1、Designation: F2528 06 (Reapproved 2014)Standard Test Methods forEnteral Feeding Devices with a Retention Balloon1This standard is issued under the fixed designation F2528; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of
2、 last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 These test methods cover the establishment of perfor-mance requirements for the utilization of a single-use, enteralf
3、eeding device with a retention balloon, used by medicalprofessionals for providing a means of nutrition and/or admin-istration of medication to patients by means of natural orifice(nasal, oral, transluminal) and or a surgically created stoma.The product is manufactured in various sizes and materials
4、such as silicone, urethane, and various polymers (as well ascombinations of these) and is provided nonsterile for steriliza-tion and sterile for single use only. Rationale for these testmethods can be found in Appendix X1.1.2 This standard does not purport to address all of thesafety concerns, if an
5、y, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and to determine theapplicability of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2F623 Performance Specification for Foley Catheter
6、2.2 Other Standard:Simulated Gastric Fluid, USP Official Compendia of Stan-dards33. Terminology3.1 Definitions:3.1.1 balloon integrity (resistance to rupture), nvolume ofliquid that corresponds with balloon failure, or bursting.3.1.2 distal, nrefers to the balloon end of the enteralfeeding device3.1
7、.3 enteral feeding device with retention balloon, natwo-way medical device intended to provide a means ofnutrition or administration of medication, or both, to patientsby means of natural orifice (nasal, oral, transluminal) or asurgically created stoma, or both, consisting of a drainagelumen and inf
8、lation lumen (see Fig. 1). Common ballooninflation sizes are 5 cm3,15cm3, and 20 cm3.3.1.4 French size (Fr), na scale used for denoting the sizeof catheters and other tubular instruments. The French sizevalue is three times the outer diameter of the tube as measuredin millimetres. For example, a dia
9、meter of 18 Fr indicates adiameter of 6 mm.3.1.5 inflation volume, nvolume of liquid used to inflatethe retention balloon of the enteral feeding device for proposedtesting in this standard.3.1.6 rated volume, nstated volume of inflation of theretention balloon of the enteral feeding device in the ma
10、nufac-turers labeling and instructions for use.3.1.7 simulated gastric fluid, na solution consisting ofhydrochloric acid, salt and pepsin with a pH of approximately1.2, per USP standard recipe.3.1.8 sterility, nthe state of being free from viable micro-organisms.4. Specimen Preparation4.1 All test s
11、pecimens for test methods listed below shallconsist of the manufacturers new, finished, untested, unsteril-ized product.At the minimum, statistically valid samples of thesmallest and the largest diameter of enteral feeding devicesshall be tested.5. Test MethodsPROCEDURE A: FLOW RATE THROUGH FEEDINGL
12、UMEN5.1 ScopeThis test method covers the determination offlow rates through the drainage lumen of the enteral feedingdevice with retention balloon.5.2 Summary of Test MethodThe apparatus is set up asshown in Fig. 2. The flow rate is adjusted through the waterinlet to a rate sufficient to maintain fl
13、ow through the overflowoutlet while each enteral feeding device is tested. A headpressure of 20 6 1.0 cm of water (196 6 10 kPa) above the1These test methods are under the jurisdiction of ASTM Committee F04 onMedical and Surgical Materials and Devices and are the direct responsibility ofSubcommittee
14、 F04.35 on GI Applications.Current edition approved June 1, 2014. Published June 2014. Originallyapproved in 2006. Last previous edition approved in 2006 as F2528 06. DOI:10.1520/F2528-06R14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serv
15、iceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3USP Official Compendia of Standards, available from U.S. Pharmacopeia(USP), 12601 Twinbrook Pkwy., Rockville, MD 20852.Copyright ASTM International, 100 Barr Harbor Driv
16、e, PO Box C700, West Conshohocken, PA 19428-2959. United States1tank bottom shall be maintained throughout the test to approxi-mate actual physiological conditions. The overflow outletshould not be covered by water.5.3 Significance and UseThe flow rate is measured inreverse flow for ease in testing,
17、 since differences in the flowrate as a result of flow direction are theoretically insignificant.5.4 Apparatus:5.4.1 Water Reservoir, capable of maintaining 20 6 1.0 cm(7.9 6 0.4 in.) of water (196 6 10 kPa) above the tip of theenteral feeding device connection throughout the test as shownin Fig. 2.
18、 (See Performance Specification F623.)5.4.2 Graduated Cylinder, calibrated for suitable measure-ment of the effluent.5.4.3 Syringe, with appropriate tip for inflation of enteralfeeding device balloon.5.5 Hazards:5.5.1 Overflow should not be covered. Head pressure mustbe kept constant; water should a
19、lways be exiting through theoverflow outlet.5.5.2 Establish equilibrium before testing.5.5.3 Flow rates through all fittings must exceed that of theenteral feeding device being tested.5.6 Procedure:5.6.1 Test at 23 6 4C (73.4 6 7F).5.6.2 Inflate the retention balloon of the test specimen withwater t
20、o labeled volume.5.6.3 Connect the enteral feeding device to enteral feedingdevice connector and open the stopcock. The tip of the enteralfeeding device connection at the junction of enteral feedingdevice on-off valve should be level with the bottom of the tank61 cm and it should deliver fluid at 20
21、 6 1 cm (196 6 10 kPa)head pressure at that junction.5.6.4 Establish flow equilibrium before taking test measure-ments.5.6.5 Record the amount of fluid through the device feedinglumen in 30 seconds.5.7 Interpretation of ResultsFlow rates for enteral feedingdevices tested must meet or exceed 9 cm3/mi
22、n.5.8 Precision and BiasTo be determined within five years.PROCEDURE B: BALLOON BURST VOLUME5.9 ScopeThis test method covers the determination ofballoon integrity of enteral feeding devices with retentionballoon.5.10 Summary of Test MethodThe enteral feeding devicewith retention balloon is submerged
23、 in a small container filledwith water. The balloon is then inflated with water untilrupture, which enables the volume at which the balloon burststo be observed.5.11 Significance and UseThe balloon burst volume ismeasured to quantify the resistance of rupture of the enteralfeeding device with retent
24、ion balloon member.5.12 ApparatusThe testing apparatus is set up as shown inFig. 3.5.12.1 System Reservoir.5.12.2 Syringe.5.12.3 Water.5.13 HazardsWater should be emptied from system res-ervoir through purge valve when fill marked is reached.5.14 Procedure:5.14.1 Test at 23 6 4C (73.4 6 7F).5.14.2 I
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF252820062014STANDARDTESTMETHODSFORENTERALFEEDINGDEVICESWITHARETENTIONBALLOON 压力 气囊 装置 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-538921.html