ASTM E2193-2016 Standard Test Method for Ultraviolet Transmittance of Monoethylene Glycol (using Ultraviolet Spectrophotometry)《单乙二醇的紫外光透射率的标准试验方法 (紫外光谱法)》.pdf
《ASTM E2193-2016 Standard Test Method for Ultraviolet Transmittance of Monoethylene Glycol (using Ultraviolet Spectrophotometry)《单乙二醇的紫外光透射率的标准试验方法 (紫外光谱法)》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2193-2016 Standard Test Method for Ultraviolet Transmittance of Monoethylene Glycol (using Ultraviolet Spectrophotometry)《单乙二醇的紫外光透射率的标准试验方法 (紫外光谱法)》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E2193 08E2193 16Standard Test Method forUltraviolet Transmittance of Monoethylene Glycol(Ultraviolet Spectrophotometric Method)(using UltravioletSpectrophotometry)1This standard is issued under the fixed designation E2193; the number immediately following the designation indicates the y
2、ear oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers a procedure for the determin
3、ation of the transmittance of monoethylene glycol (1,2-ethanediol;MEG) at wavelengths in the region 220 to 350 nm. The results provide a measure of the purity of the sample with respect toultraviolet absorbing compounds.1.2 The values stated in SI units are to be regarded as standard. No other units
4、 of measurement are included in this standard.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regula
5、torylimitations prior to use. For specific hazard statements, see Section 7.1.4 Review the current Material Safety Data Sheets (MSDS)(SDS) for detailed information concerning toxicity, first aidprocedures, and safety precautions.precautions for all materials used in this test method.2. Referenced Do
6、cuments2.1 ASTM Standards:2D1193 Specification for Reagent WaterD6299 Practice for Applying Statistical Quality Assurance and Control Charting Techniques to Evaluate Analytical Measure-ment System PerformanceE131 Terminology Relating to Molecular SpectroscopyE169 Practices for General Techniques of
7、Ultraviolet-Visible Quantitative AnalysisE180 Practice for Determining the Precision of ASTM Methods for Analysis and Testing of Industrial and Specialty Chemicals(Withdrawn 2009)3E275 Practice for Describing and Measuring Performance of Ultraviolet and Visible Spectrophotometers2.2 Other Document:M
8、anufacturers Instruction Manual of Spectrophotometer3. Summary of Test Method3.1 The product is sampled in such a way as to avoid extraneous contamination and air contact. The absorbance of the samplecontained in a 50-mm or 10-mm cell is measured against water at a series of wavelengths and the tran
9、smittance over a pathlengthof 10 mm is calculated.3.2 This test method can be performed with two options as to sample preparation prior to UV measurement.3.2.1 Option A: Nitrogen sparging of the sample (see 4.2).3.2.2 Option B: No nitrogen sparging.4. Significance and Use4.1 Knowledge of the ultravi
10、olet transmittance of monoethylene glycol is required to establish whether the product meets therequirements of its quality specifications.1 This test method is under the jurisdiction of ASTM Committee E15 on Industrial and Specialty Chemicals and is the direct responsibility of Subcommittee E15.02
11、onProduct Standards.Current edition approved April 1, 2008Jan. 1, 2016. Published May 2008January 2016. Originally approved in 2002. Last previous edition approved in 20042008 asE2193 04.E2193 08. DOI: 10.1520/E2193-08.10.1520/E2193-16.2 For referencedASTM standards, visit theASTM website, www.astm.
12、org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3 The last approved version of this historical standard is referenced on www.astm.org.This document is not an ASTM standard a
13、nd is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only
14、 the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States14.2 Dissolved oxygen i
15、n organic solvents, such as MEG, forms complexes that shift the solvent absorption from the vacuumultraviolet range into the measurable UV range (near 190 to 250 nm). Monoethylene glycol has a UV absorption peak at 180 nm.For MEG-oxygen complexes, this peak is shifted to a longer wavelength, thus in
16、creasing the absorbability at 220 nm.4.2.1 However, this effect is not observed in water. There is no significant measurable effect due to dissolved oxygen in waterthat would require nitrogen sparging prior to using for collection of the reference spectrum.4.2.2 Nitrogen sparging and re-measurement
17、of suspect or borderline glycol samples at 220 nm can be used as a tool to ruleout or confirm the presence of UV affecting contaminants other than oxygen.5. Apparatus5.1 Ultraviolet Spectrophotometer, double beam, suitable for measurement at wavelengths in the region 200 to 400 nm, havinga spectral
18、bandwidth of 2.0 nm or less at 220 nm, wavelength accuracy 60.5 nm or less at 220 nm, wavelength repeatability 0.3nm or less at 220 nm and a photometric accuracy of 60.5 % T or less, in the transmittance region above 50 % T. Stray light shallbe less than 0.1 % at 220 nm.The instrument shall be provi
19、ded with matched fused silica quartz (fused silica) cells with pathlengthsof 50 6 0.1 mm or 10 mm 6 0.01 mm. Use of 50-mm cell should provide better precision.5.1.1 Optional Single Beam SpectrophotomerSpectrophotometerUse the same cell for measurement of the blank and thesample.5.2 Nitrogen Strippin
20、g Apparatus (Option A, 3.2.1), consisting of an oil-free pressure reducing valve to fit the nitrogen cylinder,a control valve, vinyl tubing and a disposable glass pipette to be inserted in a 25-mL volumetric flask. flask or bottle. Componentsshould be clean and free of ultraviolet contaminants.Avoid
21、 contacting the sample with any plastic material containing plasticizers.Plasticizers can leach out of the material and cause erroneous results. Replace the disposable pipette by a clean, new one after eachsample handling. (See Section 910.)5.3 Bottles, capacity at least 0.5L, 0.5 L, with lined, wel
22、l-fitting cap. Use a fresh bottle for each determination.5.4 Glassware:5.4.1 Volumetric Flask, Flask or Bottle, 25-mL.25 mL.6. Reagents and Materials6.1 Purity of ReagentsUnless otherwise indicated, it is intended that all reagents shall conform to the specifications of theCommittee on Analytical Re
23、agents of the American Chemical Society where such specifications are available.4 Other grades maybe used, provided it is first ascertained that the reagent is of sufficiently high purity to permit its use without lessening the accuracyof the determination.6.2 Holmium Oxide Wavelength Calibration Fi
24、lter, calibrated (if required; see 8.1.19.1.1).NOTE 1The standard reference material SRM 2034, available from the National Institute of Standards and Technology (NIST),5 is suitable.6.3 Standard Absorbance Filter, with certified absorbance values (if required; see 8.1.29.1.2).NOTE 2The standard refe
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME21932016STANDARDTESTMETHODFORULTRAVIOLETTRANSMITTANCEOFMONOETHYLENEGLYCOLUSINGULTRAVIOLETSPECTROPHOTOMETRY

链接地址:http://www.mydoc123.com/p-530519.html