1、Designation: D4607 94 (Reapproved 2011)D4607 14Standard Test Method forDetermination of Iodine Number of Activated Carbon1This standard is issued under the fixed designation D4607; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, th
2、e year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of the relative activation level of unused or reactivated carbons
3、by adsorption ofiodine from aqueous solution. The amount of iodine absorbed (in milligrams) by 1 g of carbon using test conditions listed hereinis called the iodine number.1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.
4、3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use. Specific hazard st
5、atements are given in Section 7.2. Referenced Documents2.1 ASTM Standards:2C819 Test Method for Specific Surface Area of Carbon or Graphite (Withdrawn 1987)3D1193 Specification for Reagent WaterD2652 Terminology Relating to Activated CarbonD2867 Test Methods for Moisture in Activated CarbonD3860 Pra
6、ctice for Determination of Adsorptive Capacity of Activated Carbon by Aqueous Phase Isotherm TechniqueE11 Specification for Woven Wire Test Sieve Cloth and Test SievesE177 Practice for Use of the Terms Precision and Bias in ASTM Test MethodsE287 Specification for Laboratory Glass Graduated BuretsE28
7、8 Specification for Laboratory Glass Volumetric FlasksE300 Practice for Sampling Industrial Chemicals2.2 NIST Publication:Circular 602Testing of Glass Volumetric Apparatus43. Summary of Test Method3.1 This test method is based upon a three-point adsorption isotherm (see Practices D3860).Astandard io
8、dine solution is treatedwith three different weights of activated carbon under specified conditions. The carbon treated solutions are filtered to separate thecarbon from the treated iodine solution (filtrate). Iodine remaining in the filtrate is measured by titration. The amount of iodineremoved per
9、 gram of carbon is determined for each carbon dosage and the resulting data used to plot an adsorption isotherm. Theamount of iodine adsorbed (in milligrams) per gram of carbon at a residual iodine concentration of 0.02 N is reported as the iodinenumber.3.2 Iodine concentration in the standard solut
10、ion affects the capacity of an activated carbon for iodine adsorption. Therefore,the normality of the standard iodine solution must be maintained at a constant value (0.100 6 0.001 N ) for all iodine numbermeasurements.1 This test method is under the jurisdiction of ASTM Committee D28 on Activated C
11、arbon and is the direct responsibility of Subcommittee D28.02 on Liquid PhaseEvaluation.Current edition approved March 1, 2011March 1, 2014. Published May 2011May 2014. Originally approved in 1986. Last previous edition approved in 2006 as D4607 94(2006). (Reapproved 2011). DOI: 10.1520/D4607-94R11.
12、10.1520/D4607-14.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3 The last approved version of this historic
13、al standard is referenced on www.astm.org.4 Available from National Institute of Standards and Technology (NIST), 100 Bureau Dr., Stop 1070, Gaithersburg, MD 20899-1070, http:/www.nist.gov.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indicatio
14、n of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be consid
15、ered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.3 The apparatus required consists of various laboratory glassware used to prepare solutions and contact carbon with thestandard iodine solution. Filtration a
16、nd titration equipment are also required.4. Significance and Use4.1 The iodine number is a relative indicator of porosity in an activated carbon. It does not necessarily provide a measure ofthe carbons ability to absorb other species. Iodine number may be used as an approximation of surface area for
17、 some types ofactivated carbons (see Test Method C819). However, it must be realized that any relationship between surface area and iodinenumber cannot be generalized. It varies with changes in carbon raw material, processing conditions, and pore volume distribution(see Definitions D2652).4.2 The pr
18、esence of adsorbed volatiles, sulfur; and water extractables may affect the measured iodine number of an activatedcarbon.5. ApparatusNOTE 1All volumetric measuring equipment should meet or exceed the requirements of NIST Circular 602. Volumetric glassware meeting thesespecifications is generally des
19、ignated as “Class A”. See also Specifications E287 and E288.5.1 Analytical Balance, accuracy 60.0001 g.5.2 Buret, 10-mL capacity or 5-mL precision buret.5.3 Flasks, Erlenmeyer 250-mL capacity with ground glass stoppers.5.4 Flask, Erlenmeyer wide-mouthed, 250-mL capacity.5.5 Beakers, assorted sizes.5
20、.6 Bottles, amber, for storage of iodine and thiosulfate solutions.5.7 Funnels, 100-mm top inside diameter.5.8 Filter Paper, 18.5-cm prefolded paper, Whatman No. 2V or equivalent.5.9 Pipets, volumetric type, 5.0, 10.0, 25.0, 50.0, and 100.0-mL capacity.5.10 Volumetric Flasks, 1 L.5.11 Graduated Cyli
21、nders, 100 mL and 500 mL.6. Reagents6.1 Purity of ReagentsReagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is intended that allreagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, wheresuch specificat
22、ions are available.5 Other grades may be used, provided it is first ascertained that the reagent is of sufficiently highpurity to permit its use without lessening the accuracy of the determination.6.2 Purity of WaterReferences to water shall be understood to mean reagent water conforming to Specific
23、ation D1193 forType II reagent water.6.3 Hydrochloric Acid, concentrated.6.4 Sodium Thiosulfate, (Na2S2O3 5H2O).6.5 Iodine, United States Pharmacopeia, resublimed crystals.6.6 Potassium Iodide.6.7 Potassium Iodate, primary standard.6.8 Starch, soluble potato or arrowroot.6.9 Sodium Carbonate.7. Haza
24、rds7.1 Several potential hazards are associated with conducting this test procedure. It is not the purpose of this standard to addressall potential health and safety hazards encountered with its use. The user is responsible for establishing appropriate health andsafety practices before use of this t
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD46072014STANDARDTESTMETHODFORDETERMINATIONOFIODINENUMBEROFACTIVATEDCARBON 测定 活性炭 碘值 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-517486.html