1、Designation: D3649 06 (Reapproved 2014)Standard Practice forHigh-Resolution Gamma-Ray Spectrometry of Water1This standard is issued under the fixed designation D3649; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、 revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the measurement of gamma-rayemitting radionuclides in water by means of gamma-rayspectrometry. It is
3、applicable to nuclides emitting gamma-rayswith energies greater than 45 keV. For typical counting systemsand sample types, activity levels of about 40 Bq are easilymeasured and sensitivities as low as 0.4 Bq are found for manynuclides. Count rates in excess of 2000 counts per secondshould be avoided
4、 because of electronic limitations. Highcount rate samples can be accommodated by dilution, byincreasing the sample to detector distance, or by using digitalsignal processors.1.2 This practice can be used for either quantitative orrelative determinations. In relative counting work, the resultsmay be
5、 expressed by comparison with an initial concentrationof a given nuclide which is taken as 100 %. For quantitativemeasurements, the results may be expressed in terms of knownnuclidic standards for the radionuclides known to be present.This practice can also be used just for the identification ofgamm
6、a-ray emitting radionuclides in a sample without quan-tifying them. General information on radioactivity and themeasurement of radiation has been published (1,2).2Informa-tion on specific application of gamma spectrometry is alsoavailable in the literature (3-5). See also the referenced ASTMStandard
7、s in 2.1 and the related material section at the end ofthis standard.1.3 This standard does not purport to address the safetyconcerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety andhealth practices and determine the applicabi
8、lity of regulatorylimitation prior to use.2. Referenced Documents2.1 ASTM Standards:3D1066 Practice for Sampling SteamD1129 Terminology Relating to WaterD2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on WaterD3370 Practices for Sampling Water from C
9、losed ConduitsD3648 Practices for the Measurement of RadioactivityD4448 Guide for Sampling Ground-Water Monitoring WellsE181 Test Methods for Detector Calibration and Analysis ofRadionuclides3. Terminology3.1 DefinitionsFor definitions of terms used in thispractice, refer to Terminology D1129. For t
10、erms not defined inthis practice or in Terminology D1129, reference may be madeto other published glossaries.4. Summary of Practice4.1 Gamma ray spectra are measured with modular equip-ment consisting of a detector, high-voltage power supply,preamplifier, amplifier and analog-to-digital converter (o
11、r digi-tal signal processor), multichannel analyzer, as well as acomputer with display.4.2 High-purity germanium (HPGe) detectors, p-type orn-type, are used for the analysis of complex gamma-ray spectrabecause of their excellent energy resolution. These germaniumsystems, however, are characterized b
12、y high cost and requirecooling. Liquid nitrogen or electromechanical cooling, or both,can be used.4.3 In a germanium semiconductor detector, gamma-rayphotons produce electron-hole pairs. The charged pair is thencollected by an applied electric field. A very stable low noisepreamplifier is needed to
13、amplify the pulses of electric charge1This practice is under the jurisdiction of ASTM Committee D19 on Water andis the direct responsibility of Subcommittee D19.04 on Methods of RadiochemicalAnalysis.Current edition approved June 1, 2014. Published July 2014. Originally approvedin 1978. Last previou
14、s edition approved in 2006 as D3649 06. DOI: 10.1520/D3649-06R14.2The boldface numbers in parentheses refer to a list of references at the end ofthis standard.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of
15、ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized p
16、rinciples on standardization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.1resulting from gamma photon interactions. The output from thepreamplifi
17、er is directly proportional to the energy deposited bythe incident gamma-ray. These current pulses are fed into anamplifier of sufficient gain to produce voltage output pulses inthe amplitude range from 0 to 10 V.4.4 A multichannel pulse-height analyzer is used to deter-mine the amplitude of each pu
18、lse originating in the detector,and accumulates in a memory the number of pulses in eachamplitude band (or channel) in a given counting time. Com-puterized systems with stored programs and interface hardwarecan accomplish the same functions as hardwired multichannelanalyzers. The primary advantages
19、of the computerized systeminclude the capability of programming the multi-channel ana-lyzer functions and the ability to immediately perform datareduction calculations using the spectral data stored in thecomputer memory or mass storage device. Fora0to2-MeVspectrum, 4000 or more channels are typical
20、ly needed in orderto fully utilize a germanium detectors excellent energy reso-lution.4.5 The distribution of the amplitudes (pulse heights) of thepulses can be separated into two principal components. One ofthese components has a nearly Gaussian distribution and is theresult of total absorption of
21、the gamma-ray energy in thedetector. This peak is normally referred to as the full-energypeak or photopeak. The other component is a continuous onelower in energy than that of the photopeak. This continuouscurve is referred to as the Compton continuum and is due tointeractions wherein the gamma phot
22、ons deposit only part oftheir energy in the detector. These two portions of the curve areshown in Fig. 1. Other peaks, such as escape peaks, backscat-tered gamma rays or X rays from shields, are often superim-posed on the Compton continuum. Escape peaks will bepresent when gamma-rays with energies g
23、reater than 1.02 MeVare emitted from the sample. The positron formed in pairproduction is usually annihilated in the detector and one orboth of the 511keV annihilation quanta may escape from thedetector without interaction. This condition will cause single ordouble escape peaks at energies of 0.511
24、or 1.022 MeV lessthan the photopeak energy. In the plot of pulse height versuscount rate, the size and location of the photopeak on the pulseheight axis is proportional to the number and energy of theincident photons, and is the basis for the quantitative andqualitative application of the spectromet
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ANSIASTMD36492006STANDARDPRACTICEFORHIGHRESOLUTIONGAMMARAYSPECTROMETRYOFWATERPDF

链接地址:http://www.mydoc123.com/p-432356.html