《ASTM E2253-2011 Standard Test Method for Temperature and Enthalpy Measurement Validation of Differential Scanning Calorimeters《差示扫描热量计制温度和焓测量验证用标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2253-2011 Standard Test Method for Temperature and Enthalpy Measurement Validation of Differential Scanning Calorimeters《差示扫描热量计制温度和焓测量验证用标准试验方法》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、Designation:E2253081Designation: E2253 11Standard Test Method forTemperature and Enthalpy Measurement Validation ofDifferential Scanning Calorimeters1This standard is issued under the fixed designation E2253; the number immediately following the designation indicates the year oforiginal adoption or,
2、 in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTESections 3.1 and 6.1.3 were corrected editorially in August 2009.1. Scope1.1 This test
3、method provides procedures for validating enthalpic temperature and enthalpy measurements of differentialscanning calorimeters (DSC) and analytical methods based upon the measurement of temperature or enthalpy (or heat), orheatboth, by DSC. Performance parameters determined include temperature and c
4、alorimetric repeatability (precision), detectionlimit, quantitation limit, linearity, and bias. This test method is applicable to both exothermic and endothermic events.1.2 Validation of apparatus performance and analytical methods is requested or required for quality initiatives or where resultsmay
5、 be used for legal purposes.1.3 The values stated in SI units are to regarded as standard. No other units of measurement are included in this standard.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of thi
6、s standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E177 Practice for Use of the Terms Precision and Bias in ASTM Test MethodsE473 Terminology Relating to Thermal Analysis and
7、 RheologyE967 Test Method for Temperature Calibration of Differential Scanning Calorimeters and Differential Thermal AnalyzersE968 Practice for Heat Flow Calibration of Differential Scanning CalorimetersE1142 Terminology Relating to Thermophysical PropertiesE1860 Test Method for Elapsed Time Calibra
8、tion of Thermal AnalyzersE1970 Practice for Statistical Treatment of Thermoanalytical DataE2161 Terminology Relating to Performance Validation in Thermal Analysis2.2 Other Standard:United States Food and Drug Administration,Q2B Validation of Analytical Procedures: Methodology, 62 FR 27464, May 19,19
9、97 FDA Publications:3Q2B Validation of Analytical Procedures Methodology, 62 FR 27464, May 19, 19973. Terminology3.1Technical3.1 Technical terms used in this standard are defined in Practice E177 and in Terminologies E473, E1142, andE2161 including analyte, detection limit, differential scanning cal
10、orimetry, enthalpy, extrapolated onset value (temperature),linearity, mean, precision, quantitation limit, relative standard deviation, repeatability, standard deviation, thermal curve, andvalidation.1This test method is under the jurisdiction of ASTM Committee E37 on Thermal Measurements and is the
11、 direct responsibility of Subcommittee E37.10 on Fundamental,Statistical and Mechanical Properties.Current edition approved Sept. 1, 2008. Published November 2008. Originally approved in 2003. Last previous edition approved in 2003 as E225303. DOI:10.1520/E2253-08E01.Current edition approved Dec. 1,
12、 2011. Published February 2012. Originally approved in 2003. Last previous edition approved in 2008 as E2253 081. DOI:10.1520/E2253-11.2For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume in
13、formation, refer to the standards Document Summary page on the ASTM website.3Available from Food and Drug Administration (FDA), 5600 Fishers Ln., Rockville, MD 20857, http:/www.fda.gov.3Available from Food and Drug Administration (FDA), 10903 New Hampshire Ave., Silver Spring, MD 20993-0002, http:/w
14、ww.fda.gov.1This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consu
15、lt prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.4. Summary of Test Method4.1 Tempera
16、ture and time are the primary independent parameters and heat flow is the primary dependent experimentalparameter provided by DSC. Integration of heat flow, as a function of time, yields enthalpy (heat).4.1.1 Time, measured by the DSC apparatus, shall conform to better than 0.1 % verified by Test Me
17、thod E1860 and reported.4.1.2Heat flow, a measured value, is validated by its integration over time to obtain the desired calorimetric (enthalpic)information of interest. Determination and verification of enthalpy is the primary scope of this test method.4.1.2 Temperature is directly measured by a t
18、emperature sensor that is an integral part of the differential scanning calorimetryapparatus.4.1.3 Heat flow, a measured value, is validated by its integration over time to obtain the desired calorimetric (enthalpic)information of interest.4.2 Calorimetric validation of a differential scanning calor
19、imetric apparatus at a single temperature is performed using theindium metal melt as an analyte.4.3 Validation of a DSC method based upon enthalpic measurement may be performed using the test specimen as the analyte.4.4 The enthalpy of three (or more) specimens (nominally representing the maximum, m
20、idpoint and minimum of the range ofthe test method) are measured in triplicate (or more).Afourth blank specimen, containing no analyte, is also measured in triplicate.NOTE 1Repeatability is determined by performing a sufficient number of determinations to calculate statistically valid estimates of t
21、he standarddeviation or relative standard deviation of the measurement.4.4.1 Calorimetric linearity and bias are determined from the best-fit straight-line correlation of the results from measurementsof the three (or more) specimens.4.4.2 Calorimetric detection limit and quantitation limit are deter
22、mined from the standard deviation of the blank determination.4.4.3 Calorimetric repeatability is determined from the repeatability measurement of the three (or more) specimens.4.5 The temperature validation of a differential scanning calorimetric apparatus is performed at three temperatures using in
23、dium,bismuth, and zinc metal melts as an analyte (see Section 7).4.5.1 The melting temperature of three (or more) materials representing the maximum, midpoint, and minimum of thetemperature range of the test method are measured in triplicate (or more) (see Note 1).4.5.2 Temperature linearity and bia
24、s are determined from the best-fit straight-line correlation of the results from the temperaturemeasurements at the three (or more) temperatures.4.5.3 Temperature repeatability, detection limit and quantitation limit are determined from the standard deviation of the replicatetemperature measurements
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME22532011STANDARDTESTMETHODFORTEMPERATUREANDENTHALPYMEASUREMENTVALIDATIONOFDIFFERENTIALSCANNINGCALORIMETERS

链接地址:http://www.mydoc123.com/p-530692.html