1、ISO/ASTM 51401:2013(E)An American National StandardStandard Practice forUse of a Dichromate Dosimetry System1This standard is issued under the fixed designation ISO/ASTM 51401; the number immediately following the designation indicates theyear of original adoption or, in the case of revision, the ye
2、ar of last revision.1. Scope1.1 This practice covers the preparation, testing, and proce-dure for using the acidic aqueous silver dichromate dosimetrysystem to measure absorbed dose to water when exposed toionizing radiation. The system consists of a dosimeter andappropriate analytical instrumentati
3、on. For simplicity, the sys-tem will be referred to as the dichromate system. The dichro-mate dosimeter is classified as a type I dosimeter on the basisof the effect of influence quantities. The dichromate systemmay be used as either a reference standard dosimetry system ora routine dosimetry system
4、.1.2 This document is one of a set of standards that providesrecommendations for properly implementing dosimetry inradiation processing, and describes a means of achievingcompliance with the requirements of ISO/ASTM Practice52628 for the dichromate dosimetry system. It is intended to beread in conju
5、nction with ISO/ASTM Practice 52628.1.3 This practice describes the spectrophotometric analysisprocedures for the dichromate system.1.4 This practice applies only to gamma radiation,X-radiation/bremsstrahlung, and high energy electrons.1.5 This practice applies provided the following conditionsare s
6、atisfied:1.5.1 The absorbed dose range is from 2 103to5104Gy.1.5.2 The absorbed dose rate does not exceed 600 Gy/pulse(12.5 pulses per second), or does not exceed an equivalent doserate of 7.5 kGy/s from continuous sources (1).21.5.3 For radionuclide gamma sources, the initial photonenergy shall be
7、greater than 0.6 MeV. For bremsstrahlungphotons, the initial energy of the electrons used to produce thebremsstrahlung photons shall be equal to or greater than 2MeV. For electron beams, the initial electron energy shall begreater than 8 MeV.NOTE 1The lower energy limits given are appropriate for a
8、cylindricaldosimeter ampoule of 12 mm diameter. Corrections for displacementeffects and dose gradient across the ampoule may be required for electronbeams (2). The dichromate system may be used at lower energies byemploying thinner (in the beam direction) dosimeter containers (see ICRUReport 35).1.5
9、.4 The irradiation temperature of the dosimeter shall beabove 0C and should be below 80C.NOTE 2The temperature coefficient of dosimeter response is knownonly in the range of 5 to 50C (see 5.2). Use outside this range requiresdetermination of the temperature coefficient.1.6 This standard does not pur
10、port to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific precau-tionary statements are given
11、 in 9.3.2. Referenced documents2.1 ASTM Standards:3E170 Terminology Relating to Radiation Measurements andDosimetryE178 Practice for Dealing With Outlying ObservationsE275 Practice for Describing and Measuring Performance ofUltraviolet and Visible SpectrophotometersE666 Practice for Calculating Abso
12、rbed Dose From Gammaor X RadiationE668 Practice for Application of Thermoluminescence-Dosimetry (TLD) Systems for Determining AbsorbedDose in Radiation-Hardness Testing of Electronic DevicesE925 Practice for Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Bandwidt
13、hdoes not Exceed 2 nmE958 Practice for Estimation of the Spectral Bandwidth ofUltraviolet-Visible Spectrophotometers2.2 ISO/ASTM Standards:351261 Practice for Calibration of Routine Dosimetry Sys-tems for Radiation Processing51707 Guide for Estimating Uncertainties in Dosimetry forRadiation Processi
14、ng52628 Practice for Dosimetry in Radiation Processing1This practice is under the jurisdiction of ASTM Committee E61 on RadiationProcessing and is the direct responsibility of Subcommittee E61.02 on DosimetrySystems, and is also under the jurisdiction of ISO/TC 85/WG 3.Current edition approved Sept.
15、 14, 2013. Published November 2013. Originallypublished asASTM E 1401 91.ASTM E 1401 961was adopted by ISO in 1998with the intermediate designation ISO 15561:1998(E). The present InternationalStandard ISO/ASTM 51401:2013(E) replaces ISO 15561 and is a major revision ofthe last previous edition ISO/A
16、STM 51401:2003(E).2The boldface numbers in parentheses refer to the bibliography at the end of thispractice.3For referenced ASTM and ISO/ASTM standards, visit the ASTM website,www.astm.org, or contact ASTM Customer Service at serviceastm.org. ForAnnual Book of ASTM Standards volume information, refe
17、r to the standardsDocument Summary page on the ASTM website. ISO/ASTM International 2013 All rights reserved152701 Guide for Performance Characterization of Dosim-eters and Dosimetry Systems for Use in Radiation Pro-cessing2.3 ISO/IEC Standards:417025 General Requirements for the Competence of Testi
18、ngand Calibration Laboratories2.4 Joint Committee for Guides in Metrology (JCGM)Reports:5JCGM 100:2008, GUM 1995, with minor corrections,Evaluation of measurement data Guide to the Expres-sion of Uncertainty in Measurement2.5 International Commission on Radiation Units and Mea-surements (ICRU) Repor
19、ts:6ICRU Report 35 Radiation Dosimetry: Electrons With InitialEnergies Between 1 and 50 MeVICRU Report 80 Dosimetry Systems for Use in RadiationProcessingICRU Report 85a Fundamental Quantities and Units forIonizing Radiation3. Terminology3.1 Definitions:3.1.1 approved laboratorylaboratory that is a
20、recognizednational metrology institute; or has been formally accredited toISO/IEC 17025; or has a quality system consistent with therequirements of ISO/IEC 17025.3.1.1.1 DiscussionA recognized national metrology insti-tute or other calibration laboratory accredited to ISO/IEC17025 should be used in
21、order to ensure traceability to anational or international standard. A calibration certificateprovided by a laboratory not having formal recognition oraccreditation will not necessarily be proof of traceability to anational or international standard.3.1.2 reference standard dosimetry systemdosimetry
22、system, generally having the highest metrological qualityavailable at a given location or in a given organization, fromwhich measurements made there are derived.3.1.3 type I dosimeterdosimeter of high metrologicalquality, the response of which is affected by individual influ-ence quantities in a wel
23、l-defined way that can be expressed interms of independent correction factors.3.2 Definitions of other terms used in this practice thatpertain to radiation measurement and dosimetry may be foundin ASTM Terminology E170. Definitions in E170 are compat-ible with ICRU Report 85a; that document, therefo
24、re, may beused as an alternative reference.4. Significance and use4.1 The dichromate system provides a reliable means formeasuring absorbed dose to water. It is based on a process ofreduction of dichromate ions to chromic ions in acidic aqueoussolution by ionizing radiation.4.2 The dosimeter is a so
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMISOASTM514012013STANDARDPRACTICEFORUSEOFADICHROMATEDOSIMETRYSYSTEM 使用 重铬酸盐 剂量 测定 系统 标准 实施 规程 PDF

链接地址:http://www.mydoc123.com/p-540772.html