1、Designation: E 923 97 (Reapproved 2003)Standard Specification forGlass Westergen Tube, Reusable1This standard is issued under the fixed designation E 923; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A
2、 number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification describes requirements for a tube thatmeasures the erythrocyte sedimentation rate (ESR). ESR is thesuspension stab
3、ility of red cells in diluted, anti-coagulatedhuman blood.1.1.1 The use of the term “rate” is, strictly speaking, notcorrect. The test measures the amount of settling of red cellsafter a specified time.1.2 The tubes are used together with a special rack to ensurethey remain in a vertical position du
4、ring the test.1.3 This specification includes many dimensional require-ments that are, for the most part, in agreement with the BritishStandards Institution, German Standards Institute, Interna-tional Committee for Standardization in Haematology, and theNational Committee for Clinical Laboratory Sta
5、ndards publi-cations on Westergren tubes. The clinical procedure using thetube described in this specification is known as the “Wester-gren Method.”1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this st
6、andard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:E 438 Specification for Glasses in Laboratory Apparatus2E 920 Specification for Commercially Packaged LaboratoryApparatus2E 9
7、21 Specification for Export Packaged Laboratory Appa-ratus2E 1133 Practice for Performance Testing of Packaged Labo-ratory Apparatus for United States Government Procure-ments2E 1157 Specification for Sampling and Testing of ReusableLaboratory Glassware23. Terminology3.1 Definitions of Terms Specifi
8、c to This Standard:3.1.1 reusablecapable of being used again.3.1.2 tubethe word “tube” rather than “pipet” is used todescribe this instrument. The word “pipet” should be reservedfor volume-measuring instruments thus designated. A tubeused for measurements of blood sedimentation rate is not avolume m
9、easuring instrument. In this connection, misunder-standing can occur when a Westergren “tube” is described as a“pipet.”3.1.3 WestergrenThe surname of the individual respon-sible for the design of the Westergren tube and the method ofuse.4. Classification4.1 This specification covers a tube that is i
10、ntended to beused until it is no longer considered functional for the purposeintended. The specification is specifically written for a reusableitem and is not to be confused with a disposable tube that isdescribed in other published standards.5. Materials5.1 The tubes made to this specification shal
11、l be fabricatedfrom borosilicate glass, Type I, Class B; or soda lime glass,Type II, in accordance with Specification E 438.6. Dimensions, Mass, and Permissible Variations6.1 DesignThe Westergren tube shall be made of thick-walled glass tubing. It shall be of one-piece construction,straight and with
12、 uniform bore. The ends of the tube shall beground flat, perpendicular to the tube axis and beveled asspecified in Fig. 1.6.2 DimensionsThe tube shall be made of tubing with anoutside diameter (OD) of 6.5 6 0.5 mm with an insidediameter (ID) of 2.55 mm 6 0.15 mm. The uniformity of thebore shall be 6
13、 0.1 mm throughout the tube. The tube shall be3006 1 mm long and ground and beveled at each end. The tubeshall have an inscribed graduated scale extending over thelower 200 6 0.35 mm of the tube. The tube should containapproximately 1 mL of blood when filled and adjusted to the1This specification is
14、 under the jurisdiction of ASTM Committee E41 onLaboratory Apparatus and is the direct responsibility of Subcommittee E41.01 onGlass Apparatus.Current edition approved Nov. 10, 1997. Published February 1998. Originallypublished as E 923 83. Last previous edition E 923 93.2Annual Book of ASTM Standar
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME92319972003STANDARDSPECIFICATIONFORGLASSWESTERGRENTUBEREUSABLE 重复使用 玻璃 韦斯特格伦氏 管用 标准规范 PDF

链接地址:http://www.mydoc123.com/p-533782.html