ASTM E788-1997(2008) 866 Standard Specification for Pipet Blood Diluting《血液稀释移液管用标准规范》.pdf
《ASTM E788-1997(2008) 866 Standard Specification for Pipet Blood Diluting《血液稀释移液管用标准规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM E788-1997(2008) 866 Standard Specification for Pipet Blood Diluting《血液稀释移液管用标准规范》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 788 97 (Reapproved 2008)Standard Specification forPipet, Blood Diluting1This standard is issued under the fixed designation E 788; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number i
2、n parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. Scope1.1 This specification covers requirements for glass reusableblood
3、 diluting pipets that are used for performing red and whitecell corpuscle determinations.2. Referenced Documents2.1 ASTM Standards:2E 694 Specification for Laboratory Glass Volumetric Appa-ratusE 920 Specification for Commercially Packaged LaboratoryApparatusE 921 Specification for Export Packaged L
4、aboratory Appa-ratusE 1133 Practice for Performance Testing of Packaged Labo-ratory Apparatus for United States Government Procure-mentsE 1157 Specification for Sampling and Testing of ReusableLaboratory Glassware3. Classification3.1 This specification covers two different pipet designs andpermissib
5、le alternatives.3.1.1 Red and white cell blood diluting pipets (see Fig. 1).3.1.2 Permissible alternative designs (see Fig. 2).4. Materials4.1 The pipets shall be made of common spirit bore whiteback tubing or clear glass with a white stripe applied to theouter surface of the tubing.4.2 The beads sh
6、all be made of glass or ceramic composi-tion and shall be of red, clear, or white coloring.5. Dimensions and Permissible Variations5.1 DesignThe red and white cell blood dilution pipetsshall consist of a small uniform bore glass tube which shallhave a bulb of proper size near the proximal end (see F
7、ig. 1).As an alternative, the pipets may be constructed of three piecesof glass that are fused together to form a one-piece pipet (seeFig. 2). The bulb shall contain a nonspherical glass or ceramicbead sufficiently large enough to prevent its being impacted inthe constriction portion of the bulb. Th
8、e cross section of thepipet at any point shall be circular. The distal end of the whitecell pipet shall be pulled to a point and then ground andpolished to a tapered tip. The distal end of the red cell pipetshall be ground and polished to a tapered tip. The insidediameter (ID) of the pipet bore at t
9、he tip end of the white cellpipet shall be 0.2 to 0.5 mm. The external diameter of theground and polished tip shall not exceed 2.0 mm. The overalllength of both pipets shall be 104 to 121 mm. The proximal endmay be of funnel design, with an exterior taper, or ground andpolished with a taper of suffi
10、cient angle to permit application ofa rubber suction tube or other suction device (see Fig. 2).5.2 CapacityThe capacity of the red cell pipet bulb shallbe 0.8 to 1.2 cm3and the capacity of the white cell bulb shall be0.2 to 0.4 cm3. The capacity of the red cell pipet stem (bulb todistal end) shall b
11、e 0.008 to 0.012 cm3and the capacity of thewhite cell pipet stem shall be 0.02 to 0.04 cm3.5.3 Capacity Markings:5.3.1 Pipet Bulb MarkingsThe red and white cell pipetbulb capacity shall be confined to and indicated by calibrationlines on the pipet. These lines shall be located on both sides ofthe bu
12、lb and shall be within 2 mm to 6 mm above the bulb(proximal end) and within 3 mm to 6 mm below the bulb (distalend). The exact manner for measuring these line placements isspecified in Fig. 1.5.3.2 Pipet Stem MarkingsThe graduation lines on thestem of the red and white cell pipets shall be equally d
13、ividedinto two or ten divisions. If there are two divisions, the twocalibration lines on the stem will be numbered 0.5 and 1. Ifthere are ten divisions, the ten calibration lines will benumbered 0.5 on the fifth line and 1 on the tenth linerespectively. The top line (above bulb at the proximal end)s
14、hall be numbered 101 on the red cell pipet and 11 on the whitecell pipet, representing stem and bulb volume collectively. Atthe option of the manufacturer, pipet stems may be graduatedwith split lines or partial length lines (see Fig. 2).1This specification is under the jurisdiction of ASTM Committe
15、e E41 onLaboratory Apparatus and is the direct responsibility of Subcommittee E41.01 onApparatus.Current edition approved Nov. 1, 2008. Published January 2008. Originallyapproved in 1981. Last previous edition approved in 2003 as E 788 97(2003).2For referenced ASTM standards, visit the ASTM website,
16、 www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME78819972008866STANDARDSPECIFICATIONFORPIPETBLOODDILUTING 血液 稀释 管用 标准规范 PDF

链接地址:http://www.mydoc123.com/p-533525.html