ASTM D7677-2016 3653 Standard Test Method for the Continuous Measurement of Dissolved Ozone in Low Conductivity Water《连续测量低电导率水中溶解臭氧的标准试验方法》.pdf
《ASTM D7677-2016 3653 Standard Test Method for the Continuous Measurement of Dissolved Ozone in Low Conductivity Water《连续测量低电导率水中溶解臭氧的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7677-2016 3653 Standard Test Method for the Continuous Measurement of Dissolved Ozone in Low Conductivity Water《连续测量低电导率水中溶解臭氧的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7677 16Standard Test Method for theContinuous Measurement of Dissolved Ozone in LowConductivity Water1This standard is issued under the fixed designation D7677; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year
2、 of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the on-line and in-line determi-nation of dissolved ozone (DO3) in low conductivity water
3、inthe range from 0.001 mg/L to 5.0 mg/L DO3and conductivity90, before pumps, the tops of horizontalpipes, vertical pipes that have liquid flowing downward,immediately after increases in pipe diameter, after a depres-surization or any turbulence creating point.11. Calibration11.1 Calibration of ZeroW
4、ith the sensor removed fromthe line and still reading DO3values, place it in a location inthe air known to be free of ozone gas. Dry the sensormembrane to remove all sample water which may containDO3.Allow the sensor to equilibrate with the air temperature to62C. If there is a large temperature diff
5、erence between the airand the process, for example, 5C; this may take over an hour.Perform a zero calibration according to the manufacturersinstructions. Equivalence of zero is considered to be 0.005D7677 163mg/L (ppm) in most commercial instruments. Some polaro-graphic sensors have an intrinsic zer
6、o and do not require a zerocalibration. Follow the sensor manufacturers recommenda-tions.11.2 Two methods are used for span calibration of polaro-graphic ozone sensors: in-line calibration using a colorimetricreference and air calibration.11.2.1 Colorimetric Calibration of SpanEither of thereagent s
7、ystems noted in 8.3 can be used as the reference forspan calibration. The IOA has approved the indigo method.3With the sensor inline and reading stable DO3values, collect asample of water representative of the water the sensor ismeasuring. This can be from a port near an inserted sensor oron the out
8、let of the sensor flow housing drain. In either case,the sample must flow gently, without spray or turbulence thatcould de-gas the sample. Sample the water using three aliquotsor ampoules, simultaneously noting the reading on the on-lineinstrument. Measure all three grab samples. Discard an outlieri
9、f it differs by 0.015 mg/L (0.02 mg/L for displays not readingto the third decimal place) from the mean of the other two. Ifno two of the three agree, redo all three. Calculate the averageand input the value as the on-line instrument span calibrationaccording to manufacturers instructions.11.2.2 Air
10、 Calibration of Span (suitable for instruments soequipped)Since oxygen and ozone are similar in their per-meation rates through the sensor membrane, the rate ofpermeation of a known oxygen quantity can be equated to anozone calibration value. Single point air calibration can be ameans to calibrate s
11、pan in a DO3instrument with this capabil-ity. Remove the sensor from the line and place it in a locationknown to be free of ozone gas. Be sure the outer surface of thesensor membrane is dry and free of sample water, which maycontain DO3. Allow the sensor to equilibrate with the airtemperature within
12、 62C. If there is a large temperaturedifference between the air and the sensor, for example, 5C,this may take over an hour. Enter the air calibration mode in theanalyzer electronics which temporarily changes the polariza-tion potential applied to the cathode and changes the responseto oxygen (O2) pa
13、rtial pressure. Determination of the O2partialpressure of air is determined from the barometric pressure andtemperature of the air. Allow at least 10 minutes for the sensorto stabilize and then perform the calibration function accordingto the manufacturers instructions. Return to the ozone mea-surem
14、ent mode of the instrument and allow the extendedperiod of time required for the sensor to repolarize and stabilizeon ozone measurement per manufacturers recommendations.11.3 Calibration VerificationSpan and zero verificationare normally required.11.3.1 Colorimetric Verification of SpanWith the sens
15、orinline and reading DO3values under stable conditions, collecta sample of water representative of the water the sensor ismeasuring, this can be from a spigot near an inserted sensor oron the outlet of the flow housing drain. The DO3value forverifying the span should be 70 % of the target processcon
16、trol limit DO3value. System DO3values for verificationmay be higher than the target control limit. Sample andmeasure the ozone concentration of the water the same as forcalibration in 11.1. Compare the average of the three values tothe reading of the instrument. If they differ by 0.015 mg/L(0.02 mg/
17、L if either the spectrophotometer or ozone instru-ment displays do not read to the third decimal place) then aspan calibration is required.11.3.2 Air Verification of Span (for instruments soequipped)With the sensor removed from the line and stillreading DO3values, place it in a location known to be
18、free ofozone gas. Dry the sensor membrane to remove sample waterwhich may contain DO3. Allow the sensor to equilibrate withthe air temperature within 62C. If there is a large temperaturedifference between the air and the process, for example, 5C,this may take over an hour. To check the air calibrati
19、on value,enter into the air calibration mode (which changes the cathodepolarization voltage to make it respond to O2). Allow at least10 minutes for equilibration and record the % last calibrationvalue. If % last calibration value is not available then record %ideal value or absolute air value. Compa
20、re to % ideal orabsolute air value when calibrated. If the two agree within thequality limits set (typically 65 to 15 %), the sensor span hasbeen verified. Restore the instrument to the ozone measure-ment mode and allow the manufacturers extended repolariza-tion time before ozone measurements are us
21、ed.11.3.3 Air Verification of ZeroWith the sensor removedfrom the line and still reading DO3values, place it in a locationin the air known to be free of ozone gas. Dry the sensormembrane to remove all sample water, which may containDO3. Allow the sensor to equilibrate with the air temperaturewithin
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD767720163653STANDARDTESTMETHODFORTHECONTINUOUSMEASUREMENTOFDISSOLVEDOZONEINLOWCONDUCTIVITYWATER

链接地址:http://www.mydoc123.com/p-526088.html