ASTM C1332-2001(2013) Standard Test Method for Measurement of Ultrasonic Attenuation Coefficients of Advanced Ceramics by Pulse-Echo Contact Technique《用脉冲反射接触技术测量高级陶瓷超声衰减系数的标准试验方法》.pdf
《ASTM C1332-2001(2013) Standard Test Method for Measurement of Ultrasonic Attenuation Coefficients of Advanced Ceramics by Pulse-Echo Contact Technique《用脉冲反射接触技术测量高级陶瓷超声衰减系数的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM C1332-2001(2013) Standard Test Method for Measurement of Ultrasonic Attenuation Coefficients of Advanced Ceramics by Pulse-Echo Contact Technique《用脉冲反射接触技术测量高级陶瓷超声衰减系数的标准试验方法》.pdf(12页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C1332 01 (Reapproved 2013)Standard Test Method forMeasurement of Ultrasonic Attenuation Coefficients ofAdvanced Ceramics by Pulse-Echo Contact Technique1This standard is issued under the fixed designation C1332; the number immediately following the designation indicates the year oforigi
2、nal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes a procedure for measurementof ultraso
3、nic attenuation coefficients for advanced structuralceramic materials. The procedure is based on a broadbandbuffered piezoelectric probe used in the pulse-echo contactmode and emitting either longitudinal or shear waves. Theprimary objective of this test method is materials characteriza-tion.1.2 The
4、 procedure requires coupling an ultrasonic probe tothe surface of a plate-like sample and the recovery of succes-sive front surface and back surface echoes. Power spectra ofthe echoes are used to calculate the attenuation spectrum(attenuation coefficient as a function of ultrasonic frequency)for the
5、 sample material. The transducer bandwidth and spectralresponse are selected to cover a range of frequencies andcorresponding wavelengths that interact with microstructuralfeatures of interest in solid test samples.1.3 The purpose of this test method is to establish funda-mental procedures for measu
6、rement of ultrasonic attenuationcoefficients. These measurements should distinguish and quan-tify microstructural differences among solid samples andtherefore help establish a reference database for comparingmaterials and calibrating ultrasonic attenuation measurementequipment.1.4 This test method a
7、pplies to monolithic ceramics and alsopolycrystalline metals. This test method may be applied towhisker reinforced ceramics, particulate toughened ceramics,and ceramic composites provided that similar constraints onsample size, shape, and finish are met as described herein formonolithic ceramics.1.5
8、 This test method sets forth the constraints on samplesize, shape, and finish that will assure valid attenuationcoefficient measurements. This test method also describes theinstrumentation, methods, and data processing procedures foraccomplishing the measurements.1.6 This test method is not recommen
9、ded for highly attenu-ating materials such as very thick, very porous, rough-surfacedmonolithics or composites. This test method is not recom-mended for highly nonuniform, heterogeneous, cracked,defective, or otherwise flaw-ridden samples that are unrepre-sentative of the nature or inherent characte
10、ristics of thematerial under examination.1.7 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory
11、 limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2C1331 Test Method for Measuring Ultrasonic Velocity inAdvanced Ceramics with Broadband Pulse-Echo Cross-Correlation MethodE664 Practice for the Measurement of the Apparent Attenu-ation of Longitudinal Ultrasonic Waves by ImmersionM
12、ethodE1316 Terminology for Nondestructive ExaminationsE1495 Guide for Acousto-Ultrasonic Assessment ofComposites, Laminates, and Bonded Joints2.2 ASNT Document:Recommended Practice SNT-TC-1A for NondestructiveTesting Personnel Qualification and Certification32.3 Military Standard:MIL-STD-410 Nondest
13、ructive Testing Personnel Qualifica-tion and Certification41This test method is under the jurisdiction of ASTM Committee C28 onAdvanced Ceramics and is the direct responsibility of Subcommittee C28.03 onPhysical Properties and Non-Destructive Evaluation.Current edition approved Feb. 1, 2013. Publish
14、ed April 2013. Originallyapproved in 1996. Last previous edition approved in 2007 as C133201(2007). DOI:10.1520/C1332-01R13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, r
15、efer to the standards Document Summary page onthe ASTM website.3Available fromAmerican Society for Nondestructive Testing (ASNT), P.O. Box28518, 1711 Arlingate Ln., Columbus, OH 43228-0518, http:/www.asnt.org.4Available from Standardization Documents Order Desk, DODSSP, Bldg. 4,Section D, 700 Robbin
16、s Ave., Philadelphia, PA 19111-5098, http:/www.dodssp.daps.mil.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States12.4 Additional references are cited in the text and at end ofthis test method.3. Terminology3.1 Definitions of Terms Specif
17、ic to This Standard:3.1.1 acoustic impedance (Z)a property (1)5defined by amaterials density, p, and the velocity of sound within it, v,where Z = v.3.1.2 attenuation coeffcient ()decrease in ultrasoundintensity with distance expressed in nepers (Np) per unitlength, herein, = ln(I0/I)/d, where is att
18、enuationcoefficient, d is path length or distance, I0is original intensityand I is attenuated intensity (2).3.1.3 attenuation spectrumthe attenuation coefficient, ,expressed as a function of ultrasonic frequency, f, or plotted as versus f, over a range of ultrasonic frequencies within thebandwidth o
19、f the transducer and associated pulser-receiverinstrumentation.3.1.4 back surfacethe surface of a test sample which isopposite to the front surface and from which back surfaceechoes are returned at normal incidence directly to the trans-ducer.3.1.5 bandwidththe frequency range of an ultrasonicprobe,
20、 defined by convention as the difference between thelower and upper frequencies at which the signal amplitude is 6dB down from the frequency at which maximum signalamplitude occurs. The frequency at which the maximumoccurs is termed the center frequency of the probe or trans-ducer.3.1.6 broadband tr
21、ansduceran ultrasonic transducer ca-pable of sending and receiving undistorted signals over a broadbandwidth, consisting of thin damped piezoelectric crystal in abuffered probe (search unit).3.1.7 buffered probean ultrasonic search unit as defined inTerminology E1316 but containing a delay line or b
22、uffer rod towhich the piezoelement, that is, transducer consisting of apiezoelectric crystal, is affixed. The buffer rod separates thepiezoelement from the test sample (see Fig. 1).3.1.8 buffer rodan integral part of a buffered probe orsearch unit, usually a quartz or fused silica cylinder thatprovi
23、des a time delay between the excitation pulse from thepiezoelement and echoes returning from a sample coupled tothe free end of the buffer rod.3.1.9 free surfacethe back surface of a solid test sampleinterfaced with a very low density medium, usually air or othergas, to assure that the back surface
24、reflection coefficient equals1 to a high degree of precision.3.1.10 frequency (f)number of oscillations per second ofultrasonic waves, measured in megahertz, MHz, herein.3.1.11 front surfacethe surface of a test sample to whichthe buffer rod is coupled at normal incidence (designated as testsurface
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC133220012013STANDARDTESTMETHODFORMEASUREMENTOFULTRASONICATTENUATIONCOEFFICIENTSOFADVANCEDCERAMICSBYPULSEECHOCONTACTTECHNIQUE

链接地址:http://www.mydoc123.com/p-464266.html