ASTM E1582-2004 Standard Practice for Calibration of Temperature Scale for Thermogravimetry《热解重量测定用温标校准的标准实施规程》.pdf
《ASTM E1582-2004 Standard Practice for Calibration of Temperature Scale for Thermogravimetry《热解重量测定用温标校准的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1582-2004 Standard Practice for Calibration of Temperature Scale for Thermogravimetry《热解重量测定用温标校准的标准实施规程》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1582 04Standard Practice forCalibration of Temperature Scale for Thermogravimetry1This standard is issued under the fixed designation E 1582; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision.
2、 A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice describe the temperature calibration ofthermogravimetric analyzers over the temperature range from25 to 1500 C and is
3、 applicable to commercial and custom-built apparatus. This calibration may be accomplished by theuse of either melting point standards or magnetic transitionstandards.1.2 The mass change curve in thermogravimetry resultsfrom a number of influences, some of which are characteristicof the specimen hol
4、der assembly and atmosphere rather thanthe specimen. The variations from instrument to instrumentoccur in the point of measurement of the temperature, thenature of the material, its size and packing, the geometry andcomposition of the specimen container, the geometry anddesign of the furnace, and th
5、e accuracy and sensitivity of thetemperature sensor and displaying scales. These all contributeto differences in measured temperatures, which may exceed20 C. In addition, some sample holder assemblies will showvariations of measured temperature with sample size orheating/cooling rate, or both. Since
6、 it is neither practical noradvisable to standardize sample holders or thermobalancegeometries, instruments may be calibrated by measurement ofthe deviation of a melting or magnetic (Curie Point) transitiontemperature from the standard reference temperature. Thisdeviation can be applied as a correct
7、ion term to subsequentmeasurements.1.3 This practice assumes that the indicated temperature ofthe instrument is linear over the range defined by a two-pointcalibration and that this linearity has been verified. These twocalibration temperatures should be as close to the experimentalmeasurements to b
8、e made as possible.1.4 This practice describes three procedures for temperaturecalibration of thermogravimetric analyzers using any typebalance. Procedures A and B use melting point standards withvertical and horizontal balances. Procedure C uses magnetictransition standards for calibration. Procedu
9、re A is designedspecifically for use with horizontal-type balances using exter-nal furnaces. Procedure B is designed specifically for use withvertical hang-down balances using either internal or externalfurnaces. No procedure is restricted to the use of the furnacetype described in that procedure.1.
10、5 Computer or electronic-based instruments, techniques,or data treatment equivalent to this procedure may be used.NOTE 1Since all electronic data treatments are not equivalent, theuser shall verify equivalency prior to use.1.6 The data generated by these procedures can be used tocorrect the temperat
11、ure scale of the instrument by either apositive or negative amount using either a two-point tempera-ture calibration procedure or a multi-point temperature calibra-tion with best line fit for the generated data.NOTE 2A single-point calibration may be used where this is the onlyprocedure possible or
12、practical. The use of a single-point procedure is notrecommended.1.7 SI units are standard.1.8 This practice is related to ISO 11358 but providesinformation and methods not found in ISO 11358 .1.9 This standard does not purport to address all of thesafety problems, if any, associated with its use. I
13、t is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:E 473 Terminology Relating to Thermal Analysis2E 691 Practice for Conducting an
14、Interlaboratory Study toDetermine the Precision of a Test Method2E 967 Practice for Temperature Calibration of DifferentialScanning Calorimeters and Differential Thermal Analyz-ers2E 1142 Terminology Relating to Thermophysical Proper-ties22.2 Other Standards:ISO 11358 Thermogravimetry (TG) of Polyme
15、rs GeneralPrinciples31This practice is under the jurisdiction of ASTM Committee E 37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.01 on TestMethods and Recommended Practices.Current edition approved May 1, 2004. Published June 2004. Originallyapproved in 1993. Last pre
16、vious edition approved in 2000 as E 1582 00.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from A
17、merican National Standardards Institute, 11 West 42nd Street,13th Floor, New York, NY 10036.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3. Terminology3.1 DefinitionsTechnical terms used in this document aredefined in Terminology
18、E 473 and E 1142.3.1.1 magnetic reference temperaturethe observed tem-perature at which a change in the magnetic properties of amaterial in a magnetic field produces an apparent mass change.This temperature is read from the dynamic TG curve as thepoint of intersection of the extrapolated higher temp
19、eratureportion of the base line with a tangent drawn to the point ofgreatest slope of apparent mass-change curve. This tempera-ture most closely represents the Curie Point, that point on themass change curve where the magnetic effect of the standardmaterial has disappeared completely (see Fig. 1).NO
20、TE 3The position of the magnet and the design of the instrumentwill affect the direction of the mass change.4. Summary of Practice4.1 This practice provides a set of different procedures sincethermogravimetric apparatus is often of significantly differingdesign.4.2 Calibration of Analyzers Using Mel
21、ting PointStandardsThe calibration material is heated at a controlledrate in a controlled atmosphere through its melting region. Thetemperature of the standard is monitored and recorded continu-ously. In this practice, a small platinum mass is suspendedwithin a thermogravimetric analyzer specimen bo
22、at or panfrom a fusible link of the standard calibration material. As thestandard specimen is heated through the melting region, theplatinum mass is released. The mass is either caught in thespecimen boat or pan, producing an “action/reaction” blip onthe thermal curve, or is allowed to drop through
23、a hole in thebottom of the specimen boat or pan, producing a sharp,discontinuous mass loss. These events may be used to calibratethe thermogravimetric analyzer for the experimental conditionsused.4.3 Calibration of Analyzers Using Magnetic TransitionStandards:4.3.1 In this procedure, the apparent ma
24、ss change of one ormore of the magnetic transition standards is obtained under thenormal operating conditions of the instrument. The extrapo-lated endpoint temperature, (see Fig. 1), is determined andcompared with the established transition temperature for thematerial. The difference provides an adj
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME15822004STANDARDPRACTICEFORCALIBRATIONOFTEMPERATURESCALEFORTHERMOGRAVIMETRY 重量 测定 温标 校准 标准 实施 规程

链接地址:http://www.mydoc123.com/p-528982.html