FORD FLTM BI 109-02-2011 COLOR MEASUREMENT《颜色测量[取代 福特FLTM BO 109-02]》.pdf
《FORD FLTM BI 109-02-2011 COLOR MEASUREMENT《颜色测量[取代 福特FLTM BO 109-02]》.pdf》由会员分享,可在线阅读,更多相关《FORD FLTM BI 109-02-2011 COLOR MEASUREMENT《颜色测量[取代 福特FLTM BO 109-02]》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、 FORD LABORATORY TEST METHODFLTM BI 109-02 Date Action Revisions Rev. 2COLOR MEASUREMENT 1. Scope and Field of Application 1.1 Scope This objective method covers an instrumental comparison of a samples color against a reference by defining measurement conditions and measurement procedures for the me
2、asurement of effect colors. The aim is to instrumentally determine those color differences between sample and reference, which are caused by materials, processing, or which arise after testing of, e.g. aging. The measurement results are similar to those of the visual perception, but visual assessmen
3、t according to Ford Laboratory Test Method BI 109-01 will always override numerical measurement data and is compulsory for any type of appearance approval. 1.2 Field of Application The application differentiates solid (uni) and effect colors and is recommended, but not limited to the following mater
4、ials: Substrate Solid Colors (uni) Effect Colors Plastic - solids with or without texture - painted, excluding effect colors - foils, sheets 0: away from illumination). 2011 10 19 Revision Interior/exterior alignment for distribution within the specific region only (most be coded different to master
5、). Reference Approval A visual approval to either a master or regional standard to validate Design Intent and verify feasibility of specific materials. 2.5 Pile-Direction Sensitive A fabric that appears different after brushing over the surface with the hand. 2.6 Reference A master or any other phys
6、ical sample that the measurement of samples is correlated to. If only compared within a sample lot, the one sample is chosen to be reference that is visually closest to the idea/vision of the evaluator. 2.7 Structure Structures are imprints or regular sewing lines that generate locally different pil
7、e heights and therefore lead to visually perceivable contrast differences. FORD LABORATORY TEST METHODFLTM BI 109-02 Copyright 2011, Ford Global Technologies, Inc. Page 3 of 7 3. Requirements 3.1 General The evaluator shall be trained on the instrument to be used by either the manufacturer of the eq
8、uipment, certified dealers or by Ford personnel. 3.2 Equipment & Maintenance 3.2.1 Equipment Requirements Measurements shall be made with spectrophotometers meeting the requirements of ASTM E805 (45/0 circumferential) as well as ASTM E2194-09 & ASTM E2539-08 (multi-angle with 45 illumination). The s
9、pectral resolution of the color measuring instrument shall be 10 nm maximum, uniformly distributed over the spectral measuring range from at least 400 - 700 nm. The repeatability of measurements shall be less than 0.05 E* CIELAB units for the white CERAM Color Standards Series II (CCSSII, formerly B
10、CRA Series II) tile. The inter instrument agreement shall be less than 0.2 E* CIELAB units for the average of the CCSSII tiles Daylight illumination (CIE Illuminant D65), Incandescent illumination (CIE Illuminant A) and Fluorescent three-band illumination (CIE Illuminant F11) shall be computable as
11、well as the respective metamerism indices (according to ISO/CIE 23603). 3.2.2 Instruments proposed Aligned with the differentiation of solid (uni) and effect colors, color spectrophotometers meeting the requirements of this method and therefore recommended are: Instrument Solid (uni) Colors Effect C
12、olors Master Konica Minolta CM-2500c CT Byk mac Secondary Byk Gardner spectro guide X-Rite MA98, 96, 94 3.2.3 Maintenance and Calibration Maintenance and calibration is of prime importance in providing constant and reliable measurements. The instrument has to be maintained and calibrated in accordan
13、ce with the instructions of the manufacturer of the equipment. The instrument shall NOT be calibrated NOR charged within a series of measurements. The instrument shall be calibrated at least every 8thhour and preferably just prior to the sample measurement, using the white and black calibration tile
14、s furnished by the instrument manufacturer. At regular intervals or with the 8thhour calibration measurement equipment monitoring needs to be performed and put down in a report. A green or cyan test tile is recommended for this purpose, which is usually provided by the manufacturer of the equipment
15、(alternatively, CCSSII tiles may be used). If the instrument exceeds a tolerance of 0.2 E* for the required temperature the manufacturer of the equipment or a certified dealer needs to be contacted for assistance and support. FORD LABORATORY TEST METHODFLTM BI 109-02 Copyright 2011, Ford Global Tech
16、nologies, Inc. Page 4 of 7 4. Conditioning and Test Conditions 4.1 Conditioning of Samples Before testing all samples have to be conditioned to equilibrium at 21 2C and 50 5% RH according to SAE J1767. Textiles are considered in equilibrium after 24 h of conditioning or when the change in mass at 2
17、h intervals does not exceed 0.2%. Other materials, such as plastics, do normally require to have reached above temperature range prior to evaluation. The pile-direction sensitive fabrics have to be brushed prior to measurement: first against the pile-direction (sample appears darker), then with the
18、pile-direction (sample appears lighter). For needle punch fabrics the clean, greaseless hand can be used, while all other pile-direction sensitive fabrics require a clothbrush with polyester monofilaments of at least 20 mm length and 0.25 mm thickness. Fabrics that are not pile-direction sensitive d
19、o not need to be brushed. 4.2 Test Conditions The sample and reference shall be visually checked for inhomogeneities in texture and gloss. The surface shall be free of dust, grease, contaminations or stains. All damaged areas that are no result of specific testing methods to be evaluated (e.g. scrat
20、ch tests, etc.) need to be excluded from the measurements. If there is a defined direction of the test piece (e.g. up and down), the sample and the reference shall be measured in the same direction. For translucent materials (e.g. fabrics, etc.) the substrate shall be chosen corresponding to the sub
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- FORDFLTMBI109022011COLORMEASUREMENT 颜色 测量 取代 福特 FLTMBO10902PDF

链接地址:http://www.mydoc123.com/p-746980.html