ASTM E2253-2016 Standard Test Method for Temperature and Enthalpy Measurement Validation of Differential Scanning Calorimeters《差示扫描量热计温度和焓测量验证的标准试验方法》.pdf
《ASTM E2253-2016 Standard Test Method for Temperature and Enthalpy Measurement Validation of Differential Scanning Calorimeters《差示扫描量热计温度和焓测量验证的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2253-2016 Standard Test Method for Temperature and Enthalpy Measurement Validation of Differential Scanning Calorimeters《差示扫描量热计温度和焓测量验证的标准试验方法》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E2253 16Standard Test Method forTemperature and Enthalpy Measurement Validation ofDifferential Scanning Calorimeters1This standard is issued under the fixed designation E2253; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revi
2、sion, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method provides procedures for validatingtemperature and enthalpy measurements of differe
3、ntial scan-ning calorimeters (DSC) and analytical methods based uponthe measurement of temperature or enthalpy (or heat), or both,by DSC. Performance parameters determined include tempera-ture and calorimetric repeatability (precision), detection limit,quantitation limit, linearity, and bias. This t
4、est method isapplicable to exothermic events.1.2 Validation of apparatus performance and analyticalmethods is requested or required for quality initiatives or whereresults may be used for legal purposes.1.3 The values stated in SI units are to regarded as standard.No other units of measurement are i
5、ncluded in this standard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations pri
6、or to use.2. Referenced Documents2.1 ASTM Standards:2E177 Practice for Use of the Terms Precision and Bias inASTM Test MethodsE473 Terminology Relating to Thermal Analysis and Rhe-ologyE967 Test Method for Temperature Calibration of Differen-tial Scanning Calorimeters and Differential Thermal Ana-ly
7、zersE968 Practice for Heat Flow Calibration of DifferentialScanning CalorimetersE1142 Terminology Relating to Thermophysical PropertiesE1860 Test Method for Elapsed Time Calibration of Ther-mal AnalyzersE1970 Practice for Statistical Treatment of ThermoanalyticalDataE2161 Terminology Relating to Per
8、formance Validation inThermal Analysis and Rheology2.2 FDA Publications:3Q2B Validation of Analytical Procedures Methodology, 62FR 27464, May 19, 19973. Terminology3.1 Technical terms used in this standard are defined inPractice E177 and in Terminologies E473, E1142, and E2161including analyte, dete
9、ction limit, differential scanningcalorimetry, enthalpy, extrapolated onset value (temperature),linearity, mean, precision, quantitation limit, relative standarddeviation, repeatability, standard deviation, thermal curve, andvalidation.4. Summary of Test Method4.1 Temperature and time are the primar
10、y independentparameters and heat flow is the primary dependent experimen-tal parameter provided by DSC. Integration of heat flow, as afunction of time, yields enthalpy (heat).4.1.1 Time, measured by the DSC apparatus, shall conformto better than 0.1 % verified by Test Method E1860 andreported.4.1.2
11、Temperature is directly measured by a temperaturesensor that is an integral part of the differential scanningcalorimetry apparatus.4.1.3 Heat flow, a measured value, is validated by itsintegration over time to obtain the desired calorimetric (en-thalpic) information of interest.4.2 Calorimetric vali
12、dation of a differential scanning calo-rimetric apparatus at a single temperature is performed usingthe indium metal melt as an analyte.4.3 Validation of a DSC method based upon enthalpicmeasurement may be performed using the test specimen as theanalyte.1This test method is under the jurisdiction of
13、ASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.10 onFundamental, Statistical and Mechanical Properties.Current edition approved April 1, 2016. Published April 2016. Originallyapproved in 2003. Last previous edition approved in 2011 as E2253 11. DOI:10.
14、1520/E2253-16.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from Food and Drug Administration (F
15、DA), 10903 New HampshireAve., Silver Spring, MD 20993-0002, http:/www.fda.gov.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States14.4 The enthalpy of three (or more) specime
16、ns (nominallyrepresenting the maximum, midpoint and minimum of therange of the test method) are measured in triplicate (or more).A fourth blank specimen, containing no analyte, is alsomeasured in triplicate.NOTE 1Repeatability is determined by performing a sufficient numberof determinations to calcu
17、late statistically valid estimates of the standarddeviation or relative standard deviation of the measurement.4.4.1 Calorimetric linearity and bias are determined fromthe best-fit (linear regression) straight-line correlation of theresults from measurements of the three (or more) specimens.4.4.2 Cal
18、orimetric detection limit and quantitation limit aredetermined from the standard deviation of the blank determi-nation.4.4.3 Calorimetric repeatability is determined from the re-peatability measurement of the three (or more) specimens.4.5 The temperature validation of a differential scanningcalorime
19、tric apparatus is performed at three temperatures usingindium, bismuth, and zinc metal melts as an analyte (seeSection 7).4.5.1 The melting temperature of three (or more) materialsrepresenting the maximum, midpoint, and minimum of thetemperature range of the test method are measured in triplicate(or
20、 more) (see Note 1).4.5.2 Temperature linearity and bias are determined fromthe best-fit (linear regression) straight-line correlation of theresults from the temperature measurements at the three (ormore) temperatures.4.5.3 Temperature repeatability, detection limit and quanti-tation limit are deter
21、mined from the standard deviation of thereplicate temperature measurements.5. Significance and Use5.1 This test method may be used to determine and validatethe performance of a particular DSC apparatus.5.2 This test method may be used to determine and validatethe performance of a particular method b
22、ased upon a DSCtemperature and enthalpic measurement, or both.5.3 This test method may be used to determine the repeat-ability of a particular apparatus, operator, or laboratory.5.4 This test method may be used for specification andregulatory compliance purposes.6. Apparatus6.1 Differential Scanning
23、 Calorimeter (DSC)The essentialinstrumentation required to provide the minimum differentialscanning calorimetric capability for this test method includes:6.1.1 DSC Test Chamber, composed of:6.1.1.1 Furnace(s), to provide uniform controlled heating orcooling of a specimen and reference to a constant
24、temperatureor at a constant rate within the applicable temperature range ofthe test method.6.1.1.2 Temperature Sensor, to provide an indication of thespecimen temperature to readability required.6.1.1.3 Differential Sensor, to detect a heat flow differencebetween the specimen and reference.6.1.1.4 A
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME22532016STANDARDTESTMETHODFORTEMPERATUREANDENTHALPYMEASUREMENTVALIDATIONOFDIFFERENTIALSCANNINGCALORIMETERS

链接地址:http://www.mydoc123.com/p-530693.html