1、Beneficial comments, recommendations, additions, deletions, clarifications, etc. and any data which may improve this document should be sent to: US Army Geospatial Center, 7701 Telegraph Road, ATTN: CEAGC-GST, Alexandria, VA 22315-3864 or emailed to thomas.coxjrus.army.mil. Since contact information
2、 can change, you may want to verify the currency of this address information using the ASSIST Online database at https:/assist.daps.dla.mil. AMSC N/A FSC 6635 Distribution Statement A. Approve for public release; distribution is unlimited. A-A-52316C 05 June 2010 SUPERSEDING A-A-52316B 1 March 2005
3、COMMERCIAL ITEM DESCRIPTION FOR THE TESTER, CONSTRUCTION MATERIALS, MOISTURE & DENSITY (TCMMD) 1. SCOPE. This commercial item description covers a single unit density and moisture tester with radioactive sources. The tester is used for measuring the density and moisture content of soils, and the den
4、sity of asphalt pavement and other construction material. 2. SALIENT CHARACTERISTICS. The TCMMD will be a one-piece self contained instrument capable of measuring the density of soils, asphalt pavement, and other construction material. The TCMMD will be capable of non-destructively measuring the moi
5、sture content of soils. Its desired that the TCMMD be capable of measuring the density of soils, asphalt pavement, and other construction materials and measuring the moisture content of soils through non-destructive methods. The TCMMD shall be a measuring device comprised of radioactive sources, rad
6、ioactive detectors, a readout device, data storage, and an integral power supply. The TCMMD shall be capable of testing for density and moisture by backscatter, direct transmission, and (optional) air gap methods. 2.1 Radioactive sources. The sources shall be doubly or triply encapsulated and sealed
7、 having sealed source certification. The radiological surface dose rate shall not be greater than 40.00 millirem/hour (mrem/hr) at the surface of the TCMMD (neutron and gamma sources). 2.2 Radioactive detectors. The detectors shall be any suitable type of radioactive detectors. INCH-POUNDProvided by
8、 IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-52316C 2 2.3 Measurement Depth. The TCMMD shall be capable of depth measurements up to 12 inches in 2 inch increments or less. The index rod or index strip on the gauge allowing the trigger mechanism to be lat
9、ched or locked at the desired measurement depth. The gauge being designed so that if the operator attempts to pick it up without first returning the source rod to the shielded position, the trigger in the handle releases and the rod retracts to the shielded position. The source rod must be lockable.
10、 2.4 Visual display. The TCMMD shall have a self contained, easily readable and understandable visual display. The visual display shall provide an English readout with automatic conversion to metric units. The visual display shall not become obscured while operating in humid, wet, cold, hot, dusty,
11、or dry environments. The visual display shall incorporate an integrated back light that enables the analyst to read the visual display in reduced visibility, low light, conditions. The visual display shall provide a direct readout of Wet Density (WD) in lbs/ft3, Dry Density (DD) in lbs/ft3, moisture
12、 content in lbs/ft3, percent moisture, percent air void, percent proctor, percent marshall, and percent compaction. The TCMMD visual display shall also display the remaining power supply capacity. At a minimum, the visual display shall contain counting circuits, timing circuits, a data display, and
13、related electronic components. 2.5 Data storage and Data transfer. The TCMMD Shall have the ability to internally record a minimum of 120 test records and be provided with a Universal Serial Bus (USB) port to allow data to be transferred from the TCMMD onto external data storage devices. Each record
14、 shall include the date conducted and provision for entering a task identifier of not less than five characters. 2.6 Power supply. The TCMMD shall be provided with an operator removable, rechargeable, replaceable, extended life battery source capable of performing a minimum of 3000 one minute tests
15、before recharging is needed. The TCMMD shall be provide with a backup power supply that allows the operator to perform an additional 120 measurements in the event the main power supply becomes depleted. 2.6.1 Battery Charger. An automatic charger with quick-disconnect plug-in type connectors shall b
16、e provided. The time it takes to fully recharge the battery shall not be greater than sixteen hours. The charger shall operate from 12 volts, dc or 110 volts, ac, 60-hertz power, and shall protect the voltage regulator from damage when the battery is connected improperly. The tester shall incorporat
17、e a battery overcharge prevention feature. 2.6.2 Rechargeable Batteries. If rechargeable batteries are used, they shall be of a type that minimizes the promotion of corrosion, when left installed for extended periods. Provided by IHSNot for ResaleNo reproduction or networking permitted without licen
18、se from IHS-,-,-A-A-52316C 3 2.7 Voltage indicator. An indicator or other device shall provide a clear distinction between when the voltage is adequate for testing, low, and when the TCMMD is running on back-up power. 2.8 Housing. The power supply, sources, detectors, readout device, etc., shall be
19、in a housing of rugged construction that is moisture and dust proof. The housing shall conform to the U.S. Nuclear Regulatory Commission (NRC) requirements in Title 10 of the Code of Federal Regulations. 2.9 Tools and accessories. The TCMMD shall be furnished with a reference standard, all tools nec
20、essary for operation, site-preparation devices, technical publications, an Alternating Current (AC) battery charger, 24-V Direct Current (DC) battery charger, operating instructions, and a transport case with locking mechanism. All tools and accessories shall be stored and transported within the sam
21、e case as the TCMMD. The tools shall include, at a minimum, guide-plate/scraper, drill rod/pin, drill rod/pin drive/insertion tool, drill rod/pin extraction tool, lubrication, sign kit, and computer data download cable. 2.9.1 Reference standard. The reference standard shall be a combination type, pr
22、ovided to check equipment operation, background count, and to establish conditions for a reproducible density and moisture count rate. 2.9.2 Technical Publications. Department of the Army Technical Manuals will be used to support this system IAW MIL-STD-40051-2 and AR 25-30. Supplemental data will i
23、nclude Maintenance Allocation Charts (MAC)s, Pre-Maintenance Check (PMC), and Repair Parts Special Tools Listing (RPSTL). 2.9.3 Backscatter. Site preparation devices for the backscatter test method shall include a flat plate straightedge or other suitable tools, to be used to level the test site to
24、the required uniformity. 2.9.4 Direct Transmission. When required, the provided site preparation devices for the direct transmission test method shall include an auger, rod, or other hole forming device with marks to indicate depth. It shall have a bore diameter equal to or up to .12 inch larger tha
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ARMYAA52316C2010TESTERCONSTRUCTIONMATERIALSMOISTUREDENSITYTCMMD 建筑材料 湿度 密度 检验 PDF

链接地址:http://www.mydoc123.com/p-439928.html