1、TRACE NITROGEN IN LIGHT AROMATIC HYDROCARBONSBY CHEMILUMINESCENCEUOP Method 971-00SCOPEThis method is for determining total nitrogen in light aromatic hydrocarbons, such as benzene. It isapplicable to quantitative determination of nitrogen at a level of 30 to 400 ng/g. Samples containing morethan 40
2、0 ng/g nitrogen can be analyzed according to ASTM D 4629 or D 6069. Halogen and sulfur-containing compounds can interfere if any one or all of these elements are present in a sample inconcentration greater than 10%.OUTLINE OF METHODThe sample is injected by an autosampler into a quartz-wool filled s
3、ample boat. The volatile componentsare allowed to evaporate into an argon stream that carries them into the combustion furnace. The boat isthen moved into the furnace and the residual material is combusted in oxygen. The organic material isconverted to carbon dioxide and water. The nitrogen in the s
4、ample is converted to nitric oxide that is furtherreacted in the detector with ozone to produce nitrogen dioxide that is generated in an excited state. Theexcited state nitrogen dioxide molecules relax to the ground state, producing light (chemiluminescence),that is detected by a photomultiplier tub
5、e. The signal is proportional to the total nitrogen in the sample. Thedetector is operated at reduced pressure to decrease the probability of excited state nitrogen dioxidemolecules colliding with other molecules before emitting light, thus increasing sensitivity and signal-to-noise ratio.This proce
6、dure is not sensitive to diatomic nitrogen (N2) and has reduced sensitivity to compounds withnitrogen-nitrogen bonds that decompose to produce nitrogen gas. The relative responses of some nitrogencompounds have been determined and tabulated by Jones and Daughton.APPARATUSReferences to catalog number
7、s and suppliers are included as a convenience to the method user. Othersuppliers may be used.IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TODETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH ANDSAFETY PRACTICES ARE TO BE FOLLOW
8、ED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THISPROCEDURE IN THE MANNER PRESCRIBED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS(MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED INTHIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIAT
9、E PERSONAL PROTECTIONEQUIPMENT (PPE). COPYRIGHT 2000 UOP LLCALL RIGHTS RESERVEDUOP Methods are available through ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken PA 19428-2959,United States. The Methods may be obtained through the ASTM website, www.astm.org, or by contacting
10、 Customer Service atserviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE2 of 8971-00Chemiluminescence analyzer, Mitsubishi TN-110 analyzer with an automatic boat controller module,Cosa Instruments. This method was developed using the Mitsubishi analyzer. The procedure foranalysis may be differen
11、t for other instruments. The Mitsubishi analyzer must be equipped with thefollowing accessories:Autosampler, Mitsubishi Cat. No. ASC-150L, Cosa InstrumentsAutosampler syringes, gas tight, 100-L, Mitsubishi Cat. No. MSSG10, Cosa InstrumentsAutosampler vials, rinse, Mitsubishi Cat. No. TX3LSW, Cosa In
12、strumentsAutosampler vials, sample, Mitsubishi Cat. No. TX3LSV, Cosa InstrumentsVacuum pump, 50 L/min minimum flow capacity, Sinku Kiko GLS-050 (Mitsubishi), CosaInstruments. See NOTE 1.Filter capsule, particulate, HEPA-Cap 36, glass fiber, Fisher Scientific, Cat. No. 09-744-12. Connectthis filter b
13、efore the membrane drier.Flasks, volumetric, glass, Class A, 25-, 50 (two required), and 100-mL, Fisher Scientific, Cat. Nos.10-210A, B and C, respectivelyMembrane drier, Perma Pure, MD-110-12F. See NOTE 2.Regulator, air, stainless steel, two-stage, high purity, Matheson Gas Products, Model 3810-590
14、Regulator, argon, stainless steel, two-stage, high purity, Matheson Gas Products, Model 3810-580Regulator, oxygen, stainless steel, two-stage, high purity, Matheson Gas Products, Model 3810-540Syringes, gas tight, 50 and 100 L, Fisher Scientific, Cat. Nos. 14-824-30 and 13-684-100. Thesyringe used f
15、or pure pyridine should not be used for low nitrogen samples without first beingthoroughly cleaned.REAGENTS AND MATERIALSReferences to catalog numbers and suppliers are included as a convenience to the method user. Othersuppliers may be used.Air, compressed, dry, oil-free air for Perma Pure membrane
16、 drierAluminum foil, optional, see NOTE 3, local supplyArgon, high purity, 99.98+%, Matheson Gas ProductsBenzene, low nitrogen, as an alternative to isooctane, local supplyCatalyst, Mitsubishi Cat. No. TN5CAT, Cosa InstrumentsIsooctane, should be as low in residual nitrogen as possible to minimize t
17、he blank value, Pesticide grade,Fisher Scientific, O297-4. If available, benzene containing very low concentrations of nitrogencompounds could be substituted. The blank value must be determined for each new bottle of solventused for making standards.3 of 8971-00Oxygen, high purity, 99.7+%, Matheson
18、Gas ProductsPyridine, 99+%, Aldrich, Cat. No. 36,057-0Quartz Wool, Mitsubishi Cat. No. TNQWL, Cosa InstrumentsToluene, Fisher Scientific, Cat. No. T324-4PROCEDUREPreparation Of Standard SolutionsPyridine stock solution, 100 g N/mL1. Dispense 58 L of pyridine using a microliter syringe into a clean,
19、dry 100 mL volumetric flask.2. Dilute to the mark with toluene. Cap and mix well.The identity below shows the calculation of the concentration of nitrogen in the stock solution:()( )( )( )()()=58 0.978 14.01 1000100gN/mL79.10 100where:0.978 = density of pyridine at 20C, mg/L14.01 = molecular mass of
20、 nitrogen, g/mol58 = volume of pyridine, L100 = dilution volume, mL79.10 = molecular mass of pyridine, g/mol1000 = factor to convert mg to gCalibration standard solution, 50 ng N/mL1. Dispense 25 L of the pyridine stock solution with a microliter syringe into a clean, dry 50-mLvolumetric flask.2. Di
21、lute to the mark with a low nitrogen solvent, either isooctane or benzene. Cap the flask and mixwell. The actual concentration of this standard will be 1/2000thof the pyridine stock solution.Calibration standard solution, 100 ng N/mL1. Dispense 50 L of the pyridine stock solution with a microliter s
22、yringe into a clean, dry 50-mLvolumetric flask.2. Dilute to the mark with the low nitrogen solvent used above. Cap the flask and mix well. The actual concentration of this standard will be 1/1000thof the pyridine stock solution.4 of 8971-00Calibration standard solution, 200 ng N/mL1. Dispense 50 L o
23、f the pyridine stock solution with a microliter syringe into a clean, dry 25-mLvolumetric flask.2. Dilute to the mark with a low nitrogen solvent used above. Cap the flask and mix well. The actual concentration of this standard will be 1/500th of the pyridine stock solution.The stock solution can be
24、 retained, if refrigerated, for up to three months. The standard solutions can beretained, if refrigerated, for one week. Do not refrigerate stock and standard solutions if prepared in benzeneas the benzene will freeze, causing erratic results.SamplingSome organic nitrogen compounds will adsorb on t
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9712000TRACENITROGENINLIGHTAROMATICHYDROCARBONSBYCHEMILUMINESCENCEPDF

链接地址:http://www.mydoc123.com/p-1074593.html