1、A-A-54520 66 9999974 0062972 0 m z-=- A-A-54520 12 November 1991 COMMERCIAL ITEM DESCRIPTION THERMOMETER, SELF-INDICATING, BIMETALLIC %, The General Services Administration has authorized the use qf this Commercial Item Description in preference to Military Specification MIL-T-3462D which is cancell
2、ed. This Commercial Item Description covers a pocket size dial type thermometer, for use in liquid temperature measurements, inspection of food products, and in determining the temperature of shell eggs. Shall be a self-indicating dial type bimetallic thermometer with a O to 220 degrees Fahrenheit (
3、F) scale range and increments of 2 degrees (FI. The nominal overall length of the thermometer assembly shall be 5-1/2 inches. The thermometer shall be supplied with a carrying case that includes a pocket clip. Shall be designed to indicate temperatures by the rotation action of a bimetallic element
4、which turns a staff that is connected to a pointer. A calibration nut shall provide a means for recalibrating without disassembling the thermometer. Shall be supplied ready for use by being preset to provide thermometer readings that have a +/- 2 degree (FI accuracy. The thermometer shall be tested
5、for accuracy in accordance with the testing procedure indicated in the Quality Assurance Provisions of this document. Shall be designed to respond rapidly to temperature changes. The time for a thermometer reading to rise from 50 degrees to 180 degrees (FI shall not exceed 10 seconds, after the ther
6、mometer is immersed in a 212 degree (FI liquid bath after being transferred from a 32 degree (FI liquid bath. The thermometer shall be tested for speed of response in accordance with the testing procedure indicated in the Quality Assurance Provisions of this document. AMSC N/A FSC 6683 -_- DISTRIBUT
7、ION STATEMENT A: Approved for public release; distribution is unlimited. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A- A-54520 Shall consist of a pointed metal stem that is attached to a metal case with a calibration nut. The stem, case, and cal
8、ibration nut shall be fabricated from corrosion-resisting steel. The stem shall be positioned perpendicular to the center of the base of the case. The stem shall house a bimetallic heat-sensitive element. Shall be supplied with a pointed metal stem with an exposed length of 4-3/4 to 5-!/4 inches. Th
9、e out.side diameter of the stem shall be at least 115 inches when measured at - the midpoint. of its axial length. Shall be supplied with a metal case that encloses a graduated dial, a black pointer, and a plrotective window. The window shall be securely sealed to the case to provide a watertight th
10、eimoineter. The dimensions of the case shall aow for at least U 7,s inch diameter field view f the LS! ui-ide-, :hi- window. Shall be suyplied with an pt.ic*aly clear glass ur plastic window. The dial crystal material shall be capable of being subjecit,ed tcl a 225 degree (FI temperature for a one h
11、our period without exhibiting any evidence of discoloration 01% d e f or mat i nn . Shall be supplied with a circular dial with graduations and numerals that are black on a white or silver background. The 0 to 220 degree (FI temperature range shall extend around the circular face to cover at least a
12、 sector formed by a 300 degree arc. A minimum 20 degree arc sector shall separate the positions for the O and 2.20 degree (FI readings. The graduations shall be divided in 2 degree (FI increments with each 10 degree (FI increment longer than the other increments. At least every 20 degree (FI increme
13、nt shall include the appropriate Arabic number marking for a direct temperature reading. The graduation markings shall be permanent and clear. The dial shall also be marked to include the model number and name or registered trademark of the manufacturer or supplier. As an alternate the model number
14、may only be indicated on the printed instructions that are to be supplied with the thermometer. 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54520 66 m 9999974 O062973 4 m A-A-54520 Shall be provided with a pointer that is sufficiently dampe
15、ned to provide a temperature reading that does not change more than 2 degrees (FI when the thermometer is subjected to the stability testing pr3ocedure Indicated in the Quality Assurance Provisions of this document. Shall be fitted with a calibration nut that allows the dial adjustments to be made b
16、y using a wrench (not required to be supplied with the thermometer) to grip and turn the nut. Turning the nut shall result in the case and dial being turned and repositioned under the pointer. The calibration nut fitting shall have sufficient friction to require a torque in the range of 15 to 68 inc
17、h-oz to be applied in order to turn the calibration nut. The thermometer shall be tested to determine the tightness of the calibration nut in accordance with the testing procedure in the Quality Assurance Provisions of this document. Shall be supplied with a pocket size carrying case that includes a
18、 pocket clip which is suitably protected against corrosion. The carrying case shall be fabricated from a hard rubber or a rigid plastic. The thermometer, after insertion in the carrying case shall be held in place by a frictional spring clip type support located within the.cavity of the case. The po
19、cket clip-on case shall be designed for alternate use as a handle for the thermometer by allowing the thermometer stem to be inserted into a loop or gap in the pocket clip. Shall be provided with printed instructions, on either a product box or separate enclosure, that indicate the proper use and ca
20、re of the thermometer. A cautionary statement shall be included that indicates that the thermometer is being supplied to be accurate within 2 degrees (FI, unless damaged by shock or temperature extremes, but can be adjusted to a 1 degree (FI accuracy at any point on the dial scale by carefully recal
21、ibrating at that specific temperature. The cautionary statement may be supplied on an insert that is separate from the other instructions. _-_- WorkmanshiE, The thermometer and carrying case shall be free from defects which detract from their appearance or impair their serviceability. Unit. - Each (
22、EA). One thermometer, as specified, constitutes one unit. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-g ci- A-A-54520 bb - 9999974 0062974 b - A-A-54520 - Responsibilitx _-_-_ _- for inspection. Unless otherwise specified in the contract or pur
23、chase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contract or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified h
24、erein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements. Rg!gygg, Records of examinations and tests p
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DLAAA545201991THERMOMETERSELFINDICATINGBIMETALLIC 双金属 自动 指示 温度计 PDF

链接地址:http://www.mydoc123.com/p-686471.html