ASTM E1582-2014 Standard Practice for Calibration of Temperature Scale for Thermogravimetry《热解重量测定用温标校准的标准实施规程》.pdf
《ASTM E1582-2014 Standard Practice for Calibration of Temperature Scale for Thermogravimetry《热解重量测定用温标校准的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1582-2014 Standard Practice for Calibration of Temperature Scale for Thermogravimetry《热解重量测定用温标校准的标准实施规程》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1582 14Standard Practice forCalibration of Temperature Scale for Thermogravimetry1This standard is issued under the fixed designation E1582; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A
2、 number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice describe the temperature calibration ofthermogravimetric analyzers over the temperature range from25 to 1500C and is app
3、licable to commercial and custom-builtapparatus. This calibration may be accomplished by the use ofeither melting point standards or magnetic transition standards.1.2 The mass change curve in thermogravimetry resultsfrom a number of influences, some of which are characteristicof the specimen holder
4、assembly and atmosphere rather thanthe specimen. The variations from instrument to instrumentoccur in the point of measurement of the temperature, thenature of the material, its size and packing, the geometry andcomposition of the specimen container, the geometry anddesign of the furnace, and the ac
5、curacy and sensitivity of thetemperature sensor and displaying scales. These all contributeto differences in measured temperatures, which may exceed20C. In addition, some sample holder assemblies will showvariations of measured temperature with sample size orheating/cooling rate, or both. Since it i
6、s neither practical noradvisable to standardize sample holders or thermobalancegeometries, instruments may be calibrated by measurement ofthe deviation of a melting or magnetic (Curie Point) transitiontemperature from the standard reference temperature. Thisdeviation can be applied as a correction t
7、erm to subsequentmeasurements.1.3 This practice assumes that the indicated temperature ofthe instrument is linear over the range defined by a two-pointcalibration and that this linearity has been verified. These twocalibration temperatures should be as close to the experimentalmeasurements to be mad
8、e as possible.1.4 This practice describes two procedures for temperaturecalibration of thermogravimetric analyzers using any typebalance. Procedure A uses melting point standards for calibra-tion. Procedure B uses magnetic transition standards forcalibration.NOTE 1Since all electronic data treatment
9、s are not equivalent, the usershall verify equivalency prior to use.1.5 The data generated by these procedures can be used tocorrect the temperature scale of the instrument by either apositive or negative amount using either a two-point tempera-ture calibration procedure or a multi-point temperature
10、 calibra-tion with best line fit for the generated data.NOTE 2A single-point calibration may be used where this is the onlyprocedure possible or practical. The use of a single-point procedure is notrecommended.1.6 The values stated in SI units are to be regarded asstandard. No other units of measure
11、ment are included in thisstandard.1.7 This practice is related to ISO 11358 but providesinformation and methods not found in ISO 11358.1.8 This standard does not purport to address all of thesafety problems, if any, associated with its use. It is theresponsibility of the user of this standard to est
12、ablish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E473 Terminology Relating to Thermal Analysis and Rhe-ologyE691 Practice for Conducting an Interlaboratory Study toDetermine the Precisio
13、n of a Test MethodE967 Test Method for Temperature Calibration of Differen-tial Scanning Calorimeters and Differential Thermal Ana-lyzersE1142 Terminology Relating to Thermophysical Properties2.2 Other Standards:ISO 11358 Plastics Part 4: Thermogravimetry (TG) ofPolymers General Principles33. Termin
14、ology3.1 DefinitionsTechnical terms used in this document aredefined in Terminologies E473 and E1142.1This practice is under the jurisdiction of ASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on Calo-rimetry and Mass Loss.Current edition approved Ma
15、y 1, 2014. Published June 2014. Originallyapproved in 1993. Last previous edition approved in 2010 as E1582 10. DOI:10.1520/E1582-14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer service at serviceastm.org. For Annual Book of ASTMStandards volume infor
16、mation, refer to the standards Document Summary page onthe ASTM website.3Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. U
17、nited States13.1.1 magnetic reference temperaturethe observed tem-perature at which a change in the magnetic properties of amaterial in a magnetic field produces an apparent mass change.This temperature is read from the dynamic TG curve as thepoint of intersection of the extrapolated higher temperat
18、ureportion of the base line with a tangent drawn to the point ofgreatest slope of apparent mass-change curve. This tempera-ture most closely represents the Curie Point, that point on themass change curve where the magnetic effect of the standardmaterial has disappeared completely (see Fig. 1).3.1.1.
19、1 DiscussionThe position of the magnet and thedesign of the instrument will affect the direction of the masschange.4. Summary of Practice4.1 This practice provides a set of different procedures sincethermogravimetric apparatus is often of significantly differingdesign.4.2 Calibration of Analyzers Us
20、ing Magnetic TransitionStandards:4.2.1 In this procedure, the apparent mass change of one ormore of the magnetic transition standards is obtained under thenormal operating conditions of the instrument. The extrapo-lated endpoint temperature (see Fig. 2) is determined andcompared with the established
21、 transition temperature for thematerial. The difference provides an adjustment or calibrationthat may be applied to the temperature scale of the instrument.4.2.2 The apparent mass change of the magnetic transitionmaterials is caused by the magnetic to nonmagnetic transitionin the presence of a magne
22、tic field.4.3 Calibration of Analyzers That Have SimultaneousThermogravimetry-Differential Scanning Calorimeter orThermogravimetry-Differential Thermal Analysis CapabilityThese instruments may be calibrated using melting temperaturestandards following Practice E967.5. Significance and Use5.1 Thermog
23、ravimetric analyzers are used to characterize abroad range of materials. In most cases, one of the desiredvalues to be assigned in thermogravimetric measurements isthe temperature at which significant changes in specimen massoccur. Therefore, the temperature axis (abscissa) of allapparent-mass-chang
24、e curves must be calibrated accurately,either by direct reading of a temperature sensor, or by adjustingthe programmer temperature to match the actual temperatureover the temperature range of interest. In the latter case, this isaccomplished by the use of either melting point or magnetictransition s
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME15822014STANDARDPRACTICEFORCALIBRATIONOFTEMPERATURESCALEFORTHERMOGRAVIMETRY 重量 测定 温标 校准 标准 实施 规程

链接地址:http://www.mydoc123.com/p-528983.html