《ASTM C1590-2004(2014) Standard Practice for Alternate Actinide Calibration for Inductively Coupled Plasma-Mass Spectrometry《电感耦合等离子体质谱法的交替锕系校正的标准操作规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM C1590-2004(2014) Standard Practice for Alternate Actinide Calibration for Inductively Coupled Plasma-Mass Spectrometry《电感耦合等离子体质谱法的交替锕系校正的标准操作规程》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C1590 04 (Reapproved 2014)Standard Practice forAlternate Actinide Calibration for Inductively CoupledPlasma-Mass Spectrometry1This standard is issued under the fixed designation C1590; the number immediately following the designation indicates the year oforiginal adoption or, in the cas
2、e of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice provides guidance for an alternate linearcalibration for the determination of
3、 selected actinide isotopesin appropriately prepared aqueous solutions by InductivelyCoupled Plasma-Mass Spectrometry (ICP-MS). This alternatecalibration is mass bias adjusted using thorium-232 (232Th) anduranium-238 (238U) standards. One of the benefits of thisstandard practice is the ability to ca
4、librate for the analysis ofhighly radioactive actinides using calibration standards atmuch lower specific activities. Environmental laboratories mayfind this standard practice useful if facilities are not available tohandle the highly radioactive standards of the individualactinides of interest.1.2
5、The instrument response for a series of determinationsof known concentration of232Th and238U defines the massversus response relationship. For each standard concentration,the slope of the line defined by232Th and238U is used to derivelinear calibration curves for each mass of interest usinginterfere
6、nce equations. The mass bias corrected calibrationcurves, although generated from interference equations, arespecific to the instrument operating parameters and tuning ineffect at the time of data acquisition. Because interferenceequations are part of the normal ICP-MS manufacturerssoftware package,
7、 this calibration methodology is widelyapplicable.1.3 For this standard practice, the actinide atomic massrange that has been studied is from amu 232244. Guidance foran extended range of amu 228248 is given in this practice.1.4 Using this practice, analyte concentrations are reportedat each amu and
8、not by element total (that is,239Pu versusplutonium).1.5 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsi
9、bility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2C859 Terminology Relating to Nuclear MaterialsC1168 Practice for Preparation and Dissolution of
10、 PlutoniumMaterials for AnalysisC1347 Practice for Preparation and Dissolution of UraniumMaterials for AnalysisC1411 Practice for The Ion Exchange Separation of Ura-nium and Plutonium Prior to Isotopic AnalysisC1414 Practice for The Separation of Americium fromPlutonium by Ion ExchangeC1463 Practice
11、s for Dissolving Glass Containing Radioac-tive and Mixed Waste for Chemical and RadiochemicalAnalysisD1193 Specification for Reagent Water3. Terminology3.1 Definitions:3.1.1 For definitions of terms relating to nuclear materials,refer to Terminology C859.3.1.2 AMUatomic mass unit.4. Summary of Pract
12、ice4.1 Calibration for the actinides by ICP-MS can be per-formed in a variety of ways with varying degrees of dataquality. An alternative calibration method has been developedto compensate for instrument mass bias using a generated massresponse curve defined by the232Th and238U data points. Themass
13、response curve defined by232Th and238U approximatesthe mass response curve from amu 232244 as verifiedexperimentally and graphically depicted in Fig. 1. The massresponse curve shown reflects the operating parameters and1This practice is under the jurisdiction of ASTM Committee C26 on NuclearFuel Cyc
14、le and is the direct responsibility of Subcommittee C26.05 on Methods ofTest.Current edition approved Jan. 1, 2014. Published February 2014. Originallyapproved in 2004. Last previous edition approved in 2009 as C1590 04 (2009).DOI: 10.1520/C1590-04R14.2For referenced ASTM standards, visit the ASTM w
15、ebsite, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United St
16、ates1tune of the particular instrument in use at the time of datacollection. Different tuning parameters or instrumentationcould result in varying degrees of negative, neutral, or positivemass bias. Because the mass response curve defined by232Thand238U used in this standard practice is determined d
17、uringeach calibration, all potential linear variations in mass bias arecompensated for.4.2 The alternative calibration in this standard practicecombines the features of an external linear calibration at eachmass of interest with the mass bias correction of a mass/response curve. The correction for m
18、ass bias is integrated intothe acquisition of the standard data through the use ofinterference equations which, are part of the normal softwarepackage for correction of isobaric interferences in ICP-MSanalyses. Multipoint calibration curves are generated at eachmass of interest, resulting in more ac
19、curate quantification thanthat of the typical “semi-quantitative” single point calibrationbased on the mass/response curve alone.4.3 Sample analyses for blanks and samples are performedusing a data acquisition method file without the interferenceequations that were used to derive the calibration cur
20、ves. Aftercalibration curves are generated, the calibration files are copiedor linked to the analytical acquisition method. The sampleresponses are acquired at each mass, and concentrationscalculated from the mass bias corrected calibration curves.Some ICP-MS vendor supplied control software will pe
21、rmit thelinking of separate calibration and acquisition files (that is, youcan choose which calibration files to use to quantitate anyparticular data set regardless of the acquisition file that wasused to acquire the data).4.4 Mixed calibration standard solutions are preparedthrough dilution of sing
22、le element stock standards of thoriumand known abundance uranium (normally depleted in235U)with dilute nitric acid to develop a calibration series coveringthe desired concentration range. Standard concentrations arecalculated for232Th and238U for each calibration solution.4.5 Bismuth-209 (209Bi) is
23、used as an internal standard andis added in a fixed quantity in all standards and samples tocorrect for both instrument drift and physical sample transportfluctuations.5. Significance and Use5.1 One of the benefits of this standard practice is the abilityto calibrate for the analysis of highly radio
24、active actinidesusing calibration standards at much lower specific activities(that is,232Th and238U). Environmental laboratories may findthis standard practice useful if facilities are not available tohandle the highly radioactive standards of the individualactinides of interest.5.2 The degree of ac
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC159020042014STANDARDPRACTICEFORALTERNATEACTINIDECALIBRATIONFORINDUCTIVELYCOUPLEDPLASMAMASSSPECTROMETRY

链接地址:http://www.mydoc123.com/p-465238.html