《ASTM E1862-2014 Standard Practice for Measuring and Compensating for Reflected Temperature Using Infrared Imaging Radiometers《采用红外成像辐射仪测量和均衡反射温度的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1862-2014 Standard Practice for Measuring and Compensating for Reflected Temperature Using Infrared Imaging Radiometers《采用红外成像辐射仪测量和均衡反射温度的标准试验方法》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1862 97 (Reapproved 2010)E1862 14Standard Test Methods Practice forMeasuring and Compensating for Reflected TemperatureUsing Infrared Imaging Radiometers1This standard is issued under the fixed designation E1862; the number immediately following the designation indicates the year ofori
2、ginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 These test methods cover procedures for measuring and compensa
3、ting for reflected temperature when measuring the surfacetemperature of a specimen with an infrared imaging radiometer.1.2 These test methods may involve use of equipment and materials in the presence of heated or electrically energizedequipment, or both.1.3 This standard does not purport to address
4、 all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E1316 Terminology f
5、or Nondestructive Examinations3. Terminology3.1 Definitions:3.1.1 diffuse reflector, na surface that produces a diffuse image of a reflected source.3.1.2 infrared thermographer, nthe person using an infrared imaging radiometer.3.1.3 infrared reflector, na material with a reflectance as close as poss
6、ible to 1.00.3.1.4 reflected temperature, nthe temperature of the energy incident upon and reflected from the measurement surface of aspecimen.3.1.5 specular reflector, na surface that produces a direct image of a reflected source.3.2 See also Terminology E1316.4. Summary of Test Methods4.1 Two test
7、 methods are given for measuring the reflected temperature of a specimen, the Reflector Method and the DirectMethod.4.2 A test method is also given for compensating for the error produced by reflected temperature using the computer built intoan infrared imaging radiometer.5. Significance and Use5.1
8、The infrared energy that is reflected by a specimen can cause measurement errors for an infrared thermographer measuringits surface temperature. Two test methods are provided for measuring and compensating for this reflected temperature error source,the Reflector Method and the Direct Method.5.2 The
9、se test methods can be used in the field or laboratory using commonly available materials.1 These test methods are This practice is under the jurisdiction of ASTM Committee E07 on Nondestructive Testing and are the direct responsibility of SubcommitteeE07.10 on Specialized NDT Methods.Current editio
10、n approved May 1, 2010Dec. 1, 2014. Published March 2011December 2014. Originally approved in 1997. Last previous edition approved in 20022010 asE1862 - 97(2002)(2010).1. DOI:10.1520/E1862-97R10. DOI:10.1520/E1862-14.3 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM
11、 Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made
12、to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyri
13、ght ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15.3 These test methods can be used with any infrared radiometers that have the required computer capabilities.6. Interferences6.1 Reflector Method:6.1.1 This test method uses an infrared refle
14、ctor with an assumed reflectance of 1.00, which is an ideal property. Errors can beminimized by using a reflector having a reflectance as close as possible to 1.00.6.1.2 Specimens vary in that they can be diffuse or spectral reflectors, or both. Use of an infrared reflector with reflectancepropertie
15、s as close as possible to those of the specimen will reduce errors.6.2 Direct Method:6.2.1 The Direct Method usually does not account for the heat from the infrared thermographers body as a source of reflectedtemperature. If this heat source creates a significant error, use the Reflector Method.6.3
16、Reflected temperature errors produced by a point source, such as the sun or a lamp, are difficult to measure accurately. Theseerror sources can often be avoided by moving the infrared imaging radiometers position and angle relative to the specimen.6.4 The measured reflected temperature of a specimen
17、 may be specific to the waveband of the infrared imaging radiometer used.Therefore, the infrared imaging radiometers waveband should be noted with the measured value.6.5 The significance of the error contributed by reflected temperature can be estimated by shielding the specimen from variousangles a
18、nd observing any changes in the thermal image.6.6 The error caused by reflected temperature can be reduced by shielding the specimen from the source of the reflection.7. Apparatus7.1 Calibrated Infrared Imaging Radiometer, with a built-in computer with the capability to measure temperatures with the
19、computers emissivity control set to 1.00.7.2 Tripod, or device to support the infrared imaging radiometer.7.3 Infrared ReflectorThe reflector method also requires an infrared reflector made from a piece of metal whose reflectanceis as close as possible to 1.00. Examples are a crumpled and re-flatten
20、ed piece of aluminum foil placed shiny side up on a pieceof cardboard, or a flat piece of metal with diffuse or spectral reflection characteristics, or both, similar to those of the specimen.8. Procedure8.1 Reflector Method:8.1.1 Set the infrared imaging radiometers emissivity control to 1.00.8.1.2
21、Place the infrared imaging radiometer on the tripod or support device at the desired location and distance from thespecimen.8.1.3 Point the infrared imaging radiometer at the specimen and focus on the portion of the specimen where the reflectedtemperature is to be measured.8.1.4 Place the infrared r
22、eflector in front of, and parallel to, the specimen (see Fig. 1). Maintain a safe working distance fromany heated, electrically energized or otherwise potentially dangerous targets.8.1.5 Without moving the imager and using an appropriate measurement function (such as spot temperature, cross hairs or
23、isotherms), measure and record the apparent surface temperature of the reflector reported by the radiometers computer. This is thereflected temperature of this specimen when viewed from the position indicated in 8.1.2.8.1.6 Repeat 8.1.1 8.1.5 a minimum of three times and average the temperatures to
24、yield an average reflected temperature.8.2 Direct Method:8.2.1 Set the infrared imaging radiometers emissivity control to 1.00.8.2.2 Place the infrared imaging radiometer on the tripod or support device at the desired location and distance from thespecimen.8.2.3 Point the infrared imaging radiometer
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME18622014STANDARDPRACTICEFORMEASURINGANDCOMPENSATINGFORREFLECTEDTEMPERATUREUSINGINFRAREDIMAGINGRADIOMETERS

链接地址:http://www.mydoc123.com/p-529667.html