ASTM D1945-2003(2010) Standard Test Method for Analysis of Natural Gas by Gas Chromatography《气相色谱法分析天然气的标准试验方法》.pdf
《ASTM D1945-2003(2010) Standard Test Method for Analysis of Natural Gas by Gas Chromatography《气相色谱法分析天然气的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D1945-2003(2010) Standard Test Method for Analysis of Natural Gas by Gas Chromatography《气相色谱法分析天然气的标准试验方法》.pdf(17页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D1945 03 (Reapproved 2010)Standard Test Method forAnalysis of Natural Gas by Gas Chromatography1This standard is issued under the fixed designation D1945; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of las
2、t revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of the chemi-cal composition of natural gases and similar gaseous mixtureswith
3、in the range of composition shown in Table 1. This testmethod may be abbreviated for the analysis of lean naturalgases containing negligible amounts of hexanes and higherhydrocarbons, or for the determination of one or more compo-nents, as required.1.2 The values stated in inch-pound units are to be
4、 regardedas standard. The values given in parentheses are mathematicalconversions to SI units that are provided for information onlyand are not considered standard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the
5、 user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D2597 Test Method for Analysis of Demethanized Hydro-carbon Liquid Mixtures Containing Nitrogen and CarbonDi
6、oxide by Gas ChromatographyD3588 Practice for Calculating Heat Value, CompressibilityFactor, and Relative Density of Gaseous FuelsE260 Practice for Packed Column Gas Chromatography3. Summary of Test Method3.1 Components in a representative sample are physicallyseparated by gas chromatography (GC) an
7、d compared tocalibration data obtained under identical operating conditionsfrom a reference standard mixture of known composition. Thenumerous heavy-end components of a sample can be groupedinto irregular peaks by reversing the direction of the carrier gasthrough the column at such time as to group
8、the heavy endseither as C5and heavier, C6and heavier, or C7and heavier. Thecomposition of the sample is calculated by comparing eitherthe peak heights, or the peak areas, or both, with the corre-sponding values obtained with the reference standard.4. Significance and Use4.1 This test method is of si
9、gnificance for providing data forcalculating physical properties of the sample, such as heatingvalue and relative density, or for monitoring the concentrationsof one or more of the components in a mixture.5. Apparatus5.1 DetectorThe detector shall be a thermal-conductivitytype, or its equivalent in
10、sensitivity and stability. The thermalconductivity detector must be sufficiently sensitive to producea signal of at least 0.5 mV for 1 mol % n-butane in a 0.25-mLsample.5.2 Recording InstrumentsEither strip-chart recorders orelectronic integrators, or both, are used to display the separatedcomponent
11、s. Although a strip-chart recorder is not requiredwhen using electronic integration, it is highly desirable forevaluation of instrument performance.5.2.1 The recorder shall be a strip-chart recorder with afull-range scale of 5 mV or less (1 mV preferred). The width ofthe chart shall be not less than
12、 150 mm. A maximum penresponse time of2s(1spreferred) and a minimum chart speedof 10 mm/min shall be required. Faster speeds up to 1001This test method is under the jurisdiction ofASTM Committee D03 on GaseousFuels and is the direct responsibility of Subcommittee D03.07 on Analysis ofChemical Compos
13、ition of Gaseous Fuels.Current edition approved Jan. 1, 2010. Published March 2010. Originallyapproved in 1962. Last previous edition approved in 2001 as D194596(2001).DOI: 10.1520/D1945-03R10.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at se
14、rviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.TABLE 1 Natural Gas Components and Range ofComposition CoveredComponent Mol %Helium 0.01 to 10Hydrogen 0.01 to 10Oxygen 0.01 to 20Nitrogen 0.01 to 100Carbon dioxide 0.0
15、1 to 20Methane 0.01 to 100Ethane 0.01 to 100Hydrogen sulfide 0.3 to 30Propane 0.01 to 100Isobutane 0.01 to 10n-Butane 0.01 to 10Neopentane 0.01 to 2Isopentane 0.01 to 2n-Pentane 0.01 to 2Hexane isomers 0.01 to 2Heptanes+ 0.01 to 11*A Summary of Changes section appears at the end of this standard.Cop
16、yright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.mm/min are desirable if the chromatogram is to be interpretedusing manual methods to obtain areas.5.2.2 Electronic or Computing IntegratorsProof of sepa-ration and response equivalent to th
17、at for a recorder is requiredfor displays other than by chart recorder. Baseline trackingwith tangent skim peak detection is recommended.5.3 AttenuatorIf the chromatogram is to be interpretedusing manual methods, an attenuator must be used with thedetector output signal to maintain maximum peaks wit
18、hin therecorder chart range. The attenuator must be accurate to within0.5 % between the attenuator range steps.5.4 Sample Inlet System:5.4.1 The sample inlet system shall be constructed ofmaterials that are inert and nonadsorptive with respect to thecomponents in the sample. The preferred material o
19、f construc-tion is stainless steel. Copper, brass, and other copper-bearingalloys are unacceptable. The sample inlet system from thecylinder valve to the GC column inlet must be maintained at atemperature constant to 61C.5.4.2 Provision must be made to introduce into the carriergas ahead of the anal
20、yzing column a gas-phase sample that hasbeen entrapped in a fixed volume loop or tubular section. Thefixed loop or section shall be so constructed that the totalvolume, including dead space, shall not normally exceed 0.5mL at 1 atm. If increased accuracy of the hexanes and heavierportions of the ana
21、lysis is required, a larger sample size may beused (see Test Method D2597). The sample volume must bereproducible such that successive runs agree within 1 % oneach component. A flowing sample inlet system is acceptableas long as viscosity effects are accounted for.NOTE 1The sample size limitation of
22、 0.5 mL or smaller is selectedrelative to linearity of detector response, and efficiency of columnseparation. Larger samples may be used to determine low-quantitycomponents to increase measurement accuracy.5.4.3 An optional manifold arrangement for enteringvacuum samples is shown in Fig. 1.5.5 Colum
23、n Temperature Control:5.5.1 IsothermalWhen isothermal operation is used,maintain the analyzer columns at a temperature constant to0.3C during the course of the sample run and correspondingreference run.5.5.2 Temperature ProgrammingTemperature program-ming may be used, as feasible. The oven temperatu
24、re shall notexceed the recommended temperature limit for the materials inthe column.5.6 Detector Temperature ControlMaintain the detectortemperature at a temperature constant to 0.3C during thecourse of the sample run and the corresponding reference run.The detector temperature shall be equal to or
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD194520032010STANDARDTESTMETHODFORANALYSISOFNATURALGASBYGASCHROMATOGRAPHY 色谱 分析 天然气 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-510722.html