1、Designation: D2414 11 D2414 12Standard Test Method forCarbon BlackOil Absorption Number (OAN)1This standard is issued under the fixed designation D2414; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A n
2、umber in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. Scope1.1 This test method covers the determination of the oil ab
3、sorption number of carbon black.1.2 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the use
4、r of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1765 Classification System for Carbon Blacks Used in Rubber ProductsD1799 Practice for Carbon BlackSampling Packag
5、ed ShipmentsD1900 Practice for Carbon BlackSampling Bulk ShipmentsD4483 Practice for Evaluating Precision for Test Method Standards in the Rubber and Carbon Black Manufacturing IndustriesD4821 Guide for Carbon BlackValidation of Test Method Precision and Bias3. Summary of Test Method3.1 In this test
6、 method, oil is added by means of a constant-rate buret to a sample of carbon black in the mixer chamber of anabsorptometer. As the sample absorbs the oil, the mixture changes from a free-flowing state to one of a semiplastic agglomeration,with an accompanying increase in viscosity. This increased v
7、iscosity is transmitted to the torque-sensing system of theabsorptometer. When the viscosity of the mixture reaches a predetermined torque level, the absorptometer and buret will shut offsimultaneously. The volume of oil added is read from the direct-reading buret. The volume of oil per unit mass of
8、 carbon blackis the oil absorption number.3.2 Either DBP or paraffin oil is DBP, paraffin or epoxidized sunflower oils are acceptable for use with most standard pelletedgrades of carbon black including N-series carbon blacks found in Classification D1765. OAN testing using paraffin oil on somespecia
9、lty blacks and powder blacksoils or epoxidized sunflower oils on some standard blacks and specialty blacks includingpowder products may result in unacceptable differences as compared to OAN testing using DBP oil. While studies have showneither oil to exhibit comparable precision, paraffin oil offers
10、 the advantage of being non-hazardous; even FDA-approved gradesare available. For either oil, with DBP oil. Paraffin and epoxidized sunflower oils are considered non-hazardous; some paraffin oilsare FDA approved. For any of the oils, Sections 8-11 (Calibration, Procedure, Calculation, and Report) ar
11、e to be consistent withthe oil selected for use. Referee testing between suppliers and users should use DBP oil until such time that precision data areavailable for paraffin oil.alternate oils.4. Significance and Use4.1 The oil absorption number of a carbon black is related to the processing and vul
12、canizate properties of rubber compoundscontaining the carbon black.5. Apparatus35.1 Balance, analytical, with an 0.01-g sensitivity.5.2 Oven, gravity-convection type, capable of maintaining 125 6 5C.5.3 Spatula, rubber, 100-mm.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Co
13、nshohocken, PA 19428-2959. United States15.4 Absorptometer, equipped with a constant-rate buret that delivers 4 6 0.024 cm3/min.45.5 Desiccator.6. Reagent and Standards6.1 Purity of ReagentsReagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is intended that allreagent
14、s shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, wheresuch specifications are available.5 Other grades may be used, provided it is first ascertained that the reagent is of sufficiently highpurity to permit its use without lessening the a
15、ccuracy of the determination.6.2 n-Dibutyl Phthalate, having a density of 1.042 to 1.047 Mg/m3 at 25C and a relative density of 1.045 to 1.050 at 25C.6.3 Paraffn Oil, having a kinematic viscosity of 10 to 34 mm2/s (cSt) at 40C.NOTE 1Three paraffin oils have been found suitable including Marcol 82 fr
16、om Exxon, 80/90 White Oil from Conoco-Phillips, and LC1 oil from LabChemicals, Germany. All three oils are pharmaceutical or food grade oil, or both, based on available data.6.4 Epoxidized sunflower oils, (fatty acid ester, epoxidized) having a density of 0.90 to 0.91 g/cm3 at 20C.66.5 ASTM D24 Stan
17、dard Reference Blacks, SRB.77. Sampling7.1 Samples shall be taken in accordance with Practices D1799 and D1900.8. Calibration and Standardization8.1 Absorptometer:8.1.1 ModelThree different types of absorptometers are in use: (1) early models based on springs and mechanical indicationof torque (Type
18、 A and B), (2) second generation absorptometers equipped with load cells and digital torque display (Type E8), and(3) current model absorptometers which are designed with a torque measuring system that includes a micro-computer and softwareto continuously record torque and oil volume with time (Type
19、s H and C and modified Type E8). Types A, B, and E8 are designedto stop mixing at a predetermined, fixed torque level, which is the recommended procedure for measuring hard or tread blacks(calibration Procedure A). The computer controlled models (Types H and C and modified Type E8) are required for
20、runningcalibration Procedure B, the recommended torque curve analysis for the determination of the end-point of soft or carcass blacks.The Type H and C and modified Type E8 absorptometers can also provide an end-point at a fixed or predetermined torque levelsuch that these types of absorptometers ar
21、e well-suited for measuring OAN of both hard and soft carbon blacks. Severalcomponents influence the calibration: the dynamometer torque spring or the load cell, the torque limit switch or the indicator setpoint, the damper (oil damper or electronic damping), and the mixing head consisting of two co
22、unter rotating blades and a mixingbowl. It is necessary that all of these components are in good condition and are properly adjusted to achieve acceptable calibration.8.1.2 Mixing BowlTypically the absorptometer is delivered with either a surface-treated stainless steel or anodized aluminummixing bo
23、wl. These bowls are considered acceptable provided they give the correct reading for the appropriate SRB referencestandards. The surface finish of the mixer chamber is critical for maintaining proper calibration, and the bowl should not bemodified to achieve calibration.NOTE 2Stainless steel chamber
24、s have been found satisfactory for the test when they are manufactured to a roughness value (Ra) of 2.5 6 0.4 m(100 6 15 in.) based upon 8 measurements. No single measurement should be greater than 3.6 m (140 in.) or less than 1.5 m (60 in.). Stainlesssteel bowls purchased with an absorptometer have
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD24142012STANDARDTESTMETHODFORCARBONBLACKOILABSORPTIONNUMBEROAN 炭黑 吸油量 OAN 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-512830.html