《ASTM D7315-2012 red 2500 Standard Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode《静态模式下测定混浊度大于1浊度单位 (TU) 的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7315-2012 red 2500 Standard Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode《静态模式下测定混浊度大于1浊度单位 (TU) 的标准试验方法》.pdf(27页珍藏版)》请在麦多课文档分享上搜索。
1、Designation:D731507aD731512 Standard Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode 1 This standard is issued under the xed designation D7315; the number immediately following the designation indicates the year of original adoption or, in the case of revision,
2、the year of last revision.Anumber in parentheses indicates the year of last reapproval.A superscript epsilon () indicates an editorial change since the last revision or reapproval. 1. Scope 1.1 This test method covers the static determination of turbidity in water. Static refers to a sample that is
3、removed from its source and tested in an isolated instrument. (See Section 4.) 1.2 This test method is applicable to the measurement of turbidities greater than 1.0 turbidity unit (TU). The upper end of the measurementrangewasleftundenedbecausedifferenttechnologiesdescribedinthistestmethodcancoverve
4、rydifferentranges. The round robin study covered the range of 04000 turbidity units because instrument verication in this range can typically be covered by standards that can be consistently reproduced. 1.3 Many of the turbidity units and instrument designs covered in this test method are numericall
5、y equivalent in calibration when a common calibration standard is applied across those designs listed in Table 1. Measurement of a common calibration standard of a dened value will also produce equivalent results across these technologies. 1.3.1 In this test method calibration standards are often de
6、ned in NTU values, but the other assigned turbidity units, such as those in Table 1 are equivalent. For example, a 1 NTU formazin standard is also a 1 FNU, a 1 FAU, a 1 BU, and so forth. 1.4 This test method does not purport to cover all available technologies for high-level turbidity measurement. 1
7、.5 This test method was tested on different natural waters and wastewater, and with standards that will serve as surrogates to samples. It is the users responsibility to ensure the validity of this test method for waters of untested matrices. 1.6 Dependingontheconstituentswithinahigh-levelsample,the
8、proposedsamplepreparationandmeasurementmethodsmay or may not be applicable. Those samples with the highest particle densities typically prove to be the most difficult to measure. In these cases, and alternative measurement method such as the process monitoring method can be considered. 1.7 This stan
9、dard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Refer to the MSDSs for all
10、chemicals used in this procedure. 2. Referenced Documents 2.1 ASTM Standards: 2 D1129Terminology Relating to Water D1193Specication for Reagent Water D1889Test Method for Turbidity of Water (Withdrawn 2007) 3 D2777Practice for Determination of Precision and Bias of Applicable Test Methods of Committ
11、ee D19 on Water D4411Guide for Sampling Fluvial Sediment in Motion D5847Practice for Writing Quality Control Specications for Standard Test Methods for Water Analysis D6855Test Method for Determination of Turbidity Below 5 NTU in Static Mode E691Practice for Conducting an Interlaboratory Study to De
12、termine the Precision of a Test Method 1 This test method is under the jurisdiction ofASTM Committee D19 on Water and is the direct responsibility of Subcommittee D19.07 on Sediments, Geomorphology, and Open-Channel Flow. Current edition approved Aug. 1, 2007June 1, 2012. Published August 2007 June
13、2012. Originally approved in 2007. Last previous edition approved in 2007 as D731507.D731507A. DOI: 10.1520/D7315-07A.10.1520/D7315-12. 2 ForreferencedASTMstandards,visittheASTMwebsite,www.astm.org,orcontactASTMCustomerServiceatserviceastm.org.ForAnnualBookofASTMStandards volume information, refer t
14、o the standards Document Summary page on the ASTM website. 3 The last approved version of this historical standard is referenced on www.astm.org. This document is not anASTM standard and is intended only to provide the user of anASTM standard an indication of what changes have been made to the previ
15、ous version. Because it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version of the standard as published by ASTM is to be considered the official document. Copyright ASTM
16、International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA19428-2959. United States 1TABLE 1 Summary of Known Instrument Designs, Applications, Ranges, and Reporting Units Design and Reporting Unit ProminentApplication Key Design Features Typical Instrument Range Suggested Application
17、Ranges Nephelometric non-ratio (NTU) White light turbidimeters. Comply with USEPAMethod 180.1 for low level turbidity monitoring. Detector centered at 90 relative to the incident light beam. Uses a white light spectral source. 0.040 0.040 Regulatory Ratio White Light turbidimeters (NTRU) Complies wi
18、th ISWTR regulations and Standard Method 2130B. Can be used for both low and high level measurement. Used a white light spectral source. Primary detector centered at 90. Other detectors located at other angles.An instrument algorithm uses a combination of detector readings to generate the turbidity
19、reading. 010000 040 Regulatory 010000 other Nephelometric, near-IR turbidimeters, non-ratiometric (FNU) Complies with ISO 7027. The wavelength is less susceptible to color interferences.Applicable for samples with color and good for low level monitoring. Detector centered at 90 relative to the incid
20、ent light beam. Uses a near-IR (780900 nm) monochromatic light source. 01000 040 Regulatory (non- US) 01000 other Nephelometric near-IR turbidimeters, ratio metric (FNRU) Complies with ISO 7027. Applicable for samples with high levels of color and for monitoring to high turbidity levels. Uses a near
21、-IR monochromatic light source (780900 nm). Primary detector centered at 90. Other detectors located at other angles.An instrument algorithm uses a combination of detector readings to generate the turbidity reading. 010000 040 Regulatory 010000 other Surface Scatter Turbidimeters (NTU) Turbidity is
22、determined through light scatter from or near the surface of a sample. Detector centered at 90 relative to the incident light beam. Uses a white light spectral source. 1010000 1010000 Formazin Back Scatter (FBU) Not applicable for regulatory purposes. Best applied to high turbidity samples. Backscat
23、ter is common with but not all only probe technology and is best applied in higher turbidity samples. Uses a near-IR monochromatic light source in the 780900 nm range. Detector geometry is 90 relative to the incident light beam. 10010000+ 10010000 Formazin Back Scatter (FBU) Not applicable for regul
24、atory purposes. Best applied to high turbidity samples. Backscatter is common with but not all only probe technology and is best applied in higher turbidity samples. Uses a near-IR monochromatic light source in the 780900 nm range. Detector geometry is between 90 and 180 relative to the incident lig
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD73152012RED2500STANDARDTESTMETHODFORDETERMINATIONOFTURBIDITYABOVE1TURBIDITYUNITTUINSTATICMODE 静态

链接地址:http://www.mydoc123.com/p-525116.html