ASTM F2528-2006 Standard Test Methods for Enteral Feeding Devices with a Retention Balloon《带压力气囊的肠给食装置标准试验方法》.pdf
《ASTM F2528-2006 Standard Test Methods for Enteral Feeding Devices with a Retention Balloon《带压力气囊的肠给食装置标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2528-2006 Standard Test Methods for Enteral Feeding Devices with a Retention Balloon《带压力气囊的肠给食装置标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 2528 06Standard Test Methods forEnteral Feeding Devices with a Retention Balloon1This standard is issued under the fixed designation F 2528; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision.
2、A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 These test methods cover the establishment of perfor-mance requirements for the utilization of a single-use, enteralfeeding device w
3、ith a retention balloon, used by medicalprofessionals for providing a means of nutrition and/or admin-istration of medication to patients by means of natural orifice(nasal, oral, transluminal) and or a surgically created stoma.The product is manufactured in various sizes and materialssuch as silicon
4、e, urethane, and various polymers (as well ascombinations of these) and is provided nonsterile for steriliza-tion and sterile for single use only. Rationale for these testmethods can be found in Appendix X1.1.2 This standard does not purport to address all of thesafety concerns, if any, associated w
5、ith its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and to determine theapplicability of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2F 623 Performance Specification for Foley Catheter2.2 Other Stan
6、dard:Simulated Gastric Fluid, USP Official Compendia of Stan-dards33. Terminology3.1 Definitions:3.1.1 balloon integrity (resistance to rupture)volume ofliquid that corresponds with balloon failure, or bursting.3.1.2 distalrefers to the balloon end of the enteral feedingdevice3.1.3 enteral feeding d
7、evice with retention balloona two-way medical device intended to provide a means of nutrition oradministration of medication, or both, to patients by means ofnatural orifice (nasal, oral, transluminal) or a surgically createdstoma, or both, consisting of a drainage lumen and inflationlumen (see Fig.
8、 1). Common balloon inflation sizes are 5 cm3,15 cm3, and 20 cm3.3.1.4 French size (Fr)a scale used for denoting the size ofcatheters and other tubular instruments. The French size valueis three times the outer diameter of the tube as measured inmillimetres. For example, a diameter of 18 Fr indicate
9、s adiameter of 6 mm.3.1.5 inflation volumevolume of liquid used to inflate theretention balloon of the enteral feeding device for proposedtesting in this standard.3.1.6 rated volumestated volume of inflation of the reten-tion balloon of the enteral feeding device in the manufacturerslabeling and ins
10、tructions for use.3.1.7 simulated gastric fluida solution consisting of hy-drochloric acid, salt and pepsin with a pH of approximately1.2, per USP standard recipe.3.1.8 sterilitythe state of being free from viable micro-organisms.4. Specimen Preparation4.1 All test specimens for test methods listed
11、below shallconsist of the manufacturers new, finished, untested, unsteril-ized product.At the minimum, statistically valid samples of thesmallest and the largest diameter of enteral feeding devicesshall be tested.5. Test MethodsPROCEDURE A: FLOW RATE THROUGH FEEDINGLUMEN5.1 ScopeThis test method cov
12、ers the determination offlow rates through the drainage lumen of the enteral feedingdevice with retention balloon.1These test methods are under the jurisdiction of ASTM Committee F04 onMedical and Surgical Materials and Devices and is the direct responsibility ofSubcommittee F04.35 on GI Appliations
13、.Current edition approved June 1, 2006. Published June 2006.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3
14、USP Official Compendia of Standards, available from U.S. Pharmacopeia(USP), 12601 Twinbrook Pkwy., Rockville, MD 20852.FIG. 1 Enteral Feeding Device with Retention Balloon1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.2 Summary of
15、 Test MethodThe apparatus is set up asshown in Fig. 2. The flow rate is adjusted through the waterinlet to a rate sufficient to maintain flow through the overflowoutlet while each enteral feeding device is tested. A headpressure of 20 6 1.0 cm of water (196 6 10 kPa) above thetank bottom shall be ma
16、intained throughout the test to approxi-mate actual physiological conditions. The overflow outletshould not be covered by water.5.3 Significance and UseThe flow rate is measured inreverse flow for ease in testing, since differences in the flowrate as a result of flow direction are theoretically insi
17、gnificant.5.4 Apparatus:5.4.1 Water Reservoir, capable of maintaining 20 6 1.0 cm(7.9 6 0.4 in.) of water (196 6 10 kPa) above the tip of theenteral feeding device connection throughout the test as shownin Fig. 2. (See Performance Specification F 623.)5.4.2 Graduated Cylinder, calibrated for suitabl
18、e measure-ment of the effluent.5.4.3 Syringe, with appropriate tip for inflation of enteralfeeding device balloon.5.5 Hazards:5.5.1 Overflow should not be covered. Head pressure mustbe kept constant; water should always be exiting through theoverflow outlet.5.5.2 Establish equilibrium before testing
19、.5.5.3 Flow rates through all fittings must exceed that of theenteral feeding device being tested.5.6 Procedure:5.6.1 Test at 23 6 4C (73.4 6 7F).5.6.2 Inflate the retention balloon of the test specimen withwater to labeled volume.5.6.3 Connect the enteral feeding device to enteral feedingdevice con
20、nector and open the stopcock. The tip of the enteralfeeding device connection at the junction of enteral feedingdevice on-off valve should be level with the bottom of the tank61 cm and it should deliver fluid at 20 6 1 cm (196 6 10 kPa)head pressure at that junction.5.6.4 Establish flow equilibrium
21、before taking test measure-ments.5.6.5 Record the amount of fluid through the device feedinglumen in 30 seconds.5.7 Interpretation of ResultsFlow rates for enteral feedingdevices tested must meet or exceed 9 cm3/min.5.8 Precision and BiasTo be determined within fiveyears.PROCEDURE B: BALLOON BURST V
22、OLUME5.9 ScopeThis test method covers the determination ofballoon integrity of enteral feeding devices with retentionballoon.5.10 Summary of Test MethodThe enteral feeding devicewith retention balloon is submerged in a small container filledwith water. The balloon is then inflated with water untilru
23、pture, which enables the volume at which the balloon burststo be observed.FIG. 2 Flow Rate ApparatusF25280625.11 Significance and UseThe balloon burst volume ismeasured to quantify the resistance of rupture of the enteralfeeding device with retention balloon member.5.12 ApparatusThe testing apparatu
24、s is set up as shown inFig. 3.5.12.1 System Reservoir.5.12.2 Syringe.5.12.3 Water.5.13 HazardsWater should be emptied from system res-ervoir through purge valve when fill marked is reached.5.14 Procedure:5.14.1 Test at 23 6 4C (73.4 6 7F).5.14.2 Insert uninflated enteral feeding device into testorif
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF25282006STANDARDTESTMETHODSFORENTERALFEEDINGDEVICESWITHARETENTIONBALLOON 压力 气囊 装置 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-538920.html