《ASTM D7094-2017a red 0000 Standard Test Method for Flash Point by Modified Continuously Closed Cup (MCCCFP) Tester《用改进的连续封闭杯(MCCCFP)测试仪测定闪点的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7094-2017a red 0000 Standard Test Method for Flash Point by Modified Continuously Closed Cup (MCCCFP) Tester《用改进的连续封闭杯(MCCCFP)测试仪测定闪点的标准试验方法》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7094 17D7094 17aStandard Test Method forFlash Point by Modified Continuously Closed Cup(MCCCFP) Tester1This standard is issued under the fixed designation D7094; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the yea
2、r of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of the flash point of fuels, fuels including diesel/biodiesel blends,
3、lube oils,solvents, and other liquids by a continuously closed cup tester utilizing a specimen size of 2 mL, cup size of 7 mL, with a heatingrate of 2.5 C per minute.1.1.1 Apparatus requiring a specimen size of 1 mL, cup size of 4 mL, and a heating rate of 5.5 C per minute must be runaccording to Te
4、st Method D6450.1.2 This flash point test method is a dynamic method and depends on definite rates of temperature increase. It is one of the manyflash point test methods available and every flash point test method, including this one, is an empirical method.NOTE 1Flash point values are not a constan
5、t physical chemical property of materials tested. They are a function of the apparatus design, the conditionof the apparatus used, and the operational procedure carried out. Flash point can, therefore, only be defined in terms of a standard test method and nogeneral valid correlation can be guarante
6、ed between results obtained by different test methods or where different test apparatus is specified.1.3 This test method utilizes a closed but unsealed cup with air injected into the test chamber.1.4 This test method is suitable for testing samples with a flash point from 35 C to 225 C.NOTE 2Flash
7、point determinations below 35 C and above 225 C may be performed; however, the precision has not been determined below andabove these temperatures.1.5 If the users specification requires a defined flash point method other than this method, neither this method nor any othertest method should be subst
8、ituted for the prescribed test method without obtaining comparative data and an agreement from thespecifier.1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.Temperatures are in degrees Celsius, pressure in kilo-Pascals.1.7
9、This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine theapplicability of regulatory limitations prior
10、 to use. For specific warning statements, see 7.2, 8.5, and 10.1.2.1.8 This international standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Development of International Standards, Guides and Recommendat
11、ions issuedby the World Trade Organization Technical Barriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2D93 Test Methods for Flash Point by Pensky-Martens Closed Cup TesterD4057 Practice for Manual Sampling of Petroleum and Petroleum ProductsD4177 Practice for Automatic Sam
12、pling of Petroleum and Petroleum ProductsD6450 Test Method for Flash Point by Continuously Closed Cup (CCCFP) TesterD6708 Practice for Statistical Assessment and Improvement of Expected Agreement Between Two Test Methods that Purportto Measure the Same Property of a MaterialE300 Practice for Samplin
13、g Industrial Chemicals1 This test method is under the jurisdiction ofASTM Committee D02 on Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of SubcommitteeD02.08 on Volatility.Current edition approved May 1, 2017Dec. 1, 2017. Published June 2017February 2018. Origina
14、lly approved in 2004. Last previous edition approved in 20162017 asD7094 16.D7094 17. DOI: 10.1520/D7094-17.10.1520/D7094-17A.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information,
15、refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all
16、 changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 10
17、0 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States12.2 ISO Standards:3ISO Guide 34 Quality Systems Guidelines for the Production of Reference MaterialsISO Guide 35 Certification of Reference MaterialsGeneral and Statistical Principles3. Terminology3.1 Definitions:3.1.1
18、 dynamic, adjthe condition where the vapor above the test specimen and the test specimen are not in temperatureequilibrium at the time at which the ignition source is applied.3.1.2 flash point, nin flash point test methods, the lowest temperature of the test specimen, adjusted to account for variati
19、onsin atmospheric pressure from 101.3 kPa, at which application of an ignition source causes the vapors of the test specimen to igniteunder specified conditions of test.3.1.2.1 DiscussionFor the purpose of this test method, the test specimen is deemed to have flashed when the hot flame of the ignite
20、d vapor causesan instantaneous pressure increase of at least 20 kPa inside the closed measuring chamber.4. Summary of Test Method4.1 The lid of the test chamber is regulated to a temperature at least 18 C below the expected flash point. A 2 mL 6 0.2 mLtest specimen of a sample is introduced into the
21、 sample cup. Both specimen and cup are at a temperature at least 18 C below theexpected flash point; cooled if necessary. The cup is then raised and pressed onto the lid of specified dimensions to form thecontinuously closed but unsealed test chamber with an overall volume of 7.0 mL 6 0.3 mL.4.2 Aft
22、er closing the test chamber, the temperatures of the test specimen and the regulated lid are allowed to equilibrate to within1 C. Then the lid is heated at a prescribed, constant rate. For the flash tests, a high voltage arc of defined energy is dischargedinside the test chamber at regular intervals
23、. After each ignition, a variable amount of air (see Table 1) is introduced into the testchamber to provide the necessary oxygen for the next flash test. The pressure inside the continuously closed but unsealed testchamber remains at ambient barometric pressure except for the short time during the a
24、ir introduction and at a flash point.4.3 After each arc application, the instantaneous pressure increase above the ambient barometric pressure inside the testchamber is monitored.When the pressure increase exceeds 20 kPa, the temperature at that point is recorded as the uncorrected flashpoint, which
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD70942017ARED0000STANDARDTESTMETHODFORFLASHPOINTBYMODIFIEDCONTINUOUSLYCLOSEDCUPMCCCFPTESTER 改进 连续

链接地址:http://www.mydoc123.com/p-524476.html