ASTM E2253-2003 Standard Method for Enthalpy Measurement Validation of Differential Scanning Calorimeters《差示扫描量热计的焓测量有效性的标准方法》.pdf
《ASTM E2253-2003 Standard Method for Enthalpy Measurement Validation of Differential Scanning Calorimeters《差示扫描量热计的焓测量有效性的标准方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2253-2003 Standard Method for Enthalpy Measurement Validation of Differential Scanning Calorimeters《差示扫描量热计的焓测量有效性的标准方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 2253 03Standard Method forEnthalpy Measurement Validation of Differential ScanningCalorimeters1This standard is issued under the fixed designation E 2253; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of l
2、ast revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This standard provides procedures for validating enthal-pic measurements of differential scanning calorimeters (DSC)an
3、d analytical methods based upon the measurement of en-thalpy or heat by DSC. Performance parameters determinedinclude calorimetric repeatability (precision), detection limit,quantitation limit, linearity and bias. This method is applicableto both exothermic and endothermic events.1.2 Validation of a
4、pparatus performance and analyticalmethods is requested or required for quality initiatives or whereresults may be used for legal purposes.1.3 SI units are the standard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility o
5、f the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:E 177 Practice for Use of the Terms Precision and Bias inASTM Test Methods2E 473 Terminology Relating to
6、 Thermal Analysis2E 967 Practice for Temperature Calibration of DifferentialScanning Calorimeters and Differential Thermal Analyz-ers2E 968 Practice for Heat Flow Calibration of DifferentialScanning Calorimeters2E 1142 Terminology Relating to Thermophysical Proper-ties2E 1860 Test Method for Elapsed
7、 Time Calibration of Ther-mal Analyzers2E 1970 Practice for Statistical Treatment of Thermoanalyti-cal Data2E 2161 Terminology Relating to Performance Validation inThermal Analysis22.2 Other Standard:United States Food and Drug Administration, Q2B Valida-tion of Analytical Procedures: Methodology, 6
8、2 FR27464, May 19, 199733. Terminology3.1 Technical terms used in this standard are defined inPractice E 177 and in Terminologies E 473, E 1142, andE 2161.4. Summary of Test Method4.1 Temperature and time are the primary independentparameters and heat flow is the primary dependent experimen-tal para
9、meter provided by DSC. Integration of heat flow, as afunction of time, yields enthalpy (heat).4.1.1 Time, measured by the DSC apparatus, shall conformto better than 0.1 % verified by Test Method E 1860 andreported.4.1.2 Heat flow, a measured value, is validated by itsintegration over time to obtain
10、the desired calorimetric (enthal-pic) information of interest. Determination and verification ofenthalpy is the primary scope of this test method.4.2 Calorimetric validation of a differential scanning calo-rimetric apparatus at a single temperature is performed usingthe indium metal melt as an analy
11、te.4.3 Validation of a DSC method based upon enthalpicmeasurement may be performed using the test specimen as theanalyte.4.4 The enthalpy of three (or more) specimens (nominallyrepresenting the maximum, midpoint and minimum of therange of the test method) are measured in triplicate (or more).A fourt
12、h blank specimen, containing no analyte, is alsomeasured in triplicate.NOTE 1Repeatability is determined by performing a sufficient num-ber of determinations to calculate statistically valid estimates of thestandard deviation or relative standard deviation of the measurement.4.4.1 Calorimetric linea
13、rity and bias are determined fromthe best-fit straight-line correlation of the results from mea-surements of the three (or more) specimens.1This test method is under the jurisdiction of ASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on ThermalAnalys
14、is Methods.Current edition approved Jan. 10, 2003. Published April 2003.2Annual Book of ASTM Standards, Vol 14.02.3Available from FDA, 5600 Fishers Lane, Rockville, MD 20857.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.4.4.2 Calor
15、imetric detection limit and quantitation limit aredetermined from the standard deviation of the blank determi-nation.4.4.3 Calorimetric repeatability is determined from the re-peatability measurement of the three (or more) specimens.5. Significance and Use5.1 This method may be used to determine and
16、 validate theperformance of a particular DSC apparatus.5.2 This method may be used to determine and validate theperformance of a particular method based upon a DSC enthal-pic measurement.5.3 This method may be used to determine the repeatabilityof a particular apparatus, operator or laboratory.5.4 T
17、his method may be used for specification and regula-tory compliance purposes.6. Apparatus6.1 Differential Scanning CalorimeterThe essential in-strumentation required to provide the minimum differentialscanning calorimetric capability for this method include:6.1.1 DSC Test Chamber, composed of:6.1.1.
18、1 Furnace(s), to provide uniform controlled heating orcooling of a specimen and reference to a constant temperatureor at a constant rate within the applicable temperature range ofthe test method.6.1.1.2 Temperature Sensor, to provide an indication of thespecimen temperature to readability required.6
19、.1.1.3 Differential Sensor, to detect a heat flow differencebetween the specimen and reference.6.1.1.4 A means of sustaining a test chamber environmentof an inert purge gas at a rate of 10 to 50 mL/min 6 10 %.NOTE 2Typically, 99+ % pure nitrogen (or other inert gas, such asargon or helium) is employ
20、ed when oxidation in air is a concern. Unlessthe effects of moisture are to be studied, the use of a dry purge gas isrecommended, especially for operation at subambient temperatures.6.1.2 Temperature Controller, capable of executing a spe-cific temperature program by operating the furnace(s) between
21、selected temperature limits at a rate of temperature changeconstant to 6 1 % or at an isothermal temperature constant to6 0.5C.6.1.3 Recording Device, capable of recording and display-ing the heat flow (DSC) curve as a function of time andtemperature.6.1.4 Containers, (pans, crucibles, vials, lids,
22、closures,seals, etc.) that are inert to the specimen and referencematerials and that are of suitable structural shape and integrityto contain the specimen and reference in accordance with thespecific requirements of the test method.6.2 Balance, of 100 mg or greater capacity to weighspecimens and con
23、tainers to 6 1 g.NOTE 3A balance of this high precision is required so that weighingimprecision is not part of the overall method imprecision.7. Reagents and Materials7.1 Indium Metal, 99.99+ % purity, preferably a certifiedreference material for which the melting temperature andenthalpy of fusion a
24、re known.8. Calibration and Standardization8.1 After turning the power on, allow the instrument toequilibrate for at least one hour prior to any measurements.8.2 Perform any cleaning and calibration procedures de-scribed by the manufacturer in the apparatus OperatorsManual.8.3 If not previously esta
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME22532003STANDARDMETHODFORENTHALPYMEASUREMENTVALIDATIONOFDIFFERENTIALSCANNINGCALORIMETERS 扫描 量热计

链接地址:http://www.mydoc123.com/p-530690.html