ISA 37 16 01-2002 A Guide for the Dynamic Calibration of Pressure Transducers《压力传感器动态校准指南》.pdf
《ISA 37 16 01-2002 A Guide for the Dynamic Calibration of Pressure Transducers《压力传感器动态校准指南》.pdf》由会员分享,可在线阅读,更多相关《ISA 37 16 01-2002 A Guide for the Dynamic Calibration of Pressure Transducers《压力传感器动态校准指南》.pdf(56页珍藏版)》请在麦多课文档分享上搜索。
1、A Guide for the DynamicCalibration of PressureTransducersApproved 21 November 2002ISA-37.16.01-2002 STANDARDISA The Instrumentation,Systems, andAutomation Society TMISA37.16.012002A Guide for the Dynamic Calibration of Pressure TransducersISBN: 1-55617-838-7Copyright 2002 by ISA The Instrumentation,
2、 Systems, and Automation Society. All rights reserved.Not for resale. Printed in the United States of America. No part of this publication may be reproduced,stored in a retrieval system, or transmitted, in any form or by any means (electronic, mechanical,photocopying, recording, or otherwise), witho
3、ut the prior written permission of the Publisher.ISA67 Alexander DriveP. O. Box 12277Research Triangle Park, North Carolina 27709- 3 - ISA-37.16.01-2002PrefaceThis preface, as well as all footnotes and annexes, is included for information purposes and is not part ofISA-37.16.01-2002.This document ha
4、s been prepared as part of the service of ISA The Instrumentation, Systems, andAutomation Society toward a goal of uniformity in the field of instrumentation. To be of real value, thisdocument should not be static but should be subject to periodic review. Toward this end, the Societywelcomes all com
5、ments and criticisms and asks that they be addressed to the Secretary, Standards andPractices Board; ISA; 67 Alexander Drive; P. O. Box 12277; Research Triangle Park, NC 27709;Telephone (919) 549-8411; Fax (919) 549-8288; E-mail: standardsisa.org.The ISA Standards and Practices Department is aware o
6、f the growing need for attention to the metricsystem of units in general, and the International System of Units (SI) in particular, in the preparation ofinstrumentation standards. The Department is further aware of the benefits to USA users of ISAstandards of incorporating suitable references to the
7、 SI (and the metric system) in their business andprofessional dealings with other countries. Toward this end, this Department will endeavor to introduceSI-acceptable metric units in all new and revised standards, recommended practices, and technicalreports to the greatest extent possible. Standard f
8、or Use of the International System of Units (SI): TheModern Metric System, published by the American Society for Testing an overdamped (aperiodic) system comes to rest without overshoot; and a criticallydamped system is at the point of change between the underdamped and the overdamped conditions.4.4
9、.3 Damping ratio (z )The ratio of the actual damping to the damping required for critical damping. In Equation 4.2, thecoefficient z is the damping ratio. Typical dynamic pressure transducers have damping ratios much lessthan unity; consequently values of wdand wrvery nearly coincide. The damping ra
10、tio is definedspecifically for a linear second-order system. Transducers with more resonances are approximated byassociating a “damping ratio” with each resonant frequency.The damping ratio is a useful parameter in both the time and frequency domain. In the time domain, z isrelated to the amount of
11、overshoot (see Figure 2) and influences the number of ringing cycles presentafter a shock excitation. In the amplitude response, z is related to the height of the peak at the resonantfrequency.4.4.4 Rise timeThe length of time required for the output of a transducer to rise from 10% of its initial v
12、alue to 90% of itsfinal value when excited by a step change in measurand. Rise time is related to transducer frequencyresponse.4.4.5 OvershootThe amount of output measured beyond the final steady output value in response to a step change in themeasurand. The maximum theoretical overshoot of an ideal
13、 second-order transducer is 100 percent; thisoccurs when z is zero. Overshoot is determined as(Eq. 4.6)z-zp-=21e100overshootfor the condition z 0.1.4.4.6 Settling timeThe time required after the application of a step change in measurand for the transducer output to settlewithin a small specified per
14、centage (5 percent) of its final value. For the ideal second-order transducerwith small z(Eq. 4.7)d2s13twzz-=The settling time increases with smaller z and wd. The number of oscillations at wdrequired to settle within5 percent of final value for the ideal second-order transducer isISA-37.16.01-2002
15、- 22 -(Eq. 4.8)zpz-=213N24.4.7 Discharge time constant (DTC)Time required for a transducer or measuring system to discharge its signal to 37% of the original valuefrom a step-change measurand. It relates to low-frequency measuring capability for both transient andsinusoidal events.5 Dynamic pressure
16、 generatorsThe dynamic calibration of pressure transducers requires that the measurand produced by a dynamicpressure generator varies in time in both a known and an appropriate manner. With some generators, thepressure-time relationship can be predicted quite accurately. With others, the pressure-ti
17、me relationshipcan be established accurately only with the aid of comparison to referenced pressure transducers. Whilereproducibility is a highly desirable characteristic of the dynamic pressure generator, it is not an essentialcharacteristic. When such a characteristic is lacking in a generator, fu
18、ll dependence on the referencetransducer is required.Dynamic-pressure generators fall into two basic classes: aperiodic and periodic. The aperiodic generatorsare characterized by the pulse shapes they produce, such as the step or the peaking pulse. Quick-opening valve devices and pulse generators pr
19、oduce pressure rise times generally in the milli-secondrange or less. The rise times and the pressure amplitudes generated by these devices vary markedly fromone type of aperiodic pressure generator to another. The shock tube, for example, is capable ofgenerating pressure steps having rise times in
20、the nanosecond range. A number of the dynamiccalibrators described in this clause are now commercial products.Pressure step, as used in this document, is defined as a change in measurand in which the rise time isless than one-fifth the rise time of the transducer measuring it.Sinusoidal pressure gen
21、erators, which require the use of a transfer standard, are the most useful of thevarious periodic pressure generators available, however, and these devices are limited as to useablerange of frequency dynamic pressure ratio and dynamic amplitude. Nonsinusoidal pressure generatorsof significant useful
22、ness include the square wave or rectangular wave generators, which may beconsidered as a special case of the aperiodic or step-function generators. Figures 4 and 5 present asummary of the capabilities of the dynamic pressure generators.5.1 Shock tubeA shock tube, in its simplest form, consists of tw
23、o sections of tubing separated by a thin diaphragm. Whenthese two sections are pressurized to different pressure levels, and the diaphragm is suddenly ruptured,the higher-pressure gas will immediately begin to flow and compress the gas at a lower pressure(References 56, 57, 58).It should be noted th
24、at most cold-gas, shock-tube-development work occurred in or before the 1960s.However, in 1997, a shock tube was designed and built at a university for a transducer manufacturer. Thedevelopment report for this new shock tube, Reference 64, also updates the literature through theintervening time peri
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ISA3716012002AGUIDEFORTHEDYNAMICCALIBRATIONOFPRESSURETRANSDUCERS 压力传感器 动态 校准 指南 PDF

链接地址:http://www.mydoc123.com/p-789698.html