《ASTM E1698-1995(2010) Standard Practice for Testing Electrolytic Conductivity Detectors (ELCD) Used in Gas Chromatography《试验气相色谱测定中使用的电解电导率探测器(ELCD)的标准操作规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1698-1995(2010) Standard Practice for Testing Electrolytic Conductivity Detectors (ELCD) Used in Gas Chromatography《试验气相色谱测定中使用的电解电导率探测器(ELCD)的标准操作规程》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1698 95 (Reapproved 2010)Standard Practice forTesting Electrolytic Conductivity Detectors (ELCD) Used inGas Chromatography1This standard is issued under the fixed designation E1698; the number immediately following the designation indicates the year oforiginal adoption or, in the case
2、of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers testing the performance of anelectrolytic conductivity detector (ELCD) u
3、sed as the detectioncomponent of a gas chromatographic system.1.2 This practice is directly applicable to electrolytic con-ductivity detectors that perform a chemical reaction on a givensample over a nickel catalyst surface under oxidizing orreducing conditions and employ a scrubber, if needed, tore
4、move interferences, deionized solvent to dissolve the reac-tion products, and a conductivity cell to measure the electro-lytic conductivity of ionized reaction products.1.3 This practice covers the performance of the detectoritself, independently of the chromatographic column, in termsthat the analy
5、st can use to predict overall system performancewhen the detector is coupled to the column and other chro-matographic system components.1.4 For general gas chromatographic procedures, PracticeE260 should be followed except where specific changes arerecommended herein for the use of an electrolytic c
6、onductivitydetector. For definitions of gas chromatography and its variousterms see Practice E355.1.5 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.6 This standard does not purport to address all of thesafety concerns, if an
7、y, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E260 Practice for Packed Column Gas Chromatography
8、E355 Practice for Gas Chromatography Terms and Rela-tionships3. Significance and Use3.1 Although it is possible to observe and measure each ofthe several characteristics of the ELCD under different andunique conditions, in particular its different modes of selectiv-ity, it is the intent of this prac
9、tice that a complete set of detectorspecifications should be obtained at the same operating condi-tions, including geometry, gas and solvent flow rates, andtemperatures. It should be noted that to specify a detectorscapability completely, its performance should be measured atseveral sets of conditio
10、ns within the useful range of thedetector. The terms and tests described in this practice aresufficiently general so that they may be used at whateverconditions may be chosen for other reasons.3.2 Linearity and speed of response of the recorder usedshould be such that it does not distort or otherwis
11、e interferewith the performance of the detector. Effective recorder re-sponse should be sufficiently fast so that it can be neglected insensitivity of measurements. If additional amplifiers are usedbetween the detector and the final readout device, theircharacteristics should also first be establish
12、ed.4. Principles of Electrolytic Conductivity Detectors4.1 The principle components of the ELCD are representedin Fig. 1 and include: a control module, a reactor assembly,and, a cell assembly.4.1.1 The control module typically will house the detectorelectronics that monitor or control, or both, the
13、solvent flow,reaction temperatures, and the conductivity detector cell. It canbe functionally independent of the gas chromatography or, insome varieties, designed into the functional framework of thegas chromatograph. However, the reactor and cell assembliesare designed for specific models of gas ch
14、romatographs so it isimportant the proper components be assembled on the appro-priate chromatographic equipment.4.2 Fig. 2 is a block diagram representation of the GC/ELCD system. The electrolytic conductivity detector detectscompounds by pyrolyzing those compounds in a heated nickelcatalyst (housed
15、 in the reactor), removing interfering reactionproducts with a scrubber (if needed), dissolving the reaction1This practice is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.19 on Separation Science.Cu
16、rrent edition approved Nov. 1, 2010. Published November 2010. Originallyapproved in 1995. Last previous edition approved in 2005 as E1698 95 (2005).DOI: 10.1520/E1698-95R10.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For A
17、nnual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.products in a suitable solvent, and measuring the change inelectrical con
18、ductivity using a conductivity detector cell. Othersuitable non-catalystic reaction tubes can be used for moreselective response characteristics. Using the conditions setforth in this practice, halogen (Cl, Br, I, F) compounds,nitrogen compounds, and sulfur compounds can be measuredselectively, even
19、 in the presence of each other.4.3 The electrolytic conductivity detector pyrolyzes com-pounds as they elute from the chromatographic column througha hot nickel reaction tube. Halogen and nitrogen compoundsare detected under reducing conditions while sulfur compoundsare detected under oxidizing cond
20、itions. The effluent from thegas chromatographic column is combined with either hydrogen(reducing conditions) or air (oxidizing conditions) beforeentering the heated (800 to 1000C) nickel reaction tube. Thecompound is converted to small inorganic reaction productsdepending upon the reaction conditio
21、ns as shown in Table 1.4.4 Table 2 shows the chemistry and modes of selectiveresponse for the detector. Depending upon the mode ofoperation, various interfering reaction products are removed byemploying a selective gas scrubber before the product gasesreach the detector cell. In the nitrogen-specifi
22、c mode, halogenand sulfur products are removed by reaction with a causticscrubber. In the sulfur-specific mode, halogen products areremoved by a silver thread (or wire) scrubber. No scrubber isrequired for halogen mode operation.4.5 The reaction products pass to the conductivity cellwhere they are c
23、ombined with the solvent. The followingsolvents are typically used for normal operation in eachindicated mode. Other solvents may be used to providechanges in selectivity and sensitivity (see 6.7):Model SolventHalogen 1-PropanolSulfur 100 % MethanolNitrogen 10 %t-Butyl Alcohol/90 % Water4.6 The incr
24、ease in electrical conductivity of the solvent asa result of the introduction of the reaction products is measuredby the sensing electrodes in the conductivity cell. The solventpasses through the cell after being deionized through an ionexchange resin bed located between the conductivity cell andsol
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME169819952010STANDARDPRACTICEFORTESTINGELECTROLYTICCONDUCTIVITYDETECTORSELCDUSEDINGASCHROMATOGRAPHY

链接地址:http://www.mydoc123.com/p-529266.html