ISO 10313-1993 Ambient air determination of the mass concentration of ozone chemiluminescence method《环境空气 臭氧质量浓度的测定 化学发光法》.pdf
《ISO 10313-1993 Ambient air determination of the mass concentration of ozone chemiluminescence method《环境空气 臭氧质量浓度的测定 化学发光法》.pdf》由会员分享,可在线阅读,更多相关《ISO 10313-1993 Ambient air determination of the mass concentration of ozone chemiluminescence method《环境空气 臭氧质量浓度的测定 化学发光法》.pdf(12页珍藏版)》请在麦多课文档分享上搜索。
1、INTERNATIONAL STANDARD IS0 10313 First edition 1993-04-I 5 Ambient air - Determination of the mass concentration of ozone - Chemiluminescence method Air ambiant - D therefore, the time during which the ambient air remains in the sampling line must be sufficiently short to keep this effect to a minim
2、um. NOTE 1 It has been reported by some researchers that, at about 80 % relative humidity and 22 “C, the responses for some commercially available chemiluminescence analysers were about IO % higher than that for dry air. However, comparisons of ambient ozone measurements using com- mercially availab
3、le chemiluminescence and UV-photometric anaiysers showed no significant differences. This suggests that, in practice, any errors associated with calibrations us- ing dry air are compensated by other effects. 2 Definition For the purposes of this International Standard, the following definition appli
4、es. 2.1 transfer standard: An apparatus (transportable) with associated operational procedures that is capable of accurately measuring or reproducing ozone con- centration standards which are quantitatively related to the authoritative primary standard. 3 Principle The air sample is drawn continuous
5、ly, at a constant flow rate, through a particle filter before it enters the chemiluminescence analyser. It then flows into a re action chamber where it is mixed with an excess flow of ethylene. Ozone and ethylene react instan- taneously to produce light in the visible region with a maximum at a wave
6、length of about 400 nm. The emitted light intensity is proportional to the concen- tration of ozone in the air sample and is measured by a photomultiplier tube. The resulting voltage is ampli- fied, displayed, and calibrated in terms of the ambient ozone concentration. 4 Reagents and materials 4.1 S
7、ampling line The sampling line shall be made of material that is inert to ozone, such as glass or fluorocarbon polymer, and it shall be as short as possible to keep the time the air sample remains there to a minimum. Any am- bient nitrogen(ll) oxide present in the air sample will react with some of
8、the ozone during the time it re- mains in the sampling line. This decay of ozone is a complicated function of the mass concentration ratio, p(03)/p(NO), and the difference in mass concen- trations, (0) - p(N0). Calculations have shown that, if the residence time is less than 0,5 s, the decay in init
9、ial ozone will be less than 1 % for most ambient ozone and nitrogenIll) oxide concentrations encoun- tered. Hence, it is recommended that the sampling 1 IS0 10313:1993(E) line length and pumping speed be chosen such that the time the air remains within the sampling line is less than O,5 s. However,
10、in situations of routine monitoring, a time of up to 5 s is allowed. Any sampling line or manifold preceding the filter shall be clean. Proper precautions shall be taken to prevent condensation inside the sampling line, for example, by heating the line. 4.2 Particle filter The filter and its support
11、 shall be made of material inert to ozone, such as fluorocarbon polymer, and it shall remove all particles likely to alter the perform- ance of the analyser. It shall be changed on a regular basis, depending on the concentrations of ambient particles at the sampling site. This is necessary be- cause
12、 excessive accumulation of particles on the filter can cause loss of ozone from the air sample and an excessive pressure drop across the filter. NOTES 2 Frequently, a filter pore size of 5 pm is used. 3 Generally, new filters need some time to be conditioned by the ambient atmosphere. As a result, m
13、easured mass concentrations of ozone have been obsenred to decrease temporarily by 5 % to 10 % for periods of 5 min to 15 min immediately following filter changes. 4.4 Zero air, required in the anatyser calibration procedure. It shall be free of ozone, nitrogen oxides and any other interfering subst
14、ance that csn cause an undesired positive or negative response in either the ultraviolet photometer or the chemiluminescence analyser. The response of the ozone chemiluminescence ana- lyser is affected by the oxygen content in the gas sample. Therefore, if synthetic air is used, the oxygen content s
15、hall be within f 2 46 of the normal atmos- pheric concentration of 20,9 %. NOTE 4 Details on a scrubbing system for making zero air from ambient air may be found in I and t2 in annex B. 5 Apparatus 5.1 Chsmllumlnescence analyser, consisting of sample and ethylene flow inlet systems, a constant tempe
16、rature reaction chamber coupled to a photo- multiplier detector and accompanying signal process- ing electronics. A flowmeter, if it is not an integral part of the analyser, shall be available for verifying that the flow meets the specifications of the instrument. A schematic diagram of a typical sy
17、stem is shown in figure 1. 4.3 Ethylene, of minimum purity 995 %. WARNING - Ethylene has a lower exploslvs llmlt of 27,5 l/mR Any excess shall be vented outdoon or, tf necessary, chemically removed, especially at sites where hydrocarbons are being measured. Observe standard safety precautions when u
18、rlng this flammable gas. 59 Apparatus for callbratlon, an ozone calibration system shall be used. (See the simplified schematic diagram in figure 2.) NOTE 5 An equivalent system would ba one that usas only the air line with the ozone generator to bring zero and ozonizad air to the manifold, hence el
19、iminating the need for the four-ports two-way valve. IS0 10313:1993(E) Flow controller II Flowmefer I Temperature controller II I I II Quartz A I h Sample inlet - =f window - II M - I, ._ ,:-I Reaction chamber Particulate filter Thermo-electric cooler Flow I- controller r . Flowmeter - CX”cIU5T Figu
20、re 1 - Schematic diagram of a chemiluminescence ozone analyser 3 IS0 10313:1993(E) under callbratlon Ultraviolet photometer r”“-“-,.-”“1 I optics i I Absorptlon cell 0 I I Low pressure ; mercury vapwr ; Lamp I I ! I I Figure 2 - Schematlc dlagram of a typlcal ultmvlolet photometric callbrstlon syste
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ISO103131993AMBIENTAIRDETERMINATIONOFTHEMASSCONCENTRATIONOFOZONECHEMILUMINESCENCEMETHOD 环境 空气 臭氧 质量 浓度

链接地址:http://www.mydoc123.com/p-1249624.html