1、Designation: E 2253 08Standard Test Method forEnthalpy Measurement Validation of Differential ScanningCalorimeters1This standard is issued under the fixed designation E 2253; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year
2、 of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method provides procedures for validatingenthalpic measurements of differential scanning calorimeters(DS
3、C) and analytical methods based upon the measurement ofenthalpy or heat by DSC. Performance parameters determinedinclude calorimetric repeatability (precision), detection limit,quantitation limit, linearity and bias. This test method isapplicable to both exothermic and endothermic events.1.2 Validat
4、ion of apparatus performance and analyticalmethods is requested or required for quality initiatives or whereresults may be used for legal purposes.1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standard does not pu
5、rport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E
6、 177 Practice for Use of the Terms Precision and Bias inASTM Test MethodsE 473 Terminology Relating to Thermal Analysis and Rhe-ologyE 967 Test Method for Temperature Calibration of Differ-ential Scanning Calorimeters and Differential ThermalAnalyzersE 968 Practice for Heat Flow Calibration of Diffe
7、rentialScanning CalorimetersE 1142 Terminology Relating to Thermophysical PropertiesE 1860 Test Method for Elapsed Time Calibration of Ther-mal AnalyzersE 1970 Practice for Statistical Treatment of Thermoanalyti-cal DataE 2161 Terminology Relating to Performance Validation inThermal Analysis2.2 Othe
8、r Standard:United States Food and Drug Administration, Q2B Valida-tion of Analytical Procedures: Methodology, 62 FR27464, May 19, 199733. Terminology3.1 Technical terms used in this test method are defined inPractice E 177 and in Terminologies E 473, E 1142, andE 2161, including analyte, detection l
9、imit, differential scanningcalorimetry, enthalpy, extrapolated onset value (temperature),linearity, mean, precision, quantitation limit, relative standarddeviation, repeatability, standard deviation, thermal curve andvalidation.4. Summary of Test Method4.1 Temperature and time are the primary indepe
10、ndentparameters and heat flow is the primary dependent experimen-tal parameter provided by DSC. Integration of heat flow, as afunction of time, yields enthalpy (heat).4.1.1 Time, measured by the DSC apparatus, shall conformto better than 0.1 % verified by Test Method E 1860 andreported.4.1.2 Heat fl
11、ow, a measured value, is validated by itsintegration over time to obtain the desired calorimetric (enthal-pic) information of interest. Determination and verification ofenthalpy is the primary scope of this test method.4.2 Calorimetric validation of a differential scanning calo-rimetric apparatus at
12、 a single temperature is performed usingthe indium metal melt as an analyte.4.3 Validation of a DSC method based upon enthalpicmeasurement may be performed using the test specimen as theanalyte.4.4 The enthalpy of three (or more) specimens (nominallyrepresenting the maximum, midpoint and minimum of
13、therange of the test method) are measured in triplicate (or more).A fourth blank specimen, containing no analyte, is alsomeasured in triplicate.1This test method is under the jurisdiction ofASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on ThermalTe
14、st Methods and Practices.Current edition approved Sept. 1, 2008. Published October 2008. Originallyapproved in 2003. Last previous edition approved in 2003 as E 225303.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual
15、 Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from FDA, 5600 Fishers Lane, Rockville, MD 20857.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.NOTE 1Repeatabilit
16、y is determined by performing a sufficient num-ber of determinations to calculate statistically valid estimates of thestandard deviation or relative standard deviation of the measurement.4.4.1 Calorimetric linearity and bias are determined fromthe best-fit straight-line correlation of the results fr
17、om mea-surements of the three (or more) specimens.4.4.2 Calorimetric detection limit and quantitation limit aredetermined from the standard deviation of the blank determi-nation.4.4.3 Calorimetric repeatability is determined from the re-peatability measurement of the three (or more) specimens.5. Sig
18、nificance and Use5.1 This test method may be used to determine and validatethe performance of a particular DSC apparatus.5.2 This test method may be used to determine and validatethe performance of a particular method based upon a DSCenthalpic measurement.5.3 This test method may be used to determin
19、e the repeat-ability of a particular apparatus, operator or laboratory.5.4 This test method may be used for specification andregulatory compliance purposes.6. Apparatus6.1 Differential Scanning CalorimeterThe essential in-strumentation required to provide the minimum differentialscanning calorimetri
20、c capability for this test method include:6.1.1 DSC Test Chamber, composed of:6.1.1.1 Furnace(s), to provide uniform controlled heating orcooling of a specimen and reference to a constant temperatureor at a constant rate within the applicable temperature range ofthe test method.6.1.1.2 Temperature S
21、ensor, to provide an indication of thespecimen temperature to readability required.6.1.1.3 Differential Sensor, to detect a heat flow differencebetween the specimen and reference.6.1.1.4 A means of sustaining a test chamber environmentof an inert purge gas at a rate of 10 to 50 mL/min 6 10 %.NOTE 2T
22、ypically, 99+ % pure nitrogen (or other inert gas, such asargon or helium) is employed when oxidation in air is a concern. Unlessthe effects of moisture are to be studied, the use of a dry purge gas isrecommended, especially for operation at subambient temperatures.6.1.2 Temperature Controller, capa
23、ble of executing a spe-cific temperature program by operating the furnace(s) betweenselected temperature limits at a rate of temperature changeconstant to 6 1 % or at an isothermal temperature constant to6 0.5 C.6.1.3 Data Collection Device, to provide a means of acquir-ing, storing, and displaying
24、measured or calculated signals, orboth. The minimum output signals required for differentialscanning calorimetry are heat flow, temperature, and time.6.1.4 Containers, (pans, crucibles, vials, lids, closures,seals, etc.) that are inert to the specimen and referencematerials and that are of suitable
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME22532008STANDARDMETHODFORENTHALPYMEASUREMENTVALIDATIONOFDIFFERENTIALSCANNINGCALORIMETERSPDF

链接地址:http://www.mydoc123.com/p-530691.html