ASTM F982-1986(2002) Standard Specification for Disclosure of Characteristics of Surgically Implanted Clamps for Carotid Occlusion《静动脉连接用外科植入夹的特性表示》.pdf
《ASTM F982-1986(2002) Standard Specification for Disclosure of Characteristics of Surgically Implanted Clamps for Carotid Occlusion《静动脉连接用外科植入夹的特性表示》.pdf》由会员分享,可在线阅读,更多相关《ASTM F982-1986(2002) Standard Specification for Disclosure of Characteristics of Surgically Implanted Clamps for Carotid Occlusion《静动脉连接用外科植入夹的特性表示》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 982 86 (Reapproved 2002)Standard Specification forDisclosure of Characteristics of Surgically ImplantedClamps for Carotid Occlusion1This standard is issued under the fixed designation F 982; the number immediately following the designation indicates the year oforiginal adoption or, in
2、 the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers requirements for the disclosureof specific characteristi
3、cs of screw-type adjustable clamps thatare designed for the gradual permanent occlusion of carotidarteries. These devices consist of an implantable portion and anexternally projecting removable screwdriver (see Fig. 1).1.2 The following precautionary caveat pertains only to thetest method portion, S
4、ection 5, of this specification: Thisstandard does not purport to address all of the safety concerns,if any, associated with its use. It is the responsibility of the userof this standard to establish appropriate safety and healthpractices and determine the applicability of regulatory limita-tions pr
5、ior to use.2. Referenced Documents2.1 The designations and titles of the applicable documentsfor this specification are listed in Annex A1 in the followinggroups:2.1.1 Materials,2.1.2 Finishing,2.1.3 Biocompatibility,2.1.4 Handling, and2.1.5 Analysis.3. Terminology3.1 Descriptions of Terms Specific
6、to This Standard (seeFig. 1):3.1.1 access plateportion of the device that closes theframe.3.1.2 capcovering device to seal the lumen of the stemwhen the screwdriver is not in place.3.1.3 collarthreaded portion of the frame that acts as aguide and counter torque surface for the pressure plate screw.3
7、.1.4 frameencircling portion of the device, usuallyU-shaped.3.1.5 guidecylinder within the stem to provide countertorque and guidance for the screwdriver.3.1.6 hingemeans of attaching the access plate to theframe.3.1.7 pressure platemovable compressing plate.3.1.8 pressure plate screwthreaded shaft
8、that advancesthe pressure plate.3.1.9 screwdriverdevice used to provide torque to thepressure plate screw. The screwdriver should have permanentlymarked scale indicating advance ratio by millimetres.3.1.10 set screwscrew that secures the access plate to theframe.3.1.11 stemcylinder that is used to h
9、old the frame and toprovide counter torque for the screwdriver.4. General Requirements4.1 This section contains requirements for disclosure ofinformation on screw-type adjustable clamps.4.2 Performance Disclosure:4.2.1 Materials:4.2.1.1 The manufacturer shall disclose the generic namesof the materia
10、ls used in the manufacture of the clamp.1This specification is under the jurisdiction of ASTM Committee F04 onMedical and Surgical Materials and Devices and is the direct responsibility ofSubcommittee F04.31 on Neurosurgical Standards.Current edition approved March 27, 1986. Published May 1986.FIG.
11、1 Screw-Type Adjustable Clamp Components1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.Whenever available, ASTM material specification nomencla-ture shall be used (Annex A1). If multiple components areused, the names of each compone
12、nt shall be disclosed.4.2.1.2 The metals and alloys or other materials used inclamps that conform to this specification should be fabricatedof approved materials in accordance with the ASTM specifi-cations listed in A1.2.1.4.2.2 FinishingSurface cleanliness and characteristicsshould meet the require
13、ments of the ASTM specifications listedin A1.2.2. There should be no debris visible at 203 and noimperfections visible to the naked eye.4.2.3 BiocompatibilityClamps should be biocompatiblewith the tissue in which they are intended to be implanted.Metal components shall meet ASTM biologic compatibili
14、tyrequirements or equivalents listed in A1.2.3. Nonporous poly-meric materials should conform to the ASTM requirements orequivalents listed in the Annex.4.2.4 HandlingHandling procedures should be similar tothose suggested by several ASTM standards listed in A1.2.4.4.2.5 AnalysisAnalysis of clips re
15、moved for any reasonshould resemble that specified for removal of orthopedicimplants (see A1.2.5).4.2.6 Reporting of FailuresAll failures should be reportedboth to the manufacturer and to the Food and Drug Adminis-tration (FDA).4.2.7 Advance RatioThe manufacturer shall disclose thedistance (millimet
16、res) advanced by the pressure plate for eachfull revolution of the screwdriver (see also 5.1).4.2.8 Pressure Plate Induced Laceration of VesselThemanufacturer shall disclose the torque at which the pressureplate will cause vessel laceration (see also 5.2).4.2.9 Slip Resistance of the ClampThe manufa
17、cturer shalldisclose whether the set screw will unwind and the pressureplate will retreat in the face of pulsatile pressure of 150/80 at80 cpm applied to the pressure plate when it is 2 mm fromclosure and when it is at the closed position (see also 5.3).5. Test Methods5.1 Advance RatioThis measureme
18、nt must be accurate to60.35 mm.5.2 Vessel DamageImplant the clamp aseptically around adog carotid artery and close using a torque wrench. Implantseveral animals, each having their clamp tightened to adifferent torque. Sacrifice the animals two weeks later andexamine microscopically as well as histol
19、ogically to determineif there is laceration. Clamps can be tightened to a torque justbelow that which will cause laceration.5.3 Slip Resistance Perform the study in vitro using dogcarotid arteries (average diameter 5 mm) or tubes of similardistensibility and a pulse duplicator system. Set the pulsed
20、uplicator to create a pulsatile cycle of pressure similar to thephysiologic systolic-diastolic pattern (150/80 at 80 cpm).During the experiment, keep the artery in a normal saline bathand connect it to the pulse duplicator system. Close the clampto a gap of 2 mm. Turn on the pulse duplicator and mea
21、sure theposition of the pressure plate relative to the basis of the frameevery 24 h for 72 h to determine if there has been any retreatof the pressure plate. Perform a second test keeping the pulseduplicator functioning at the same setting but with the clampclosed down to occlude the artery using th
22、e torque determinedsafe as detailed in 5.2. Again, measure the position of thepressure plate every 24 h for 72 h. A backoff of 0.2 mm (90turn) will be the maximum permitted. Measurement must beaccurate to 60.35 mm.6. Labeling Requirements6.1 All labeling must be consistent with applicable FederalReg
23、ulations. In addition, the labeling for carotid occlusiveclamps within the scope of this specification should complywith the following requirements:6.1.1 Package LabelThe following information shall beavailable with the unit package:6.1.1.1 Manufacturers name,6.1.1.2 Trade name,6.1.1.3 Catalog numbe
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF98219862002STANDARDSPECIFICATIONFORDISCLOSUREOFCHARACTERISTICSOFSURGICALLYIMPLANTEDCLAMPSFORCAROTIDOCCLUSION

链接地址:http://www.mydoc123.com/p-537662.html