《ASTM F2182-2011a Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging《测定在磁共振成像中靠近被植入物或在其上方导致发热的无线电频.pdf》由会员分享,可在线阅读,更多相关《ASTM F2182-2011a Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging《测定在磁共振成像中靠近被植入物或在其上方导致发热的无线电频.pdf(15页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2182 11aStandard Test Method forMeasurement of Radio Frequency Induced Heating On orNear Passive Implants During Magnetic ResonanceImaging1This standard is issued under the fixed designation F2182; the number immediately following the designation indicates the year oforiginal adoption
2、or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers measurement of radio fre-quency (RF) induced heating
3、on or near a passive medicalimplant and its surroundings during magnetic resonance imag-ing (MRI).1.2 This test method is one required to determine if thepresence of a passive implant may cause injury to the patientwith the implant during an MR procedure. Other safety issuesthat should be addressed
4、include magnetically induced dis-placement force and torque, as well as proper device functionwhile in various configurations in the MR environment.1.3 The amount of RF-induced temperature rise for a givenspecific absorption rate (SAR) will depend on the RF fre-quency, which is dependent on the stat
5、ic magnetic fieldstrength of the MR system. While the focus in this test methodis on 1.5 Tesla (T) or 3 Tesla cylindrical bore MR systems, theRF-induced temperature rise for an implant in MR systems ofother static magnetic field strengths or magnet designs can beevaluated by suitable modification of
6、 the method describedherein.1.4 This test method assumes that testing is done on devicesthat will be entirely inside the body. For other implantationconditions (for example, external fixation devices, percutane-ous needles, catheters or tethered devices such as ablationprobes), modifications of this
7、 test method are necessary.1.5 This test method applies to whole body magneticresonance equipment, as defined in section 2.2.103 of the IECStandard 60601-2-33, Ed. 2.0, with a whole body RF transmitcoil as defined in section 2.2.100. The RF coil is assumed tohave quadrature excitation.1.6 The values
8、 stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.7 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safet
9、y and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2F2052 Test Method for Measurement of Magnetically In-duced Displacement Force on Medical Devices in theMagnetic Resonance EnvironmentF2119 Test Method for Evaluat
10、ion of MR Image Artifactsfrom Passive ImplantsF2213 Test Method for Measurement of Magnetically In-duced Torque on Medical Devices in the Magnetic Reso-nance EnvironmentF2503 Practice for Marking Medical Devices and OtherItems for Safety in the Magnetic Resonance Environment2.2 IEC Standard:360601-2
11、-33, Ed. 2.0 Medical Electrical EquipmentPart 2:Particular Requirements for the Safety of Magnetic Reso-nance Equipment for Medical Diagnosis, 20022.3 NEMA Standard:4NEMA MS 82008 Characterization of the Specific Ab-sorption Rate for Magnetic Resonance Imaging Systems3. Terminology3.1 Definitions:3.
12、1.1 gelled salinephantom medium consisting of sodiumchloride and polyacrylic acid or sodium chloride and hydroxy-ethylcellulose in water as specified in this test method.1This test method is under the jurisdiction of ASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct r
13、esponsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved April 15, 2011. Published August 2011. Originallyapproved in 2002. Last previous edition approved in 2011 as F2182 11. DOI:10.1520/F2182-11A.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orc
14、ontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from the International Electrotechnical Commission (IEC), 3 rue deVarembe, Case postale 131, CH-1211 Geneva 20, Switzerland
15、.4Available from National Electrical Manufacturers Association (NEMA), 1300N. 17th St., Suite 1752, Rosslyn, VA 22209, http:/www.nema.org.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.3.1.2 implant, nin medicine, an object, structu
16、re, or de-vice intended to reside within the body for diagnostic, pros-thetic, or other therapeutic purposes.3.1.3 isocentergeometric center of the gradient coil sys-tem, which generally is the geometric center of a scanner witha cylindrical bore.3.1.4 local SARspecific absorption rate (SAR) average
17、dover any 10 g of tissue of the patient body and over a specifiedtime. 60601-2-33, Ed. 2.03.1.5 magnetic resonance (MR) environmentvolumewithin the 0.50 mT (5 gauss (G) line of an MR system, whichincludes the entire three dimensional volume of space sur-rounding the MR scanner. For cases where the 0
18、.50 mT line iscontained within the Faraday shielded volume, the entire roomshall be considered the MR environment.3.1.6 magnetic resonance imaging (MRI)imaging tech-nique that uses static and time varying magnetic fields toprovide images of tissue by the magnetic resonance of nuclei.3.1.7 magnetic r
19、esonance system (MR system)ensembleof MR equipment, accessories including means for display,control, energy supplies, and the MR environment.60601-2-33, Ed. 2.03.1.8 MR Conditionalan item that has been demonstratedto pose no known hazards in a specified MR environment withspecified conditions of use
20、. Field conditions that define thespecified MR environment include field strength, spatial gra-dient, dB/dt (time rate of change of the magnetic field), radiofrequency (RF) fields, and specific absorption rate (SAR).Additional conditions, including specific configurations of theitem, may be required
21、.3.1.9 MR Safean item that poses no known hazards in allMR environments.NOTE 1MR Safe items include nonconducting, nonmagnetic itemssuch as a plastic petri dish. An item may be determined to be MR Safe byproviding a scientifically based rationale rather than test data.3.1.10 MR test systemMR system
22、or an apparatus thatreproduces the RF field of this type of system.3.1.11 MR Unsafean item that is known to pose hazardsin all MR environments.NOTE 2MR Unsafe items include magnetic items such as a pair offerromagnetic scissors.3.1.12 passive implantan implant that serves its functionwithout supply
23、of electrical power.3.1.13 radio frequency (RF) magnetic fieldthe magneticfield in MRI that is used to flip the magnetic moments. Thefrequency of the RF field is gB0where g is the gyromagneticconstant, 42.56 MHz/T for protons, and B0is the staticmagnetic field in Tesla.3.1.14 specific absorption rat
24、e (SAR)the mass normalizedrate at which RF energy is deposited in biological tissue. SARis typically indicated in W/kg.4. Summary of Test Method4.1 The implant to be tested is placed in a phantom materialthat simulates the electrical and thermal properties of thehuman body. The implant is placed at
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF21822011ASTANDARDTESTMETHODFORMEASUREMENTOFRADIOFREQUENCYINDUCEDHEATINGONORNEARPASSIVEIMPLANTSDURINGMAGNETICRESONANCEIMAGING

链接地址:http://www.mydoc123.com/p-537985.html