《ASTM D7315-2007a Standard Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode《静态模式下测定混浊度大于1浊度单位(TU)的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7315-2007a Standard Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode《静态模式下测定混浊度大于1浊度单位(TU)的标准试验方法》.pdf(21页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 7315 07aStandard Test Method forDetermination of Turbidity Above 1 Turbidity Unit (TU) inStatic Mode1This standard is issued under the fixed designation D 7315; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the yea
2、r of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the static determination ofturbidity in water. Static refers to a sample that is removed
3、from its source and tested in an isolated instrument. (SeeSection 4.)1.2 This test method is applicable to the measurement ofturbidities greater than 1.0 turbidity unit (TU). The upper endof the measurement range was left undefined because differenttechnologies described in this test method can cove
4、r verydifferent ranges. The round robin study covered the range of04000 turbidity units because instrument verification in thisrange can typically be covered by standards that can beconsistently reproduced.1.3 Many of the turbidity units and instrument designscovered in this test method are numerica
5、lly equivalent incalibration when a common calibration standard is appliedacross those designs listed in Table 1. Measurement of acommon calibration standard of a defined value will alsoproduce equivalent results across these technologies.1.3.1 In this test method calibration standards are oftendefi
6、ned in NTU values, but the other assigned turbidity units,such as those in Table 1 are equivalent. For example,a1NTUformazin standard is alsoa1FNU,a1FAU,a1BU,andsoforth.1.4 This test method does not purport to cover all availabletechnologies for high-level turbidity measurement.1.5 This test method
7、was tested on different natural watersand wastewater, and with standards that will serve as surro-gates to samples. It is the users responsibility to ensure thevalidity of this test method for waters of untested matrices.1.6 Depending on the constituents within a high-levelsample, the proposed sampl
8、e preparation and measurementmethods may or may not be applicable. Those samples with thehighest particle densities typically prove to be the most difficultto measure. In these cases, and alternative measurementmethod such as the process monitoring method can be consid-ered.1.7 This standard does no
9、t purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Refer to the MSDSsfor all chemicals use
10、d in this procedure.2. Referenced Documents2.1 ASTM Standards:2D 1129 Terminology Relating to WaterD 1193 Specification for Reagent WaterD 1889 Test Method for Turbidity of Water3D 2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on WaterD 4411 Guide f
11、or Sampling Fluvial Sediment in MotionD 5847 Practice for Writing Quality Control Specificationsfor Standard Test Methods for Water AnalysisD 6855 Test Method for Determination of Turbidity Below5 NTU in Static ModeE 691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a
12、 Test Method2.2 Other Referenced Standards:USEPA Method 180.1 Methods for Chemical Analysis ofWater and Wastes, Turbidity4ISO 7027 Water QualityDetermination of Turbidity5United States Geological Survey (USGS) National FieldManual for the Collection of Water Quality Data63. Terminology3.1 Definition
13、sFor definitions of terms used in this testmethod refer to Terminology D 1129.3.2 Definitions of Terms Specific to This Standard:1This test method is under the jurisdiction of ASTM Committee D19 on Waterand is the direct responsibility of Subcommittee D19.07 on Sediments, Geomor-phology, and Open-Ch
14、annel Flow.Current edition approved Aug. 1, 2007. Published August 2007. Originallyapproved in 2007. Last previous edition approved in 2007 as D 7315 07.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMSt
15、andards volume information, refer to the standards Document Summary page onthe ASTM website.3Withdrawn.4Available from United States Environmental Protection Association (EPA),Ariel Rios Bldg., 1200 Pennsylvania Ave., NW, Washington, DC 20460, http:/www.epa.gov.5Available from International Organiza
16、tion for Standardization (ISO), 1 rue deVaremb, Case postale 56, CH-1211, Geneva 20, Switzerland, http:/www.iso.ch.6Available from United Stated Geological Survey (USGS), 12201 Sunrise ValleyDrive, Reston, VA 20192, http:/www.usgs.gov.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700
17、, West Conshohocken, PA 19428-2959, United States.3.2.1 turbidityan expression of the optical properties of asample that causes light rays to be scattered and absorbedrather than transmitted in straight lines through the sample.Turbidity of water is caused by the presence of suspended anddissolved m
18、atter such as clay, silt, finely divided organicmatter, plankton, other microscopic organisms, organic acids,and dyes.3.2.2 turbidimeteran instrument that measures light scat-ter, attenuation, or both in a sample and quantitatively convertsthe light scatter, the attenuation, or both to a displayed v
19、alue.The location and type and number of detectors used will dictatethe relative sensitivity for a typical technology. See Table 1 forexamples of designs.3.2.3 reference turbidity standarda standard that is syn-thesized reproducibly from traceable raw materials by a skilledanalyst. All other standar
20、ds are traced back to this standard.The reference standard for turbidity is formazin (see 9.2.2).3.2.4 calibration turbidity standarda turbidity standardthat is traceable and equivalent to the reference turbiditystandard to within defined accuracy, including commerciallyprepared 4000 NTU Formazin, s
21、tabilized formazin (see 9.2.3),and styrenedivinylbenzene (SDVB) (see 9.2.4). These stan-dards may be used to calibrate the instrument.NOTE 1Calibration standards may be instrument design specific.NOTE 2Calibration standards that exceed 10 000 turbidity units arecommercially available.TABLE 1 Summary
22、 of Known Instrument Designs, Applications, Ranges, and Reporting UnitsDesign andReporting UnitProminent Application Key Design FeaturesTypicalInstrument RangeSuggestedApplication RangesNephelometric non-ratio(NTU)White light turbidimeters. Complywith USEPA Method 180.1 for lowlevel turbidity monito
23、ring.Detector centered at 90 relativeto the incident light beam. Uses awhite light spectral source.0.040 0.040 RegulatoryRatio White Lightturbidimeters (NTRU)Complies with ISWTR regulationsand Standard Method 2130B. Canbe used for both low and highlevel measurement.Used a white light spectralsource.
24、 Primary detector centeredat 90. Other detectors located atother angles. An instrumentalgorithm uses a combination ofdetector readings to generate theturbidity reading.010 000 040 Regulatory010 000 otherNephelometric, near-IRturbidimeters, non-ratiometric(FNU)Complies with ISO 7027. Thewavelength is
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD73152007ASTANDARDTESTMETHODFORDETERMINATIONOFTURBIDITYABOVE1TURBIDITYUNITTUINSTATICMODE 静态 模式 测定

链接地址:http://www.mydoc123.com/p-525115.html