《ASTM D3871-1984(2003) Standard Test Method for Purgeable Organic Compounds in Water Using Headspace Sampling《用筒底间隙取样技术对水中可提纯有机化合物的试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D3871-1984(2003) Standard Test Method for Purgeable Organic Compounds in Water Using Headspace Sampling《用筒底间隙取样技术对水中可提纯有机化合物的试验方法》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 3871 84 (Reapproved 2003)Standard Test Method forPurgeable Organic Compounds in Water Using HeadspaceSampling1This standard is issued under the fixed designation D 3871; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision
2、, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of mostpurgeable organic compounds that boil below 200C and a
3、reless than 2 % soluble in water. It covers the low g/L to lowmg/L concentration range (see Section 15 and Appendix X1).1.2 This test method was developed for the analysis ofdrinking water. It is also applicable to many environmental andwaste waters when validation, consisting of recovering knowncon
4、centrations of compounds of interest added to representa-tive matrices, is included.1.3 Volatile organic compounds in water at concentrationsabove 1000 g/L may be determined by direct aqueousinjection in accordance with Practice D 2908.1.4 It is the users responsibility to assure the validity of the
5、test method for untested matrices.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limita
6、tions prior to use. Specific precau-tionary statements are given in 8.5.5.1.2. Referenced Documents2.1 ASTM Standards:D 1129 Terminology Relating to Water2D 1193 Specification for Reagent Water2D 2908 Practice for Measuring Volatile Organic Matter inWater by Aqueous-Injection Gas Chromatography3E 35
7、5 Practice for Gas Chromatography Terms and Rela-tionships43. Terminology3.1 DefinitionsFor definitions of terms used in this testmethod, refer to Terminology D 1129 and Practice E 355.3.2 Description of Term Specific to This Standard:3.2.1 purgeable organicany organic material that is re-moved from
8、 aqueous solution under the purging conditionsdescribed in this test method (10.1.1).4. Summary of Test Method4.1 An inert gas is bubbled through the sample to purgevolatile compounds from the aqueous phase. These compoundsare then trapped in a column containing a suitable sorbent.After purging is c
9、omplete, trapped components are thermallydesorbed onto the head of a gas chromatographic column forseparation and analysis. Measurement is accomplished with anappropriate detector.5. Significance and Use5.1 Purgeable organic compounds, including organohalides,have been identified as contaminants in
10、raw and drinkingwater. These contaminants may be harmful to the environmentand man. Dynamic headspace sampling is a generally appli-cable method for concentrating these components prior to gaschromatographic analysis (1 to 5).5This test method can beused to quantitatively determine purgeable organic
11、 compoundsin raw source water, drinking water, and treated effluent water.6. Interferences6.1 Purgeable compounds that coelute with components ofinterest and respond to the detector will interfere with thechromatographic measurement. Likelihood of interference maybe decreased by using dissimilar col
12、umns or a more selectivedetector for the chromatographic step.7. Apparatus7.1 Purging DeviceCommercial devices are available forthis analysis. Either commercial apparatus or the equipmentdescribed below may be used for this analysis. Devices usedshall be capable of meeting the precision and bias sta
13、tementsgiven in 15.1.7.1.1 Glass Purging Device having a capacity of 5 mL isshown in Fig. A1.1. Construction details are given in AnnexA1. A glass frit is installed at the base of the sample reservoirto allow finely divided gas bubbles to pass through the aqueous1This test method is under the jurisd
14、iction of ASTM Committee D19 on Waterand is the direct responsibility of Subcommittee D19.06 on Methods forAnalysis forOrganic Substances in Water.Current edition approved Jan. 10, 2003. Published January 2003. Originallyapproved in 1979. Last previous edition approved in 1995 as D3871 84 (1995)e1.2
15、Annual Book of ASTM Standards, Vol 11.01.3Annual Book of ASTM Standards, Vol 11.02.4Annual Book of ASTM Standards, Vol 14.02.5The boldface numbers in parentheses refer to the references at the end of thistest method.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken
16、, PA 19428-2959, United States.sample while the sample is restrained above the frit. Thesample reservoir is designed to provide maximum bubblecontact time and efficient mixing.7.1.2 Gaseous volumes above the sample reservoir are keptto a minimum to provide efficient transfer and yet large enoughto a
17、llow sufficient space for foams to disperse. Inlet and exitports are constructed from 6.4-mm (14-in.) outside diametermedium-wall tubing so that leak-free removable connectionscan be made using“ finger-tight” compression fittings contain-ing plastic ferrules. The optional foam trap is used to contro
18、loccasional samples that foam excessively.7.2 TrapAshort section of stainless steel or glass tubing ispacked with a suitable sorbent. Traps should be conditionedbefore use (Section 11). While other trap designs and sorbentmaterials may be used (see Section 12), the trap and sorbentdescribed here are
19、 recommended and were used to collectprecision and bias data. If another trap design or sorbentmaterial is used, these precision and bias statements should beverified. A suitable trap design is 150 mm long by 3.17-mmoutside diameter (2.54-mm inside diameter). The front 100 mmis packed with 60 to 80
20、mesh 2,6-diphenyl-p-phenylene oxidefollowed by 50 mm of 35 to 60-mesh silica gel. One trapdesign is shown in Fig. A1.2, with details in Annex A1. Thebody assembly acts as a seal for the exit end of the trap. Themodified stem assembly is used to seal the inlet end of the trapwhen it is not in use.7.3
21、 Desorber consists of a trap heater and an auxiliarycarrier gas source to backflush the trap at elevated temperaturesdirectly onto the gas-chromatographic column. Desorber 1(Fig. A1.3 and Annex A1) is dedicated to one gas chromato-graph, but Desorber 2 can be used as a universal desorber formany gas
22、 chromatographs with a septum-type liquid-inletsystem.7.3.1 Desorber 1 is attached directly onto the gas-chromatograph liquid-inlet system after removing the septumnut, the septum, and the internal injector parts. The modifiedbody assembly is screwed onto the inlet system using the PTFEgasket as a s
23、eal. A plug is attached to one of the stemassemblies.7.3.1.1 The assembled parts, simply called “the plug,” areused to seal the desorber whenever the trap is removed tomaintain the flow of carrier gas through the gas-chromatographic column.7.3.1.2 The flow controller, PTFE tubing, and stem assem-bly
24、 are used to provide the trap-backflush flow. This entireassembly also provides gas flow to operate the purging device.7.3.2 Desorber 2 (Fig. A1.4 and Annex A1) may be at-tached to any gas chromatograph by piercing the gas-chromatographic liquid-inlet septum with the needle.7.3.2.1 The desorber is a
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD387119842003STANDARDTESTMETHODFORPURGEABLEORGANICCOMPOUNDSINWATERUSINGHEADSPACESAMPLING 用筒底 间隙 取样

链接地址:http://www.mydoc123.com/p-515681.html