ASTM E1247-2003 Standard Practice for Detecting Fluorescence in Object-Color Specimens by Spectrophotometry 《用分光光度法探测物体彩色样品的荧光粉的标准实施规程》.pdf
《ASTM E1247-2003 Standard Practice for Detecting Fluorescence in Object-Color Specimens by Spectrophotometry 《用分光光度法探测物体彩色样品的荧光粉的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1247-2003 Standard Practice for Detecting Fluorescence in Object-Color Specimens by Spectrophotometry 《用分光光度法探测物体彩色样品的荧光粉的标准实施规程》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1247 03Standard Practice forDetecting Fluorescence in Object-Color Specimens bySpectrophotometry1This standard is issued under the fixed designation E 1247; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of
2、 last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice provides spectrophotometric methods fordetecting the presence of fluorescence in object-color speci-me
3、ns.NOTE 1Since the addition of fluorescing agents (colorants, whiteningagents, etc.) is often intentional by the manufacturer of a material,information on the presence or absence of fluorescent properties in aspecimen may often be obtained from the maker of the material.1.2 This practice requires th
4、e use of a spectrophotometerthat both irradiates the specimen over the wavelength rangefrom 340 to 700 nm and allows the spectral distribution ofillumination on the specimen to be altered as desired.1.3 Within the above limitations, this practice is general inscope rather than specific as to instrum
5、ent or material.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use
6、.2. Referenced Documents2.1 ASTM Standards:D 2244 Practice for Calculation of Color Tolerances andColor Differences from Intrumentally Measured ColorCoordinates2E 284 Terminology of Appearance2E 308 Practice for Computing the Colors of Objects byUsing the CIE System2E 313 Practice for Calculating Ye
7、llowness and WhitenessIndices from Instrumentally Measured Color Coordinates2E 991 Practice for Color Measurement of FluorescentSpecimens2E 1164 Practice for Obtaining Spectrophotometric Data forObject-Color Evaluation2E 1331 Test Method for Reflectance Factor and Color bySpectrophotometry Using Hem
8、ispherical Geometry2E 1348 Test Method for Transmittance and Color by Spec-trophotometry Using Hemispherical Geometry2E 1349 Test Method for Reflectance Factor and Color bySpectrophotometry Using Bidirectional Geometry2E 2152 Practice for Computing the Colors of FluorescentObjects from Bispectral Ph
9、otometric Data2E 2153 Practice for Obtaining Bispectral Photometric Datafor Evaluation of Fluorescent Color23. Terminology3.1 The definitions in Terminology E 284, Practices E 991,E 2152, and E 2153 are applicable to this practice.4. Significance and Use4.1 Several standards, including Practices E 9
10、91, E 1164,and Test Methods E 1331, E 1348 and E 1349, require eitherthe presence or absence of fluorescence exhibited by thespecimen for correct application. This practice provides spec-trophotometric procedures for identifying the presence offluorescence in materials.4.2 This practice is applicabl
11、e to all object-color specimens,whether opaque, translucent, or transparent, meeting the re-quirements for specimens in the appropriate standards listed in2.1. Translucent specimens should be measured by reflectance,with a standard non-fluorescent backing material, usually butnot necessarily black,
12、placed behind the specimen duringmeasurement.4.3 This practice requires the use of a spectrophotometer inwhich the spectral distribution of the illumination on thespecimen can be altered by the user in one of several ways. Themodification of the illumination can either be by the insertionof optical
13、filters between the illuminating source and thespecimen, without interfering with the detection of the radia-tion from the specimen, or by interchange of the illuminatingand detecting systems of the instrument or by scanning of boththe illuminating energy and detection output as in the two-monochrom
14、ator method.1This practice is under the jurisdiction of ASTM Committee E 12 on Color andAppearance and is the direct responsibility of Subcommittee E12.05 on Fluores-cence.Current edition approved Jan. 10, 2003. Published March 2003. Originallyapproved in 1988. Last previous edition approved in 2000
15、 as E 1247 92 (2000).2Annual Book of ASTM Standards, Vol 06.01.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.4.4 The confirmation of the presence of fluorescence ismade by the comparison of spectral curves, color difference, orsing
16、le parameter difference such as DY between the measure-ments.NOTE 2In editions of E 1247 - 92 and earlier, the test of fluorescencewas the two sets of spectral transmittances or radiance factor (reflectancefactors) differ by 1 % of full scale at the wavelength of greatest difference.4.5 Either bidir
17、ectional or hemispherical instrument geom-etry may be used in this practice. The instrument must becapable of providing either broadband (white light) irradiationon the specimen or monochromatic irradiation and monochro-matic detection.4.6 Either bidirectional or hemispherical instrument geom-etry m
18、ay be used in this practice. The instrument must becapable of providing either broadband (white light) irradiationon the specimen or monochromatic irradiation and monochro-matic detection.5. Instrumental Requirements5.1 This practice requires instrumentation meeting the fol-lowing requirements.5.1.1
19、 The instrument source shall provide sufficient irradia-tion energy at the sample port to excite fluorescent emission, ifpresent.5.1.2 The instrument must provide one of the followingillumination/viewing combinations:5.1.2.1 Monochromatic illumination and monochromaticviewing (that is, a two-monochr
20、omator spectrophotometersometimes called a bispectrometer or spectrofluorimeter).5.1.2.2 Polychromatic illumination and monochromaticviewing.5.1.2.3 Reversible illumination/viewing to allow both poly-chromatic illumination with monochromatic viewing andmonochromatic illumination with polychromatic v
21、iewing.5.1.3 The instrument and associated computer softwareshall allow the standardization/calibration of the instrumentusing user modified standardization/calibration values, whichis a requirement for using any of the filter methods describedin this practice.NOTE 3Repeatable and accurate applicati
22、on of this practice requiresspecialized instrumentation. Some commercial one-monochromator spec-trometers are limited in their ability to allow for the insertion of opticalfilters and re-standardization with the filter in place as required in thisprocedure.6. Procedures6.1 There are three general ty
23、pes of procedures to detect thepresence of fluorescence instrumentally. Each has its advan-tages and shortcomings depending on the wavelength andintensity of the fluorescent emission and the instrumentationavailable to the user.6.2 Two-Monochromator Method: This method requires acolorimetric measuri
24、ng instrument that is equipped with twoseparate monochromators: the first, the illumination monochro-mator, irradiates the specimen with monochromatic light andthe second, the viewing monochromator, analyzes the radiationleaving the specimen. A two-dimensional array of bispectralphotometric values i
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME12472003STANDARDPRACTICEFORDETECTINGFLUORESCENCEINOBJECTCOLORSPECIMENSBYSPECTROPHOTOMETRY 分光光度法

链接地址:http://www.mydoc123.com/p-528245.html