ASTM E925-2009(2014) 8272 Standard Practice for Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Bandwidth does not Exceed 2 nm《光谱缝宽不超过2 nm的可见紫外线.pdf
《ASTM E925-2009(2014) 8272 Standard Practice for Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Bandwidth does not Exceed 2 nm《光谱缝宽不超过2 nm的可见紫外线.pdf》由会员分享,可在线阅读,更多相关《ASTM E925-2009(2014) 8272 Standard Practice for Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Bandwidth does not Exceed 2 nm《光谱缝宽不超过2 nm的可见紫外线.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E925 09 (Reapproved 2014)Standard Practice forMonitoring the Calibration of Ultraviolet-VisibleSpectrophotometers whose Spectral Bandwidth does notExceed 2 nm1This standard is issued under the fixed designation E925; the number immediately following the designation indicates the year of
2、original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONIn the application of spectrophotometric methods of analysi
3、s it is the responsibility of the analystto verify and validate that the instrument is functioning properly and is capable of providingacceptable analytical results. It is preferable that the verification of instrument performance beaccomplished through the use of reference materials whose propertie
4、s have been accuratelydetermined. Such materials are readily available, and their use in the tests and measurements describedin this practice is satisfactory for evaluating the performance of spectrophotometers whose spectralbandwidth does not exceed the value for which the intrinsic or certified pr
5、operties are valid. Acompromise maximum permissible spectral bandwidth of 2 nm is recommended for the referencematerials and error tolerances recommended here.This practice covers some of the essential instrumental parameters that should be evaluated toensure the acceptability of the analytical data
6、 routinely obtained on the instrument. These parametersinclude the accuracy of the wavelength and absorbance scales and stray radiant power levels.The accuracy of the wavelength scale in both the ultraviolet (UV) and visible regions is determinedusing the sharp absorption bands of a holmium oxide gl
7、ass or solution filter. The absorbance scaleaccuracy in the UV region (235 to 350 nm) is determined using acidic solutions of potassiumdichromate. In the visible region (440 to 635 nm) the absorbance accuracy is determined usingindividually certified neutral density glass filters. The use of these r
8、eference materials provides a validand relatively simple means to test the errors in the wavelength and absorbance scales of small spectralbandwidth spectrophotometers in the spectral ranges indicated. A simplified version of the opaquefilter method is provided as a test for excessive stray radiant
9、energy.1. Scope1.1 This practice covers the parameters of spectrophotomet-ric performance that are critical for testing the adequacy ofinstrumentation for most routine tests and methods2within thewavelength range of 200 to 700 nm and the absorbance rangeof 0 to 2.The recommended tests provide a meas
10、urement of theimportant parameters controlling results in spectrophotometricmethods, but it is specifically not to be inferred that all factorsin instrument performance are measured.1.2 This practice may be used as a significant test of theperformance of instruments for which the spectral bandwidthd
11、oes not exceed 2 nm and for which the manufacturersspecifications for wavelength and absorbance accuracy do notexceed the performance tolerances employed here. This prac-tice employs an illustrative tolerance of 61 % relative for theerror of the absorbance scale over the range of 0.2 to 2.0, andof 6
12、1.0 nm for the error of the wavelength scale. A suggestedmaximum stray radiant power ratio of410-4yields 99.99 %), where value assignment is by self assertion (Note 1).3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual
13、 Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.4Available from National Technical Information Service (NTIS), 5301 ShawneeRd., Alexandria, VA 22312, http:/www.ntis.gov.5Available from International Organization for Standardization (ISO), 1,
14、 ch. dela Voie-Creuse, CP 56, CH-1211 Geneva 20, Switzerland, http:/www.iso.org.6The boldface numbers in parentheses refer to a list of references at the end ofthis standard.E925 09 (2014)2Apurchaser should require certification by the supplier that thewavelengths of the absorption bands are within
15、0.2-nm of thevalues given in Ref. (2), and reported below. The appropriatesolution standard is 4 % (mass fraction) holmium oxide in10 % (volume fraction) perchloric acid, contained in a 10-mmpath length cuvette. For this material, the transmittance minimaof 18 absorption bands have been certified by
16、 a multi-laboratory inter-comparison, at the highest level, allowing thepeak value assignments as an intrinsic wavelength standard (3).Absorbance maxima or transmittance minima must be locatedwithin 61 nm of the wavelengths given below:Glass FilterADilute Acidic SolutionB241.5 nmC241.1 nm. . . 249.9
17、 nm279.3 nm 278.1 nm287.6 nm 287.2 nm333.8 nm 333.5 nm. . . 345.4 nm360.8 nm 361.3 nm385.8 nm 385.6 nm418.5 nm 416.3 nm453.4 nm . . .D459.9 nm 467.8 nm. . . 485.3 nm536.4 nm 536.6 nm637.5 nm 640.5 nmAWavelengths taken from Ref. (2) for Corning Glass Works Code 3130 glass,superceded by Corning Glass
18、Works Code 3131 glass and Kopp Glass Code 3131glass, for which the wavelengths are also valid.BWavelengths rounded to 0.1 nm for a 1-nm spectral bandwidth taken from Ref.(3).CMay not be usable, depending on the base glass of the filter.DPeak omitted because it resolves into a doublet at spectral ban
19、dwidth valuesless than 1 nm.NOTE 1Self assertion may take the form of value assignment andcertification in many forms. Some specific examples are:(1) By a national metrology institute (NMI),(2) By an ISO 17025 and ISO Guide 34 accredited Reference Mate-rial producer, and(3) By a laboratory claiming
20、traceability to an NMI.In all cases, the user should be satisfied that the quality of the valueassignment data meets the laboratory requirements.7.1.1 If the observed absorption bands of the holmium oxideglass or solution deviate by more than 61 nm from the valuesstated, then corrective service must
21、 by performed on theinstrument by qualified personnel. If the user performs thisservice, the manufacturers recommended procedure should befollowed carefully.7.1.2 The wavelength accuracy is dependent on the spectralbandwidth and thus on the physical bandwidth. Spectralbandwidths may be determined fr
22、om the manufacturers speci-fications.7.1.3 Computer based peak location algorithms that may beused to assign absorbance maxima or transmittance minima arediscussed in 7.6 of Guide E1866. It should be noted that peakasymmetries in the holmium oxide reference materials are suchthat digital filter widt
23、hs should be smaller than the full-width-half-maximum recommendation of that guide.7.1.4 In the absence of drift or slippage in the wavelengthdrive train, repeatability of the band positions should be on theorder of 60.1 nm for a given instrument, especially with theuse of a computer based peak loca
24、tion algorithm.7.2 Procedure:7.2.1 Examine the holmium oxide reference material andremove any surface contamination using a soft brush orlint-free cloth. Measure the temperature of the sample com-partment by placing an appropriate sensor into the cell com-partment of a stabilized instrument and repl
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME925200920148272STANDARDPRACTICEFORMONITORINGTHECALIBRATIONOFULTRAVIOLETVISIBLESPECTROPHOTOMETERSWHOSESPECTRALBANDWIDTHDOESNOTEXCEED2NM

链接地址:http://www.mydoc123.com/p-533787.html