《ASTM F2213-2006(2011) Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment《测量磁共振环境中医疗设备上的磁感应扭矩的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2213-2006(2011) Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment《测量磁共振环境中医疗设备上的磁感应扭矩的标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2213 06 (Reapproved 2011)Standard Test Method forMeasurement of Magnetically Induced Torque on MedicalDevices in the Magnetic Resonance Environment1This standard is issued under the fixed designation F2213; the number immediately following the designation indicates the year oforiginal
2、adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the measurement of the mag-netically induced
3、 torque produced by the static magnetic fieldin the magnetic resonance environment on medical devices andthe comparison of that torque to the equivalent torque appliedby the gravitational force to the implant.1.2 This test method does not address other possible safetyissues which include but are not
4、 limited to issues of magneti-cally induced force due to spatial gradients in the staticmagnetic field, RF heating, induced heating, acoustic noise,interaction among devices, and the functionality of the deviceand the MR system.1.3 The torque considered here is the magneto-static torquedue to the in
5、teraction of the MRI static magnetic field with themagnetization in the implant. The dynamic torque due tointeraction of the static field with eddy currents induced in arotating device is not addressed in this test method. Currents inlead wires may induce a torque as well.1.4 The sensitivity of the
6、torque measurement apparatusmust be greater than110 the “gravity torque,” the product of thedevices maximum linear dimension and its weight.1.5 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.6 This standard does not purport t
7、o address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2F2052 Te
8、st Method for Measurement of Magnetically In-duced Displacement Force on Medical Devices in theMagnetic Resonance EnvironmentF2119 Test Method for Evaluation of MR Image Artifactsfrom Passive ImplantsF2182 Test Method for Measurement of Radio FrequencyInduced Heating On or Near Passive Implants Duri
9、ngMagnetic Resonance ImagingF2503 Practice for Marking Medical Devices and OtherItems for Safety in the Magnetic Resonance Environment2.2 Other Standards:IEC 60601-2-33 Ed. 2.0 Medical Electrical EquipmentPart 2: Particular Requirements for the Safety of MagneticResonance Equipment for Medical Diagn
10、osis, 20023ISO 13485:2003(E) Medical DevicesQuality Manage-ment SystemsRequirements for Regulatory Purposes,definition 3.733. Terminology3.1 DefinitionsFor the purposes of this test method, thedefinitions in 3.1.1-3.1.18 shall apply:3.1.1 diamagnetic materiala material whose relative per-meability i
11、s less than unity.3.1.2 ferromagnetic materiala material whose magneticmoments are ordered and parallel producing magnetization inone direction.3.1.3 magnetic induction or magnetic flux density (B inT)that magnetic vector quantity which at any point in amagnetic field is measured either by the mecha
12、nical forceexperienced by an element of electric current at the point, or bythe electromotive force induced in an elementary loop duringany change in flux linkages with the loop at the point. Themagnetic induction is frequently referred to as the magneticfield. B0is the static field in an MR system.
13、 Plain type indicatesa scalar (for example, B) and bold type indicates a vector (forexample, B).3.1.4 magnetic field strength (H in A/m)strength of theapplied magnetic field.3.1.5 magnetic resonance (MR)resonant absorption ofelectromagnetic energy by an ensemble of atomic particlesituated in a magne
14、tic field.1This test method is under the jurisdiction of ASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved Oct. 1, 2011. Published October 2011. Originallyapproved in 2002. Last pre
15、vious edition approved in 2006 as F2213 06. DOI:10.1520/F2213-06R11.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM w
16、ebsite.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.1.6 magnetic resonance diagnostic devicea device
17、in-tended for general diagnostic use to present images whichreflect the spatial distribution or magnetic resonance spectra, orboth, which reflect frequency and distribution of nuclei exhib-iting nuclear magnetic resonance. Other physical parametersderived from the images or spectra, or both, may als
18、o beproduced.3.1.7 magnetic resonance (MR) environmentvolumewithin the 0.50 mT (5 gauss (G) line of an MR system, whichincludes the entire three dimensional volume of space sur-rounding the MR scanner. For cases where the 0.50 mT line iscontained within the Faraday shielded volume, the entire roomsh
19、all be considered the MR environment.3.1.8 magnetic resonance equipmentmedical electricalequipment which is intended for in-vivo magnetic resonanceexamination of a patient. The MR equipment comprises allparts in hardware and software from the supply mains to thedisplay monitor. The MR equipment is a
20、 ProgrammableElectrical Medical System (PEMS).3.1.9 magnetic resonance examination (MRExamination)process of acquiring data by magnetic reso-nance from a patient.3.1.10 magnetic resonance imaging (MRI)imaging tech-nique that uses static and time varying magnetic fields toprovide images of tissue by
21、the magnetic resonance of nuclei.3.1.11 magnetic resonance system (MR System)ensembleof MR equipment, accessories including means for display,control, energy supplies, and the MR environment.IEC 606012333.1.12 magnetically induced displacement forceforce pro-duced when a magnetic object is exposed t
22、o the spatial gradientof a magnetic field. This force will tend to cause the object totranslate in the gradient field.3.1.13 magnetically induced torquetorque produced whena magnetic object is exposed to a magnetic field. This torquewill tend to cause the object to align itself along the magneticfie
23、ld in an equilibrium direction that induces no torque.3.1.14 magnetization (M in T)magnetic moment per unitvolume.3.1.15 medical deviceany instrument, apparatus, imple-ment, machine, appliance, implant, in vitro reagent or calilbra-tor, software, material, or other similar or related article,intende
24、d by the manufacturer to be used, alone or in combi-nation, for human beings for one or more of the specificpurpose(s) of:(1) diagnosis, prevention, monitoring, treatment, or alleviation of dis-ease,(2) diagnosis, monitoring, treatment, alleviation of, or compensation foran injury,(3) investigation,
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF221320062011STANDARDTESTMETHODFORMEASUREMENTOFMAGNETICALLYINDUCEDTORQUEONMEDICALDEVICESINTHEMAGNETICRESONANCEENVIRONMENT

链接地址:http://www.mydoc123.com/p-538068.html