ASTM F746-2004(2014) Standard Test Method for Pitting or Crevice Corrosion of Metallic Surgical Implant Materials《金属外科植入材料的凹痕或裂隙腐蚀的标准试验方法》.pdf
《ASTM F746-2004(2014) Standard Test Method for Pitting or Crevice Corrosion of Metallic Surgical Implant Materials《金属外科植入材料的凹痕或裂隙腐蚀的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F746-2004(2014) Standard Test Method for Pitting or Crevice Corrosion of Metallic Surgical Implant Materials《金属外科植入材料的凹痕或裂隙腐蚀的标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F746 04 (Reapproved 2014)Standard Test Method forPitting or Crevice Corrosion of Metallic Surgical ImplantMaterials1This standard is issued under the fixed designation F746; the number immediately following the designation indicates the year of originaladoption or, in the case of revisi
2、on, the year of last revision.Anumber in parentheses indicates the year of last reapproval.Asuperscriptepsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of resistanceto either pitting or crevice corrosion of metals an
3、d alloys fromwhich surgical implants will be produced. It is a modifiedversion of an established test2and is used as a screening test torank surgical implant alloys in order of their resistance tolocalized corrosion.1.2 This test method applies only to passive metals andalloys. Nonpassive alloys (ot
4、her than noble alloys) are suscep-tible to general corrosion and are not normally suitable forimplant use.1.3 This test method is intended for use as a laboratoryscreening test for metals and alloys which undergo pitting orcrevice corrosion, or both.1.4 The values stated in either SI units or inch-p
5、ound unitsare to be regarded separately as standard. The values stated ineach system may not be exact equivalents; therefore, eachsystem shall be used independently of the other. Combiningvalues from the two systems may result in non-conformancewith the standard.1.5 This standard does not purport to
6、 address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:3D1193 Spe
7、cification for Reagent WaterF86 Practice for Surface Preparation and Marking of Metal-lic Surgical ImplantsF2129 Test Method for Conducting Cyclic PotentiodynamicPolarization Measurements to Determine the CorrosionSusceptibility of Small Implant DevicesG3 Practice for Conventions Applicable to Elect
8、rochemicalMeasurements in Corrosion TestingG5 Reference Test Method for Making PotentiodynamicAnodic Polarization MeasurementsG15 Terminology Relating to Corrosion and Corrosion Test-ing (Withdrawn 2010)43. Summary of Test Method3.1 Acylindrical specimen fitted with an inert tapered collaris immerse
9、d in a phosphate buffered saline electrolyte at 37Cfor1htoestablish a corrosion potential. Pitting (or crevicecorrosion) is then stimulated by potentiostatically polarizingthe specimen to a potential much more noble than the corrosionpotential. Stimulation of pitting (or crevice corrosion) will bema
10、rked by a large and generally increasing polarizing current.3.2 Immediately after the stimulation step, the potential isdecreased as rapidly as possible to one of several preselectedpotentials at, or more noble than, the corrosion potential. If thealloy is susceptible to pitting (or crevice corrosio
11、n) at thepreselected potential, the polarizing current will remain atrelatively high values and will fluctuate or increase with time.A post-test examination of the metal specimen establisheswhether localized corrosion has occurred by pitting of theexposed surface or by preferential attack at the cre
12、vice formedby the tapered collar, or both.3.3 If the pit (or crevice) surface repassivates at the pre-selected potential and localized corrosion is halted, the polar-izing current will drop to values typical for passive surfacesand the current will decrease continuously. The parameter ofinterest, th
13、e critical potential for pitting (or crevice corrosion),is defined as the highest (most noble) pre-selected potential atwhich pit (or crevice) surfaces repassivate after the stimulationstep.4. Significance and Use4.1 This test method is designed solely for determiningcomparative laboratory indices o
14、f performance. The results1This test method is under the jurisdiction ofASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved Oct. 1, 2014. Published November 2014. Originallyapproved i
15、n 1981. Last previous edition approved in 2009 as F746 04(2009)1.DOI: 10.1520/F0746-04R14.2Syrett, B. C., Corrosion, Vol 33, 1977, p. 221.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume
16、information, refer to the standards Document Summary page onthe ASTM website.4The last approved version of this historical standard is referenced onwww.astm.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1may be used for ranking a
17、lloys in order of increasing resistanceto pitting and crevice corrosion under the specific conditions ofthis method. It should be noted that the method is intentionallydesigned to reach conditions that are sufficiently severe tocause breakdown of at least one alloy (Type 316 L stainlesssteel) curren
18、tly considered acceptable for surgical implant use,and that those alloys which suffer pitting or crevice corrosionduring the more severe portions of the test do not necessarilysuffer localized corrosion when placed within the human bodyas a surgical implant.5. Apparatus5.1 The following required equ
19、ipment is described in Ref-erence Test Method G5:5.1.1 Standard Polarization Cell, of 1000 cm3.5.1.2 Electrode Holders, for auxiliary and working elec-trodes.5.1.3 Potentiostat, calibrated in accordance with ReferenceTest Method G5.5.1.4 Potential-Measuring Instrument.5.1.5 Current-Measuring Instrum
20、ent.5.1.6 Anodic Polarization Circuit.5.1.7 Platinum Auxiliary Electrodes.5.1.8 Saturated Calomel Electrode (SCE).5.1.9 Salt Bridge Probe.5.2 A cylindrical working electrode is fabricated from thetest material by machining, grinding, and suggested finalpolishing with 600-grit metallographic paper. I
21、t is suggestedthat the part of the cylindrical specimen that is exposed to thetest solution have a length of 20.00 6 1.00 mm 0.787 6 0.039in. and a diameter of 6.35 6 0.03 mm 0.250 6 0.001 in. (seeFig. 1).5.3 A crevice is created by fitting the cylindrical specimenwith a tapered collar, machined fro
22、m commercial puritypolytetrafluoroethylene (PTFE). The collar should have anouter diameter of 12.70 6 0.05 mm 0.500 6 0.002 in. and athickness of 3.18 6 0.20 mm 0.125 6 0.008 in. The insidediameter of the tapered collar should range from 0.38 mm0.015 in. smaller than the diameter of the specimen to
23、0.38mm 0.015 in. larger. To be consistent with the dimensionssuggested in 5.2, the inside diameter should taper from 5.97 60.05 mm 0.235 6 0.002 in. to 6.73 6 0.05 mm 0.265 60.002 in. See Fig. 1 for drawing of the tapered collar. Therelatively fine tolerances are needed to ensure a reproducible fita
24、nd crevice.5.4 In Reference Test Method G5, the method of specimenattachment is to drill and tap the specimen to receive a threadedstainless steel connection rod. A4-40 thread is used, typically.However, because many surgical implant alloys are not easilydrilled, external threads may also be machine
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF74620042014STANDARDTESTMETHODFORPITTINGORCREVICECORROSIONOFMETALLICSURGICALIMPLANTMATERIALS 金属 外科

链接地址:http://www.mydoc123.com/p-537206.html