9、 policy of ISA to encourage and welcome the participation of all concerned individuals and interests in the development of ISA standards, recommended practices, and technical reports. Participation in the ISA standards-making process by an individual in no way constitutes endorsement by the employer
10、 of that individual, of ISA, or of any of the standards which ISA develops.This Standard was originally prepared by the SP37.10 Committee which operated under the guidance of SP37.The following individuals served on the original SP37.10 Subcommittee:NAME COMPANYN. Keast, Chairman Bolt, Beranek and N
11、ewman, IncorporatedG. T. Cozad McDonnell Douglas CorporationL. Horn National Bureau of StandardsR. W. Lally PCB PiezotronicsJ. Rhodes Endevco Corporation4 ISA-S37.10-1982 (R1995)The following individuals served on ISA Committee SP37, who reaffirmed ISA-S37.10 in 1995:NAME COMPANYE. Icayan, Chairman
12、Westinghouse Hanford Co.J. Weiss Electric Power Research Inst.P. Bliss, Deceased ConsultantM. Brigham The Supply SystemD. Hayes LA Dept. Water i.e., Temperature: 25 10C (77 18F); Relative Humidity: not to exceed 90 percent; Barometric Pressure: 730 70 millimeters Hg (29 3 in. Hg).Terminology used he
13、re is defined either in ISA-S37.1 or in 4.3 of this document. An asterisk appears beside the paragraph number of those terms defined in S37.1.It is important that all transducer performance characteristics be listed independent of the characteristics of any amplifier and/or non-integral cables suppl
14、ied with the transducer. (Performance characteristics may also be supplied including the effects of amplifiers and cables, if so identified as supplemental information.) In those cases where such characteristics cannot be stated independent of amplifier properties, the pertinent amplifier and cable
15、type, part number, and properties shall also be specified.All performance characteristics are applicable under the conditions specified for Load Impedance. Note that this practice is used in lieu of specifying “open-circuit“ output characteristics (an earlier practice, which did not permit verificat
16、ion of such characteristics since all ancillary equipment has a finite value of input impedance).NOTE In the following, separate statements are given for pressure transducers and for piezoelectric sound-pressure transducers in those cases where current terminologies differ for the two applications.4
17、.2.1 Required performance characteristics4.2.1.1 Range*a) Pressure transducers. The range, usually expressed as “ _Pa (psi)“ or “0 to _Pa (psi.)“b) Piezoelectric Sound-Pressure Transducers. The range is usually expressed as “_dB sound pressure level (SPL) to _dB SPL re 20 Pa.“*Defined in ISA-S37.114
18、 ISA-S37.10-1982 (R1995)4.2.1.2 Sensitivity, charge or voltage*a) Pressure Transducers. The voltage sensitivity is expressed as “_mV per Pa (psi) _%,“ or as “_ _mV per Pa (psi).“ Equivalently, charge sensitivity may be expressed as “_ picocoulombs per Pa (psi) _%,“ or as “_ _ picocoulombs per Pa (ps
19、i).“In any case, it is assumed that the pressure and electrical parameters are both reported as peak, average, or rms values unless stated otherwise.b) Piezoelectric sound-pressure transducers. The voltage sensitivity level is expressed in dB as 20 times the logarithm to the base 10 of the ratio of
20、the sensitivity to the reference sensitivity. That is, sensitivity level S, re 1V/Pa, is S = LV Lp= 20 log10V/pwhere Lvis the output voltage, re 1V, produced by applied sound pressure level Lp,re 20Pa. Appropriate tolerances should be shown for a specified nominal value.Alternatively, the charge sen
21、sitivity level may be expressed as “_ _dBre 1 picocoulomb per Pascal.“4.2.1.3 Frequency response*a) Pressure transducers. This is expressed as “within _% of the sensitivity at _ Hz from _ to _ Hz.“ The method for determining this frequency response should be described.b) Piezoelectric sound-pressure
22、 transducers. Frequency response shall be specified as “_ type frequency response) within 3 dB (or, alternatively, within 1 dB) from _ Hz to _ Hz.“ The quantity entered into the first blank shall be one of the following (as defined in 4.3): calculated frequency response*, pressure response, free-fie
23、ld grazing incidence response, free-field normal incidence response, or free-field random incidence response.Frequency response shall be referred to a frequency within the specified frequencyrange of the transducer and to a specific fluid. The methods of mounting the trans-ducer and applying the tes
24、t fluid should both be specified.4.2.1.4 Linearity*a) Pressure transducers. Linearity is normally expressed as “_ linearity within _% of full (or a specified partial) scale output.“ The type of linearity to be entered in the first blank above shall be one of the straight line types defined in ISA-S3
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便