ASTM D7315-2017 9375 Standard Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode《测定静态模式下混浊度大于1浊度单位(TU)的标准试验方法》.pdf
《ASTM D7315-2017 9375 Standard Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode《测定静态模式下混浊度大于1浊度单位(TU)的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7315-2017 9375 Standard Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode《测定静态模式下混浊度大于1浊度单位(TU)的标准试验方法》.pdf(22页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7315 17Standard Test Method forDetermination of Turbidity Above 1 Turbidity Unit (TU) inStatic Mode1This standard is issued under the fixed designation D7315; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year o
2、f last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the static determination ofturbidity in water. Static refers to a sample that is removedfro
3、m its source and tested in an isolated instrument. (SeeSection 4.)1.2 This test method is applicable to the measurement ofturbidities greater than 1.0 turbidity unit (TU). The upper endof the measurement range was left undefined because differenttechnologies described in this test method can cover v
4、erydifferent ranges. The round robin study covered the range of04000 turbidity units because instrument verification in thisrange can typically be covered by standards that can beconsistently reproduced.1.3 Many of the turbidity units and instrument designscovered in this test method are numerically
5、 equivalent incalibration when a common calibration standard is appliedacross those designs listed in Table 1. Measurement of acommon calibration standard of a defined value will alsoproduce equivalent results across these technologies.1.3.1 In this test method calibration standards are oftendefined
6、 in NTU values, but the other assigned turbidity units,such as those in Table 1 are equivalent. For example,a1NTUformazin standard is alsoa1FNU,a1FAU,a1BU,andsoforth.1.4 This test method does not purport to cover all availabletechnologies for high-level turbidity measurement.1.5 This test method was
7、 tested on different natural watersand wastewater, and with standards that will serve as surro-gates to samples. It is the users responsibility to ensure thevalidity of this test method for waters of untested matrices.1.6 Depending on the constituents within a high-levelsample, the proposed sample p
8、reparation and measurementmethods may or may not be applicable. Those samples with thehighest particle densities typically prove to be the most difficultto measure. In these cases, and alternative measurementmethod such as the process monitoring method can be consid-ered.1.7 This standard does not p
9、urport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Refer to the MSDSsfor all chemicals used i
10、n this procedure.1.8 This international standard was developed in accor-dance with internationally recognized principles on standard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization Tec
11、hnicalBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating to WaterD1193 Specification for Reagent WaterD2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on WaterD4411 Guide for Sampling Fluvial Sedime
12、nt in MotionD5847 Practice for Writing Quality Control Specificationsfor Standard Test Methods for Water AnalysisD6855 Test Method for Determination of Turbidity Below 5NTU in Static ModeE691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test Method2.2 Other Referen
13、ced Standards:U.S. EPA Method 180.1 Methods for Chemical Analysis ofWater and Wastes, Turbidity3ISO 7027 Water QualityDetermination of Turbidity41This test method is under the jurisdiction of ASTM Committee D19 on Waterand is the direct responsibility of Subcommittee D19.07 on Sediments,Geomorpholog
14、y, and Open-Channel Flow.Current edition approved July 1, 2017. Published July 2017. Originally approvedin 2007. Last previous edition approved in 2012 as D7315 12. DOI: 10.1520/D7315-17.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at servicea
15、stm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from United States Environmental ProtectionAgency (EPA), WilliamJefferson Clinton Bldg., 1200 Pennsylvania Ave., NW, Washington, DC 20460,http:/www.epa.gov.4Availa
16、ble from International Organization for Standardization (ISO), ISOCentral Secretariat, BIBC II, Chemin de Blandonnet 8, CP 401, 1214 Vernier,Geneva, Switzerland, http:/www.iso.org.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO
17、 Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued b
18、y the World Trade Organization Technical Barriers to Trade (TBT) Committee.1USGS National Field Manual for the Collection of WaterQuality Data53. Terminology3.1 Definitions:3.1.1 For definitions of terms used in this standard, refer toTerminology D1129.3.2 Definitions of Terms Specific to This Stand
19、ard:3.2.1 attenuation, vthe amount of incident light that isscattered and absorbed before reaching a detector, which isgeometrically centered at 0 relative to the centerline of theincident light beam.3.2.1.1 DiscussionAttenuation is inversely proportional totransmitted signal.Attenuated Turbidity 5
20、Absorbed Light1Scattered LightThe application of attenuation in this test method is as adistinct means of measuring turbidity. When measuring inthe FAU or AU mode, the turbidity value is a combinationof scattered (attenuated) plus absorbed light. The scattered5Available from United Stated Geological
21、 Survey (USGS), 12201 Sunrise ValleyDrive, Reston, VA 20192, http:/www.usgs.gov.TABLE 1 Summary of Known Instrument Designs, Applications, Ranges, and Reporting UnitsDesign andReporting UnitProminent Application Key Design FeaturesTypicalInstrument RangeSuggestedApplication RangesNephelometric non-r
22、atio(NTU)White light turbidimeters. Complywith U.S. EPA Method 180.1 forlow level turbidity monitoring.Detector centered at 90 relativeto the incident light beam. Usesa white light spectral source.0.040 0.040 RegulatoryRatio White Light turbidime-ters (NTRU)Complies with ISWTR regulationsand Standar
23、d Method 2130B.Can be used for both low andhigh level measurement.Used a white light spectralsource. Primary detector cen-tered at 90. Other detectorslocated at other angles. An in-strument algorithm uses a com-bination of detector readings togenerate the turbidity reading.010 000 040 Regulatory010
24、000 otherNephelometric, near-IRturbidimeters, non-ratiometric (FNU)Complies with ISO 7027. Thewavelength is less susceptibleto color interferences. Appli-cable for samples with colorand good for low level monitor-ing.Detector centered at 90 relativeto the incident light beam. Usesa near-IR (780900 n
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD731520179375STANDARDTESTMETHODFORDETERMINATIONOFTURBIDITYABOVE1TURBIDITYUNITTUINSTATICMODE 测定 静态

链接地址:http://www.mydoc123.com/p-525117.html